Literature DB >> 22825798

A thrombus susceptibility comparison of two pulsatile Penn State 50 cc left ventricular assist device designs.

Michael A Navitsky1, Steven Deutsch, Keefe B Manning.   

Abstract

Left ventricular assist devices (LVADs) have proven successful as bridge to transplant devices for patients awaiting donor organs. While survival rates continue to increase, destination therapy remains hindered by thrombus formation within the device. Research has shown that thrombosis is correlated to the fluid dynamics within the device and may be a result of sustained shear rates below 500 s(-1) on the polyurethane blood sac used in the Penn State pulsatile LVAD. Particle image velocimetry is used to compare flow within two 50 cc LVAD designs to assess fluid patterns and quantify wall shear rates in regions known from in vivo studies to be susceptible to thrombus formation. The two designs differ in their front face geometry. The V-1 model has an outward-facing "dome" whereas the face of the V-2 model is flat. A thrombus susceptibility metric, which uses measured wall shear rates and exposure times, was applied to objectively compare pump designs over the entire cardiac cycle. For each design, there are regions where wall shear rates remained below 500 s(-1) for the entire cardiac cycle resulting in high thrombus susceptibility potential. Results of this study indicate that the V-2 device had an overall lower propensity for thrombus formation in the current region of interest.

Entities:  

Mesh:

Year:  2012        PMID: 22825798      PMCID: PMC3524351          DOI: 10.1007/s10439-012-0627-z

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  18 in total

1.  Flow-induced platelet activation in mechanical heart valves.

Authors:  Danny Bluestein; Wei Yin; Klaus Affeld; Jolyon Jesty
Journal:  J Heart Valve Dis       Date:  2004-05

2.  Wall shear-rate estimation within the 50cc Penn State artificial heart using particle image velocimetry.

Authors:  Pramote Hochareon; Keefe B Manning; Arnold A Fontaine; John M Tarbell; Steven Deutsch
Journal:  J Biomech Eng       Date:  2004-08       Impact factor: 2.097

3.  Correlation of in vivo clot deposition with the flow characteristics in the 50 cc penn state artificial heart: a preliminary study.

Authors:  Pramote Hochareon; Keefe B Manning; Arnold A Fontaine; John M Tarbell; Steven Deutsch
Journal:  ASAIO J       Date:  2004 Nov-Dec       Impact factor: 2.872

4.  Off-design considerations of the 50cc Penn State Ventricular Assist Device.

Authors:  Leslie A Oley; Keefe B Manning; Arnold A Fontaine; Steven Deutsch
Journal:  Artif Organs       Date:  2005-05       Impact factor: 3.094

5.  Fluid dynamic analysis of the 50 cc Penn State artificial heart under physiological operating conditions using particle image velocimetry.

Authors:  Pramote Hochareon; Keefe B Manning; Arnold A Fontaine; John M Tarbell; Steven Deutsch
Journal:  J Biomech Eng       Date:  2004-10       Impact factor: 2.097

6.  Viscoelasticity of pediatric blood and its implications for the testing of a pulsatile pediatric blood pump.

Authors:  Jennifer A Long; Akif Undar; Keefe B Manning; Steven Deutsch
Journal:  ASAIO J       Date:  2005 Sep-Oct       Impact factor: 2.872

7.  LDA measurements of mean velocity and Reynolds stress fields within an artificial heart ventricle.

Authors:  J T Baldwin; S Deutsch; D B Geselowitz; J M Tarbell
Journal:  J Biomech Eng       Date:  1994-05       Impact factor: 2.097

8.  The LionHeart LVD-2000: a completely implanted left ventricular assist device for chronic circulatory support.

Authors:  S M Mehta; W E Pae; G Rosenberg; A J Snyder; W J Weiss; J P Lewis; D J Frank; J J Thompson; W S Pierce
Journal:  Ann Thorac Surg       Date:  2001-03       Impact factor: 4.330

9.  Visualization and analysis of mural thrombogenesis on collagen, polyurethane and nylon.

Authors:  J A Hubbell; L V McIntire
Journal:  Biomaterials       Date:  1986-09       Impact factor: 12.479

10.  The effect of fluid shear and co-adsorbed proteins on the stability of immobilized fibrinogen and subsequent platelet interactions.

Authors:  V Balasubramanian; S M Slack
Journal:  J Biomater Sci Polym Ed       Date:  2002       Impact factor: 3.517

View more
  4 in total

1.  The Use of Fluid Mechanics to Predict Regions of Microscopic Thrombus Formation in Pulsatile VADs.

Authors:  Stephen R Topper; Michael A Navitsky; Richard B Medvitz; Eric G Paterson; Christopher A Siedlecki; Margaret J Slattery; Steven Deutsch; Gerson Rosenberg; Keefe B Manning
Journal:  Cardiovasc Eng Technol       Date:  2014-03-01       Impact factor: 2.495

2.  Platelet adhesion to polyurethane urea under pulsatile flow conditions.

Authors:  Michael A Navitsky; Joshua O Taylor; Alexander B Smith; Margaret J Slattery; Steven Deutsch; Christopher A Siedlecki; Keefe B Manning
Journal:  Artif Organs       Date:  2014-04-09       Impact factor: 3.094

3.  Refining a numerical model for device-induced thrombosis and investigating the effects of non-Newtonian blood models.

Authors:  Ling Yang; Nicolas Tobin; Keefe B Manning
Journal:  J Biomech       Date:  2021-03-23       Impact factor: 2.712

4.  In vitro real-time magnetic resonance imaging for quantification of thrombosis.

Authors:  Ling Yang; Thomas Neuberger; Keefe B Manning
Journal:  MAGMA       Date:  2020-07-29       Impact factor: 2.310

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.