| Literature DB >> 22734738 |
Yu Wang1, Todd N Erpelding, Ladislav Jankovic, Zijian Guo, Jean-Luc Robert, Guillaume David, Lihong V Wang.
Abstract
We present an integrated photoacoustic and ultrasonic three-dimensional (3-D) volumetric imaging system based on a two-dimensional (2-D) matrix array ultrasound probe. A wavelength-tunable dye laser pumped by a Q-switched Nd:YAG laser serves as the light source and a modified commercial ultrasound imaging system (iU22, Philips Healthcare) with a 2-D array transducer (X7-2, Philips Healthcare) detects both the pulse-echo ultrasound and photoacoustic signals. A multichannel data acquisition system acquires the RF channel data. The imaging system enables rendering of co-registered 3-D ultrasound and photoacoustic images without mechanical scanning. The resolution along the azimuth, elevation, and axial direction are measured to be 0.69, 0.90 and 0.84 mm for photoacoustic imaging. In vivo 3-D photoacoustic mapping of the sentinel lymph node was demonstrated in a rat model using methylene blue dye. These results highlight the clinical potential of 3-D PA imaging for identification of sentinel lymph nodes for cancer staging in humans.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22734738 PMCID: PMC3380932 DOI: 10.1117/1.JBO.17.6.061208
Source DB: PubMed Journal: J Biomed Opt ISSN: 1083-3668 Impact factor: 3.170