Achieving the relocation accuracy of stereotactic frame-based cranial radiotherapy in a three-point thermoplastic shell.

N A Rosenfelder1, L Corsini, H McNair, K Pennert, K Burke, C M Lamb, A Aitken, S Ashley, V Khoo, M Brada.   

Abstract

AIMS: To compare the accuracy of fractionated cranial radiotherapy in a standard three-point thermoplastic shell using daily online correction with accuracy in a Gill-Thomas-Cosman relocatable stereotactic frame.
MATERIALS AND METHODS: All patients undergoing fractionated radiotherapy for benign intracranial tumours between March 2009 and August 2010 were included. Patients were immobilised in the frame with those unable to tolerate it immobilised in the shell. The ExacTrac imaging system was used for verification/correction. Daily online imaging before and after correction was carried out for shell patients and systematic and random population set-up errors calculated. These were compared with frame patients who underwent standard departmental imaging/correction with fractions 1-3 and weekly thereafter. Set-up margins were calculated from population errors.
RESULTS: Systematic and random errors were 0.3-0.7 mm/° before correction and 0.1-0.2 mm/° after correction in all axes in the frame, and 0.6-1.5 mm/° before correction and 0.1-0.4 mm/° after correction in the shell. Isotropic margins required for patient set-up could be reduced from 2 mm to <1 mm in the frame and from 5 mm to <1 mm in the shell.
CONCLUSION: Similar set-up accuracy can be achieved in the standard thermoplastic shell as in a relocatable frame despite less precise immobilisation. The use of daily online correction precludes the need for larger set-up margins.
Copyright © 2012. Published by Elsevier Ltd.

Entities:  

Mesh:

Year:  2012        PMID: 22795232     DOI: 10.1016/j.clon.2012.06.007

Source DB:  PubMed          Journal:  Clin Oncol (R Coll Radiol)        ISSN: 0936-6555            Impact factor:   4.126


  2 in total

1.  Evaluation of accuracy and reproducibility of a relocatable maxillary fixation system for fractionated intracranial stereotactic radiation therapy.

Authors:  Majella Russo; Rebecca Owen; Anne Bernard; Vaughan Moutrie; Matthew Foote
Journal:  J Med Radiat Sci       Date:  2015-09-22

2.  Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases.

Authors:  Kasirajan Tamil Selvan; Ganesan Padma; Murugesan Kesavan Revathy; N Arunai Nambi Raj; K Senthilnathan; P Ramesh Babu
Journal:  J Med Phys       Date:  2019 Apr-Jun
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.