Reliability of the pinch strength with digitalized pinch dynamometer.

Heesuk Shin1, Seung Won Moon, Gab-Soon Kim, Jung Dong Park, Jin Hoan Kim, Mi Jin Jung, Chul Ho Yoon, Eun Shin Lee, Min-Kyun Oh.   

Abstract

OBJECTIVE: To examine the intra-rater, inter-rater, and inter-instrumental reliability of the digitalized pinch muscle strength dynamometer.
METHOD: Thirty normal subjects were examined for pinch strength, using both the Preston pinch gauge and the digitalized pinch dynamometer. The participants performed all pinch strength tests in the seated position as recommended by the American Society of Hand Therapists (ASHT). Three successive measurements were taken for each hand. The mean of the three trials was used for data analysis. The pinch strength tests performed used a repeated measure design and measurements were taken by each rater.
RESULTS: The relationship between the Preston pinch gauge and the digitalized pinch dynamometer in pinch strength was reliable (the ICC were 0.821 and 0.785 in rater 1 and rater 2 respectively). The relationship between the first session and second session in pinch strength using the digitalized pinch dynamometer was reliable (the ICC were 0.872 and 0.886 in rater A and rater B respectively). The relationship between rater A and rater B in pinch strength using the digitalized pinch dynamometer was reliable (the ICC was 0.754).
CONCLUSION: The pinch strength measurement using the digitalized pinch dynamometer is reliable within the rater and between raters. Thus, the Preston pinch gauge and the digitalized dynamometer measure grip strength equivalently, and can be used interchangeably.

Entities:  

Keywords:  Muscle strength dynamometer; Pinch strength; Reliability

Year:  2012        PMID: 22837976      PMCID: PMC3400880          DOI: 10.5535/arm.2012.36.3.394

Source DB:  PubMed          Journal:  Ann Rehabil Med        ISSN: 2234-0645


  13 in total

1.  Comparison of Rolyan and Jamar dynamometers for measuring grip strength.

Authors:  Virgil Mathiowetz
Journal:  Occup Ther Int       Date:  2002       Impact factor: 1.448

2.  Grip test; the use of a dynamometer with adjustable handle spacings.

Authors:  C O BECHTOL
Journal:  J Bone Joint Surg Am       Date:  1954-07       Impact factor: 5.284

3.  Grip strength: effects of testing posture and elbow position.

Authors:  J A Balogun; C T Akomolafe; L O Amusa
Journal:  Arch Phys Med Rehabil       Date:  1991-04       Impact factor: 3.966

4.  Reliability and validity of the DynEx dynamometer.

Authors:  Orit Shechtman; Lisa Gestewitz; Christine Kimble
Journal:  J Hand Ther       Date:  2005 Jul-Sep       Impact factor: 1.950

5.  A test case: does the availability of visual feedback impact grip strength scores when using a digital dynamometer?

Authors:  Gwen Weinstock-Zlotnick; Jane Bear-Lehman; Tzu-Ying Yu
Journal:  J Hand Ther       Date:  2011-03-31       Impact factor: 1.950

6.  Grip and pinch strength: norms for 6- to 19-year-olds.

Authors:  V Mathiowetz; D M Wiemer; S M Federman
Journal:  Am J Occup Ther       Date:  1986-10

7.  The need for reliability and validity in hand assessment instruments.

Authors:  E E Fess
Journal:  J Hand Surg Am       Date:  1986-09       Impact factor: 2.230

8.  Grip strength as measured by the Jamar dynamometer.

Authors:  R T Schmidt; J V Toews
Journal:  Arch Phys Med Rehabil       Date:  1970-06       Impact factor: 3.966

9.  Reliability and validity of grip and pinch strength evaluations.

Authors:  V Mathiowetz; K Weber; G Volland; N Kashman
Journal:  J Hand Surg Am       Date:  1984-03       Impact factor: 2.230

10.  Hand grip function in patients with rheumatoid arthritis.

Authors:  D B Myers; D M Grennan; D G Palmer
Journal:  Arch Phys Med Rehabil       Date:  1980-08       Impact factor: 3.966

View more
  5 in total

1.  Assessment of Motor Function in Peripheral Nerve Injury and Recovery.

Authors:  Albin John; Stephen Rossettie; John Rafael; Cameron Cox; Ivica Ducic; Brendan Mackay
Journal:  Orthop Rev (Pavia)       Date:  2022-09-13

2.  Hand Function Following Simulated Fusion of Thumb Metacarpophalangeal and Interphalangeal Joints.

Authors:  Joseph Paul Letzelter; Romana Ahmad; Jennifer Tagliarino; Jessica Woeckener; Ricardo Bello; Eitan Melamed
Journal:  Hand (N Y)       Date:  2020-02-26

3.  Reliability of pinch strength testing in elderly subjects with unilateral thumb carpometacarpal osteoarthritis.

Authors:  Jorge H Villafañe; Kristin Valdes
Journal:  J Phys Ther Sci       Date:  2014-07-30

4.  Assessment of hand function in women with lymphadenopathy after radical mastectomy.

Authors:  Subham Mistry; Taimul Ali; Mohammed Qasheesh; Rashid Ali Beg; Mohammad Abu Shaphe; Fuzail Ahmad; Faizan Z Kashoo; Amr S Shalaby
Journal:  PeerJ       Date:  2021-04-13       Impact factor: 2.984

5.  Relationship Between Grip and Pinch Strength and Activities of Daily Living in Stroke Patients.

Authors:  Jung Hyun Bae; Si Hyun Kang; Kyung Mook Seo; Don-Kyu Kim; Hyun Iee Shin; Hye Eun Shin
Journal:  Ann Rehabil Med       Date:  2015-10-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.