Fractured Ozurdex™ implant in the vitreous cavity.

Pukhraj Rishi, Gaurav Mathur, Ekta Rishi.   

Abstract

Entities:  

Mesh:

Year:  2012        PMID: 22824616      PMCID: PMC3442483          DOI: 10.4103/0301-4738.98734

Source DB:  PubMed          Journal:  Indian J Ophthalmol        ISSN: 0301-4738            Impact factor:   1.848


× No keyword cloud information.
Dear Editor, Retinal vein occlusion is the second most common retinal vascular disease after diabetic retinopathy.[1] Ozurdex™, an intravitreal implant, has been studied in the GENEVA trial.[2] Recently, there have been reports of migration of Ozurdex™ implant into the anterior chamber in aphakic eyes.[34] We report a case of fractured Ozurdex™ implant noted immediately after intravitreal injection. A 55-year-old-gentleman reported with blurring of vision in the right eye since 8 months. Best-corrected visual acuity (BCVA) in the right eye was 20/60, N8 and 20/20, N6 in the left eye. Right fundus revealed scattered retinal hemorrhages, old laser photocoagulation marks and macular edema [Fig. 1a]. Optical coherence tomography (OCT) revealed cystoid macular edema with foveolar detachment [Fig. 1b]. Electroretinogram (ERG) revealed reduced and delayed responses [Fig. 1c]. Clinical findings and investigations confirmed nonischemic Central retinal vein occlusion (CRVO) with persisting macular edema. The patient underwent intravitreal injection Ozurdex™ in the right eye. Fundus examination immediately after injection revealed the implant to be fractured into two pieces [Fig. 1d]. The injector was inspected for any residual drug pellet, and none was found. Having been assured of entire drug contents being delivered within the eye, our next concern was about increased elution of drug subsequent to increased surface area of pellets following its segmentation. Two months later, BCVA improved to 20/60, N6. Fundus examination revealed resolving retinal hemorrhages [Fig. 2a], dry-appearing macula and broken Ozurdex™ implant in vitreous cavity [Fig. 2b] and foveal contour was restored [Fig. 2c].
Figure 1

Right fundus revealed scattered retinal hemorrhages, old laser photocoagulation marks and macular edema (a). Optical coherence tomography of the right eye revealed cystoid macular edema with foveolar detachment; central macular thickness measured 553 microns (b). Electroretinogram revealed reduced and delayed scotopic and photopic responses (c). Fundus examination after the injection; the implant was noticed to be fractured in two pieces (d)

Figure 2

Two months after the injection, fundus examination revealed resolving retinal hemorrhages (a), a dry-appearing macula and a broken Ozurdex™ implant in the vitreous cavity (b). A significant decrease in the macular edema with a normal foveal dip was seen on optical coherence tomography. The central macular thickness measured 184 microns. No cystic spaces or subretinal fluid was seen (c)

Right fundus revealed scattered retinal hemorrhages, old laser photocoagulation marks and macular edema (a). Optical coherence tomography of the right eye revealed cystoid macular edema with foveolar detachment; central macular thickness measured 553 microns (b). Electroretinogram revealed reduced and delayed scotopic and photopic responses (c). Fundus examination after the injection; the implant was noticed to be fractured in two pieces (d) Two months after the injection, fundus examination revealed resolving retinal hemorrhages (a), a dry-appearing macula and a broken Ozurdex™ implant in the vitreous cavity (b). A significant decrease in the macular edema with a normal foveal dip was seen on optical coherence tomography. The central macular thickness measured 184 microns. No cystic spaces or subretinal fluid was seen (c) The recommended technique for intraocular injection of implant is described. The patient is prepared in the operating room with the same standards as those followed for intraocular procedures. Maintaining aseptic techniques, carefully remove the cap from the applicator. The safety tab is pulled straight off the applicator without twisting or flexing it. The long axis of the applicator is held parallel to the limbus and the sclera is engaged at an oblique angle with bevel of needle up to create a shelved scleral path, 4 mm away from limbus. The needle tip is advanced within the sclera for about 1 mm, then redirected towards the center of the eye and advanced until penetration of the sclera is completed and vitreous cavity is entered. The needle should not be advanced past the point where the sleeve touches the conjunctiva. The actuator button is slowly depressed until an audible click is noted. Before withdrawing the applicator from the eye, one makes sure that the actuator button is fully depressed and has locked flush with the applicator surface. The needle is removed in the same direction as while entering the vitreous. The fundus is examined. Despite following the prescribed technique for injection, we encountered an unusual adverse event, as described above. Possible causes could be (i) implant got cracked during manufacturing/packaging and “broke” during injection, being unable to endure the force of the injection process and (ii) possible misalignment of implant within the injector, leading to shearing forces breaking it during injection. In conclusion, the fractured Ozurdex™ implant was functionally stable despite its segmentation; it is difficult to say what exactly caused it. To the best of our knowledge, this “adverse event” has not been reported before and its possibility must be borne in mind by the treating physicians.
  3 in total

Review 1.  Retinal vein occlusion: beyond the acute event.

Authors:  Justis P Ehlers; Sharon Fekrat
Journal:  Surv Ophthalmol       Date:  2011-05-24       Impact factor: 6.048

2.  Randomized, sham-controlled trial of dexamethasone intravitreal implant in patients with macular edema due to retinal vein occlusion.

Authors:  Julia A Haller; Francesco Bandello; Rubens Belfort; Mark S Blumenkranz; Mark Gillies; Jeffrey Heier; Anat Loewenstein; Young-Hee Yoon; Marie-Louise Jacques; Jenny Jiao; Xiao-Yan Li; Scott M Whitcup
Journal:  Ophthalmology       Date:  2010-04-24       Impact factor: 12.079

3.  Wandering Ozurdex(®) implant.

Authors:  Reema Bansal; Pooja Bansal; Pandurang Kulkarni; Vishali Gupta; Aman Sharma; Amod Gupta
Journal:  J Ophthalmic Inflamm Infect       Date:  2011-09-30
  3 in total
  5 in total

1.  Surprises during intravitreal drug delivery-a report of three cases.

Authors:  Pradeep Venkatesh; Rohan Chawla; Bhavin Shah; Sat Pal Garg
Journal:  Digit J Ophthalmol       Date:  2017-03-15

2.  Comparison of the release profile and pharmacokinetics of intact and fragmented dexamethasone intravitreal implants in rabbit eyes.

Authors:  Rahul Bhagat; Jean Zhang; Sidiq Farooq; Xiao-Yan Li
Journal:  J Ocul Pharmacol Ther       Date:  2014-12       Impact factor: 2.671

Review 3.  Dexamethasone intravitreal implant in the treatment of diabetic macular edema.

Authors:  Pravin U Dugel; Francesco Bandello; Anat Loewenstein
Journal:  Clin Ophthalmol       Date:  2015-07-16

4.  Splitting of a Dexamethasone Implant (Ozurdex) following the Injection.

Authors:  Oya Donmez; Melih Parlak; Aylin Yaman; Ali Osman Saatci
Journal:  Case Rep Ophthalmol Med       Date:  2013-07-30

Review 5.  Role of implants in the treatment of diabetic macular edema: focus on the dexamethasone intravitreal implant.

Authors:  Zafer Cebeci; Nur Kir
Journal:  Diabetes Metab Syndr Obes       Date:  2015-11-16       Impact factor: 3.168

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.