Accelerating image acquisition in 64-MDCT: the influence of scan parameters on image resolution and quality in a phantom study.

Thorsten Klink1, Marc Regier, Udo van Stevendaal, Michael Grass, Gerhard Adam, Philipp Begemann.   

Abstract

Computed tomographic (CT) image resolution and quality were evaluated utilizing varying scan protocols with accelerated image acquisition. A resolution phantom with hole diameters from 0.2 to 1.0 mm was scanned in axial, coronal, and sagittal plane using a 64-slice multidetector CT with varying scan parameters. No relevant differences in image resolution and quality were detected between the fastest scan protocol, with the shortest rotation time and highest pitch, and the slowest protocol. Accelerated CT protocols resulted in diagnostic images with adequate resolution and quality.
Copyright © 2012 Elsevier Inc. All rights reserved.

Mesh:

Year:  2012        PMID: 22726972     DOI: 10.1016/j.clinimag.2011.11.006

Source DB:  PubMed          Journal:  Clin Imaging        ISSN: 0899-7071            Impact factor:   1.605


  3 in total

1.  The effect of scan parameters on cone beam CT trabecular bone microstructural measurements of the human mandible.

Authors:  N Ibrahim; A Parsa; B Hassan; P van der Stelt; I H A Aartman; D Wismeijer
Journal:  Dentomaxillofac Radiol       Date:  2013-10-16       Impact factor: 2.419

2.  Influence of object location in different FOVs on trabecular bone microstructure measurements of human mandible: a cone beam CT study.

Authors:  N Ibrahim; A Parsa; B Hassan; P van der Stelt; I H A Aartman; P Nambiar
Journal:  Dentomaxillofac Radiol       Date:  2013-11-21       Impact factor: 2.419

3.  The effect of variations in CT scan protocol on femoral finite element failure load assessment using phantomless calibration.

Authors:  Ali Ataei; Jelle Eikhout; Ruud G H van Leeuwen; Esther Tanck; Florieke Eggermont
Journal:  PLoS One       Date:  2022-03-18       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.