Detection of bacterial contamination in platelet concentrates by a sensitive flow cytometric assay (BactiFlow): a multicentre validation study.

T Vollmer1, J Dreier, V Schottstedt, J Bux, K Tapernon, W Sibrowski, K Kleesiek, C Knabbe.   

Abstract

BACKGROUND: Bacterial contamination of platelet concentrates (PCs) still represents an ongoing risk. As a result of septic complications, particularly observed with older PCs, the shelf life of PCs has been reduced in Germany to 4 days. In this study, bacterial screening of PCs by BactiFlow (BF) flow cytometry was introduced in three German blood services to evaluate the robustness and applicability of the assay. Results were used to discuss the potential for the extension of PC shelf life to 5 days. STUDY DESIGN AND METHODS: A total of 1956 PCs were tested on days 4 or 5+ after PC production using the BF, whereas the BacT/Alert culture system served as reference method.
RESULTS: Two PCs were confirmed positive by culture only and were identified as Propionibacterium acnes and Staphylococcus species. Two PCs were confirmed positive for Streptococcus mitis by BF and culture. Additionally, two PCs were culture-positive only in one culture bottle (aerobic: S. mitis and anaerobic: S. hominis). Retrospective analysis of bacterial growth kinetics provide the indication that corresponding bacterial titres were most likely below the BF analytical detection limit (<150 CFU mL(-1) ) and had probably no transfusion relevance. All remaining specimens were tested negative.
CONCLUSIONS: Testing of PCs by BF was successfully implemented. The BF proved sufficient as a rapid screening method to improve PC safety. This study further provides data supporting the extension of PC shelf life to 5 days after negative BF testing on day 4.
© 2012 The Authors. Transfusion Medicine © 2012 British Blood Transfusion Society.

Entities:  

Mesh:

Year:  2012        PMID: 22724544     DOI: 10.1111/j.1365-3148.2012.01166.x

Source DB:  PubMed          Journal:  Transfus Med        ISSN: 0958-7578            Impact factor:   2.019


  4 in total

1.  Extension of the Storage Period of Platelet Concentrates in Germany to 5 Days by Bacterial Testing: Is it Worth the Effort?

Authors:  Tanja Vollmer; Dennis Hinse; Jürgen Diekmann; Cornelius Knabbe; Jens Dreier
Journal:  Transfus Med Hemother       Date:  2019-03-14       Impact factor: 3.747

Review 2.  Diagnostic methods for platelet bacteria screening: current status and developments.

Authors:  Melanie Störmer; Tanja Vollmer
Journal:  Transfus Med Hemother       Date:  2013-12-30       Impact factor: 3.747

3.  Bacterial screening of platelet concentrates on day 2 and 3 with flow cytometry: the optimal sampling time point?

Authors:  Tanja Vollmer; Volkmar Schottstedt; Juergen Bux; Gabriele Walther-Wenke; Cornelius Knabbe; Jens Dreier
Journal:  Blood Transfus       Date:  2014-01-02       Impact factor: 3.443

4.  Bench Test for the Detection of Bacterial Contamination in Platelet Concentrates Using Rapid and Cultural Detection Methods with a Standardized Proficiency Panel.

Authors:  Tanja Vollmer; Cornelius Knabbe; Wolf-Jochen Geilenkeuser; Michael Schmidt; Jens Dreier
Journal:  Transfus Med Hemother       Date:  2015-07-16       Impact factor: 3.747

  4 in total