Nathalie Morel1, Hervé Volland, Julie Dano, Patricia Lamourette, Patricia Sylvestre, Michèle Mock, Christophe Créminon.
Abstract
Bacillus anthracis is one of the most dangerous potential biological weapons, and it is essential to develop a rapid and simple method to detect B. anthracis spores in environmental samples. The immunoassay is a rapid and easy-to-use method for the detection of B. anthracis by means of antibodies directed against surface spore antigens. With this objective in view, we have produced a panel of monoclonal antibodies against B. anthracis and developed colorimetric and electrochemiluminescence (ECL) immunoassays. Using Meso Scale Discovery ECL technology, which is based on electrochemiluminescence (ECL) detection utilizing a sulfo-Tag label that emits light upon electrochemical stimulation (using a dedicated ECL plate reader, an electrical current is placed across the microplate with electrodes integrated into the bottom of the plate, resulting in a series of electrically induced reactions leading to a luminescent signal), a detection limit ranging between 0.3 × 10(3) and 10(3) CFU/ml (i.e., 30 to 100 spores per test), depending on the B. anthracis strain assayed, was achieved. In complex matrices (5 mg/ml of soil or simulated powder), the detection level (without any sample purification or concentration) was never altered more than 3-fold compared with the results obtained in phosphate-buffered saline.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22773632 PMCID: PMC3426714 DOI: 10.1128/AEM.01282-12