P Kanyong1, R M Pemberton, S K Jackson, J P Hart.
Abstract
A screen-printed carbon electrode (SPCE) incorporating the electrocatalyst cobalt phthalocyanine (CoPC), fabricated using a water-based ink formulation, has been investigated as the base transducer for a uric acid biosensor. A sandwich biosensor was fabricated by first depositing cellulose acetate (CA) onto this transducer (CoPC-SPCE), followed by uricase (UOX) and finally a polycarbonate (PC) membrane; this device is designated PC-UOX-CA-CoPC-SPCE. This biosensor was used in conjunction with chronoamperometry to optimize the conditions for the analysis of urine: temperature, 35°C; buffer, pH 9.2; ionic strength, 50 mM; uricase, 0.6 U; incubation time, 180 s. The proposed biosensor was applied to urine from a healthy subject. The precision determined on unspiked urine (n=6) was 5.82%. Urine was fortified with 0.225 mM UA, and the resulting precision and recovery were 4.21 and 97.3%, respectively. The linear working range of the biosensor was found to be 0.015 to 0.25 mM (the former represents the detection limit), and the sensitivity was calculated to be 2.10 μA/mM.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22705172 DOI: 10.1016/j.ab.2012.05.027
Source DB: PubMed Journal: Anal Biochem ISSN: 0003-2697 Impact factor: 3.365