A patient-specific quality assurance study on absolute dose verification using ionization chambers of different volumes in RapidArc treatments.

S A Syam Kumar1, Prabakar Sukumar, Padmanaban Sriram, Dhanabalan Rajasekaran, Srinu Aketi, Nagarajan Vivekanandan.   

Abstract

The recalculation of 1 fraction from a patient treatment plan on a phantom and subsequent measurements have become the norms for measurement-based verification, which combines the quality assurance recommendations that deal with the treatment planning system and the beam delivery system. This type of evaluation has prompted attention to measurement equipment and techniques. Ionization chambers are considered the gold standard because of their precision, availability, and relative ease of use. This study evaluates and compares 5 different ionization chambers: phantom combinations for verification in routine patient-specific quality assurance of RapidArc treatments. Fifteen different RapidArc plans conforming to the clinical standards were selected for the study. Verification plans were then created for each treatment plan with different chamber-phantom combinations scanned by computed tomography. This includes Medtec intensity modulated radiation therapy (IMRT) phantom with micro-ionization chamber (0.007 cm(3)) and pinpoint chamber (0.015 cm(3)), PTW-Octavius phantom with semiflex chamber (0.125 cm(3)) and 2D array (0.125 cm(3)), and indigenously made Circular wax phantom with 0.6 cm(3) chamber. The measured isocenter absolute dose was compared with the treatment planning system (TPS) plan. The micro-ionization chamber shows more deviations when compared with semiflex and 0.6 cm(3) with a maximum variation of -4.76%, -1.49%, and 2.23% for micro-ionization, semiflex, and farmer chambers, respectively. The positive variations indicate that the chamber with larger volume overestimates. Farmer chamber shows higher deviation when compared with 0.125 cm(3). In general the deviation was found to be <1% with the semiflex and farmer chambers. A maximum variation of 2% was observed for the 0.007 cm(3) ionization chamber, except in a few cases. Pinpoint chamber underestimates the calculated isocenter dose by a maximum of 4.8%. Absolute dose measurements using the semiflex ionization chamber with intermediate volume (0.125 cm(3)) shows good agreement with the TPS calculated among the detectors used in this study. Positioning is very important when using smaller volume chambers because they are more sensitive to geometrical errors within the treatment fields. It is also suggested to average the dose over the sensitive volume for larger-volume chambers. The ionization chamber-phantom combinations used in this study can be used interchangeably for routine RapidArc patient-specific quality assurance with a satisfactory accuracy for clinical practice.
Copyright © 2012 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22626968     DOI: 10.1016/j.meddos.2012.04.001

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers.

Authors:  Henry Finlay Godson; Ravikumar Manickam; Sathiyan Saminathan; Kadirampatti Mani Ganesh; Retna Ponmalar
Journal:  Radiol Phys Technol       Date:  2016-12-01

2.  Calibration and error analysis of metal-oxide-semiconductor field-effect transistor dosimeters for computed tomography radiation dosimetry.

Authors:  Sigal Trattner; Peter Prinsen; Jens Wiegert; Elazar-Lars Gerland; Efrat Shefer; Tom Morton; Carla M Thompson; Yoad Yagil; Bin Cheng; Sachin Jambawalikar; Rani Al-Senan; Maxwell Amurao; Sandra S Halliburton; Andrew J Einstein
Journal:  Med Phys       Date:  2017-10-26       Impact factor: 4.071

3.  A robust measurement point for dose verification in delivery quality assurance for a robotic radiosurgery system.

Authors:  Keita Kurosu; Iori Sumida; Hiroya Shiomi; Hirokazu Mizuno; Hiroko Yamaguchi; Hirofumi Okubo; Keisuke Tamari; Yuji Seo; Osamu Suzuki; Seiichi Ota; Shinichi Inoue; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2017-05-01       Impact factor: 2.724

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.