Literature DB >> 22684053

Estimating thyroid dose in pediatric CT exams from surface dose measurement.

Rani Al-Senan1, Deborah L Mueller, Mustapha R Hatab.   

Abstract

The purpose of this study was to investigate the possibility of estimating pediatric thyroid doses from CT using surface neck doses. Optically stimulated luminescence dosimeters were used to measure the neck surface dose of 25 children ranging in ages between one and three years old. The neck circumference for each child was measured. The relationship between obtained surface doses and thyroid dose was studied using acrylic phantoms of various sizes and with holes of different depths. The ratios of hole-to-surface doses were used to convert patients' surface dose to thyroid dose. ImPACT software was utilized to calculate thyroid dose after applying the appropriate age correction factors. A paired t-test was performed to compare thyroid doses from our approach and ImPACT. The ratio of thyroid to surface dose was found to be 1.1. Thyroid doses ranged from 20 to 80 mGy. Comparison showed no statistical significance (p = 0.18). In addition, the average of surface dose variation along the z-axis in helical scans was studied and found to range between 5% (in 10 cm diameter phantom/24 mm collimation/pitch 1.0) and 8% (in 16 cm diameter phantom/12 mm collimation/pitch 0.7). We conclude that surface dose is an acceptable predictor for pediatric thyroid dose from CT. The uncertainty due to surface dose variability may be reduced if narrower collimation is used with a pitch factor close to 1.0. Also, the results did not show any effect of thyroid depth on the measured dose.

Entities:  

Mesh:

Year:  2012        PMID: 22684053     DOI: 10.1088/0031-9155/57/13/4211

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  A method to acquire CT organ dose map using OSL dosimeters and ATOM anthropomorphic phantoms.

Authors:  Da Zhang; Xinhua Li; Yiming Gao; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

2.  Thyroid doses and risk to paediatric patients undergoing neck CT examinations.

Authors:  Maria Vittoria Spampinato; Sameer Tipnis; Joshua Tavernier; Walter Huda
Journal:  Eur Radiol       Date:  2015-02-02       Impact factor: 5.315

3.  Radiation dose in the thyroid and the thyroid cancer risk attributable to CT scans for pediatric patients in one general hospital of China.

Authors:  Yin-Ping Su; Hao-Wei Niu; Jun-Bo Chen; Ying-Hua Fu; Guo-Bing Xiao; Quan-Fu Sun