Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses.

T L Rose1, L S Robblee.   

Abstract

The charge injection limits of a Pt electrode using 0.2 ms charge balanced, biphasic current pulses ranged from 50 to 150 microC/cm2 geometric if the potential excursions of the electrode are kept below those at which H2 or O2 is produced. These charge densities are three to ten times smaller than the currently accepted value based on earlier experiments in which the reversible surface reactions were fully utilized and the pulse widths were longer.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2276759     DOI: 10.1109/10.61038

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  64 in total

1.  Resolution of the epiretinal prosthesis is not limited by electrode size.

Authors:  Matthew R Behrend; Ashish K Ahuja; Mark S Humayun; Robert H Chow; James D Weiland
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2011-04-19       Impact factor: 3.802

2.  Tissue and electrode capacitance reduce neural activation volumes during deep brain stimulation.

Authors:  Christopher R Butson; Cameron C McIntyre
Journal:  Clin Neurophysiol       Date:  2005-10       Impact factor: 3.708

3.  Simulation of intra-orbital optic nerve electrical stimulation.

Authors:  M Oozeer; C Veraart; V Legat; J Delbeke
Journal:  Med Biol Eng Comput       Date:  2005-09       Impact factor: 2.602

4.  Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes.

Authors:  S F Cogan; T D Plante; J Ehrlich
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

5.  Miniature Elastomeric Valve Design for Safe Direct Current Stimulator.

Authors:  Chaojun Cheng; Raviraj Thakur; Ankitha Rajagopalan Nair; Scott Sterrett; Gene Fridman
Journal:  IEEE Biomed Circuits Syst Conf       Date:  2018-03-29

6.  The development of a multichannel electrode array for retinal prostheses.

Authors:  Yasuo Terasawa; Hiroyuki Tashiro; Akihiro Uehara; Tohru Saitoh; Motoki Ozawa; Takashi Tokuda; Jun Ohta
Journal:  J Artif Organs       Date:  2006-12-21       Impact factor: 1.731

7.  High-resolution electrical stimulation of primate retina for epiretinal implant design.

Authors:  Chris Sekirnjak; Pawel Hottowy; Alexander Sher; Wladyslaw Dabrowski; Alan M Litke; E J Chichilnisky
Journal:  J Neurosci       Date:  2008-04-23       Impact factor: 6.167

8.  Safety assessment of epidural wire electrodes for cough production in a chronic pig model of spinal cord injury.

Authors:  Krzysztof E Kowalski; Tomasz Kowalski; Anthony F DiMarco
Journal:  J Neurosci Methods       Date:  2016-05-07       Impact factor: 2.390

9.  Highly stable carbon nanotube doped poly(3,4-ethylenedioxythiophene) for chronic neural stimulation.

Authors:  Xiliang Luo; Cassandra L Weaver; David D Zhou; Robert Greenberg; Xinyan T Cui
Journal:  Biomaterials       Date:  2011-05-20       Impact factor: 12.479

10.  Focal electrical stimulation of major ganglion cell types in the primate retina for the design of visual prostheses.

Authors:  Lauren H Jepson; Pawel Hottowy; Keith Mathieson; Deborah E Gunning; Wladyslaw Dabrowski; Alan M Litke; E J Chichilnisky
Journal:  J Neurosci       Date:  2013-04-24       Impact factor: 6.167