Paul Philipp Heinisch1, Oliver Richter, Michael Schünke, Rene Bombien Quaden.
Abstract
Transcatheter heart valve implantation can be performed transapically and transfemorally. The transfemoral way to the aortic valve is significantly longer than the transapical one. The aim of this study was to analyse the intima of 15 human aortas after the deployment of different conventional valved stent delivery systems. Fifteen human aortas have been analysed (77 ± 8.4 years). These aortas were preserved with formalin and explanted from the common iliac arteries to the ascending aorta. After protocolling all relevant vascular parameters, the deployment force of different conventional valved stent delivery systems was analysed. After that, the intima was closely investigated by endoscopy. The deployment force of the old catheter was not different from the actual system. The endoscopic investigation revealed significant intimal damages in all parts of the aorta after deployment of the delivery system. This study demonstrated that the passage to the aortic valve can result in significant intimal damage regardless of the used deployment catheter. Efforts are necessary to lower the profile of the deployment devices to increase their flexibility. The intima of the aorta and possible damage have to get back into the physicians' focus to avoid possible late aortic complications.Entities:
Mesh:
Year: 2012 PMID: 22659269 PMCID: PMC3422938 DOI: 10.1093/icvts/ivs225