| Literature DB >> 22737012 |
Zdeněk Buchta1, Simon Reřucha, Břetislav Mikel, Martin Cížek, Josef Lazar, Ondřej Cíp.
Abstract
In this paper, a novel principle of contactless gauge block calibration is presented. The principle of contactless gauge block calibration combines low-coherence interferometry and laser interferometry. An experimental setup combines Dowell interferometer and Michelson interferometer to ensure a gauge block length determination with direct traceability to the primary length standard. By monitoring both gauge block sides with a digital camera gauge block 3D surface measurements are possible too. The principle presented is protected by the Czech national patent No. 302948.Entities:
Keywords: gauge block; low-coherence interferometry; nanometrology
Year: 2012 PMID: 22737012 PMCID: PMC3376568 DOI: 10.3390/s120303350
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.Optical setup for gauge blocks measurement. CP1, CP2 and CP3 are compensating plates, RS is a reference surface, λ/4 is a retardation quarterwave plate.
Figure 2.Sketch of the screenshot recorded by the CCD camera.
Figure 3.The Interference Detecting Signal (IDS) recorded during a pilot experiment with gauge block measurement. In the case shown in this figure, a gauge block with the length 17 mm was measured.
Figure 4.Sketch of the practical realization of the designed experimental setup.
Refractive index of air—standard uncertainty.
| xi | U(xi) | |||
|---|---|---|---|---|
| Atmospheric pressure | 99 kPa | 10 | 3E-9 /Pa | 2.7E-08 |
| Air temperature | 20°C | 0,01 | 1E-6 /K | 1.0E-08 |
| Relative humidity | 60% | 1 | 1E-8 /% | 1.0E-08 |
| CO2 concentration | 700 ppm | 10 | 1.5E-10/ppm | 1.5E-09 |
| Edlen formula | 1.0002650 | 2.E-08 | 1 | 2.0E-08 |
| RIA relative uncertainty | 3.6E-08 |
Uncertainty part given by mechanical parameters of the experimental setup
| xi | U(xi) | |||
|---|---|---|---|---|
| Gauge block thermal expansion | 20°C | 0.01 | 10.6E-6 | 1.1E-07 |
| Air temperature | 20°C | 0,01 | 12.3E-6 | 1.2E-07 |
| Mechanical components thermal expansion (gauge block holder, reference surface holder) | 20°C | 0,01 | 10.6E-6 | 1.1E-07 |
| Total mechanical relative uncertainty | 1.9E-07 |