BACKGROUND: Central arterial pressure is a better predictor of adverse cardiovascular outcomes than brachial blood pressure, but noninvasive measurement by applanation tonometry is technically demanding. METHOD: Pulsecor R6.5 is a novel device adapted from a standard sphygmomanometer which estimates the central aortic pressure from analysis of low-frequency suprasystolic waveforms at the occluded brachial artery. A physics-based model, which simulates the arterial system using elastic, thin-walled tube elements and Navier-Stokes equations, is used to calculate arterial pressure and flow propagation. To determine the reliability of the device, we compared 94 central systolic pressures estimated by Pulsecor to the simultaneous directly measured central aortic pressures at the time of coronary angiography in 37 individuals. RESULTS: There was good correlation in central SBP between catheter measurements and Pulsecor estimates by either invasive or noninvasive calibration methods (r = 0.99, P < 0.0001 and r = 0.95, P < 0.0001, respectively). The mean difference in central systolic pressure was 2.78 (SD 3.90) mmHg and coefficient of variation was 0.03 when the invasive calibration method was used.When the noninvasive calibration method was used, the mean difference in central systolic pressure was 0.25 (SD 6.31) mmHg and coefficient of variation was 0.05. CONCLUSION: We concluded that Pulsecor R6.5 provides a simple and easy method to noninvasively estimate central SBP, which has highly acceptable accuracy.
BACKGROUND: Central arterial pressure is a better predictor of adverse cardiovascular outcomes than brachial blood pressure, but noninvasive measurement by applanation tonometry is technically demanding. METHOD: Pulsecor R6.5 is a novel device adapted from a standard sphygmomanometer which estimates the central aortic pressure from analysis of low-frequency suprasystolic waveforms at the occluded brachial artery. A physics-based model, which simulates the arterial system using elastic, thin-walled tube elements and Navier-Stokes equations, is used to calculate arterial pressure and flow propagation. To determine the reliability of the device, we compared 94 central systolic pressures estimated by Pulsecor to the simultaneous directly measured central aortic pressures at the time of coronary angiography in 37 individuals. RESULTS: There was good correlation in central SBP between catheter measurements and Pulsecor estimates by either invasive or noninvasive calibration methods (r = 0.99, P < 0.0001 and r = 0.95, P < 0.0001, respectively). The mean difference in central systolic pressure was 2.78 (SD 3.90) mmHg and coefficient of variation was 0.03 when the invasive calibration method was used.When the noninvasive calibration method was used, the mean difference in central systolic pressure was 0.25 (SD 6.31) mmHg and coefficient of variation was 0.05. CONCLUSION: We concluded that Pulsecor R6.5 provides a simple and easy method to noninvasively estimate central SBP, which has highly acceptable accuracy.
Authors: C J Boos; E Vincent; A Mellor; D R Woods; C New; R Cruttenden; M Barlow; M Cooke; K Deighton; P Scott; S Clarke; J O'Hara Journal: J Hum Hypertens Date: 2017-05-25 Impact factor: 3.012
Authors: John D Sluyter; Alun D Hughes; Carlos A Camargo; Simon A McG Thom; Kim H Parker; Bernhard Hametner; Siegfried Wassertheurer; Robert Scragg Journal: Hypertension Date: 2019-08-26 Impact factor: 10.190
Authors: Nima Fazeli; Chang-Sei Kim; Mohammad Rashedi; Alyssa Chappell; Shaohua Wang; Roderick MacArthur; M Sean McMurtry; Barry Finegan; Jin-Oh Hahn Journal: Med Biol Eng Comput Date: 2014-09-03 Impact factor: 2.602
Authors: Paul A Roberts; Aaron C W Lin; Brett R Cowan; Alistair A Young; Ralph Stewart Journal: Int J Cardiovasc Imaging Date: 2017-11-08 Impact factor: 2.357
Authors: Anish N Bhuva; A D'Silva; C Torlasco; N Nadarajan; S Jones; R Boubertakh; J Van Zalen; P Scully; K Knott; G Benedetti; J B Augusto; Rachel Bastiaenen; G Lloyd; S Sharma; J C Moon; K H Parker; C H Manisty; Alun D Hughes Journal: Eur Heart J Cardiovasc Imaging Date: 2020-07-01 Impact factor: 6.875
Authors: James Faulkner; Louis Martinelli; Kirsty Cook; Lee Stoner; Helen Ryan-Stewart; Eloise Paine; Helen Hobbs; Danielle Lambrick Journal: J Spinal Cord Med Date: 2019-09-16 Impact factor: 1.985
Authors: John D Sluyter; Alun D Hughes; Andrew Lowe; Kim H Parker; Carlos A Camargo; Bernhard Hametner; Siegfried Wassertheurer; Robert K R Scragg Journal: Int J Cardiol Date: 2016-06-15 Impact factor: 4.164