Change in intraocular pressure measurement 2 years after myopic laser-assisted subepithelial keratectomy.

Gauti Jóhannesson1, Per Hallberg, Anders Eklund, Timo Koskela, Christina Lindén.   

Abstract

PURPOSE: To evaluate intraocular pressure (IOP) measurements 24 months after laser-assisted subepithelial keratectomy (LASEK) with 3 tonometry methods-Goldmann applanation tonometry (GAT), Pascal dynamic contour tonometry (DCT), and applanation resonance tonometry (ART)-and to compare them with measurements taken preoperatively and 3 and 6 months postoperatively.
SETTING: Department of Ophthalmology, Umeå University, Umeå, Sweden.
DESIGN: Prospective randomized single-center.
METHODS: One randomized eye per person from myopic healthy individuals who had LASEK was included. Visual acuity, central corneal thickness (CCT), corneal curvature, and IOP were measured at all time points. Six IOP measurements were performed per method, with a 5-minute lapse between methods.
RESULTS: The study evaluated 38 eyes. Two years after LASEK, dynamic ART was the only method that did not measure significantly lower IOP than before LASEK (mean -0.2 mm Hg ± 1.6 [SD]). The greatest difference was with GAT (mean -1.9 ± 1.7 mm Hg) followed by static ART (mean -1.2 ± 1.6 mm Hg) and then DCT (mean -0.9 ± 1.3 mm Hg). All methods measured significantly lower IOP at 3 months and 6 months.
CONCLUSIONS: Dynamic ART, which analyses IOP measurement during indentation, showed no significant difference in measured IOP after 24 months. The convex tip of the ART device and the continuous sampling of data during corneal indentation seem to be factors in why dynamic ART was less affected by corneal properties and therefore more suitable for IOP measurement in LASEK-treated eyes, although this must be confirmed in larger trials.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22789494     DOI: 10.1016/j.jcrs.2012.04.033

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Influence of LASEK on Schiøtz, Goldmann and dynamic contour Tonometry.

Authors:  Marco Sales-Sanz; Esther Arranz-Marquez; Carolina Arruabarrena; Miguel A Teus
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-10-14       Impact factor: 3.117

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.