Literature DB >> 22624783

The durability of the intrascaphoid compression of headless compression screws: in vitro study.

D S Gruszka1, K J Burkhart, T E Nowak, T Achenbach, P M Rommens, L P Müller.   

Abstract

PURPOSE: To test a new generation of compression screws: the Acumed Acutrak 2 Mini (AA; Acumed, Hillsboro, OR), the Stryker TwinFix (ST; Stryker, Kalamazoo, MI), and the Synthes 3.0 headless compression screw (SH; Synthes, Solothurn, Switzerland).
METHODS: We used 40 fresh-frozen human scaphoids for this study. Bone density was measured. A K-wire was inserted centrally. A perpendicular osteotomy was created in the middle third (Herbert B2 fracture). A custom-made load sensor was placed between the bone fragments. All screws were implanted according to the manufacturers' instructions. The Synthes 2.0 cortical screw (SC), implanted as a lag screw, was used as a reference. The compression force during each experiment was digitally monitored for 12 hours while the data were acquired. The data were analyzed using analysis of variance with the Bonferroni correction.
RESULTS: Immediately after screw insertion, ST reached 226 N, followed by AA with 191 N, SH with 137 N, and SC with 72 N. After 12 hours, ST displayed the highest residual compression force, 141 N, followed by AA with 121 N, SH with 78 N, and SC with 32 N. The differences were significant for ST and AA compared to SC. The loss of compression force over 12 hours was 39% for ST, 42% for AA, 49% for SH, and 55% for SC.
CONCLUSIONS: The new generation of headless compression screws, especially ST and AA, provided significantly higher compression forces after 12 hours, as well as the least loss of compression force over time, in comparison to a classic cortical lag screw. CLINICAL RELEVANCE: A new generation of headless compression screws, by producing higher compression forces, increase stability at the fracture site and might thereby promote bone healing.
Copyright © 2012 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22624783     DOI: 10.1016/j.jhsa.2012.03.018

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  12 in total

1.  Insertion profiles of 4 headless compression screws.

Authors:  Adam Hart; Edward J Harvey; Louis-Philippe Lefebvre; Francois Barthelat; Reza Rabiei; Paul A Martineau
Journal:  J Hand Surg Am       Date:  2013-06-25       Impact factor: 2.230

2.  Variable Bone Density of Scaphoid: Importance of Subchondral Screw Placement.

Authors:  Morgan M Swanstrom; Kyle W Morse; Joseph D Lipman; Krystle A Hearns; Michelle G Carlson
Journal:  J Wrist Surg       Date:  2017-08-08

3.  Short-Term Outcomes of Herbert Screw Fixation for Isolated Olecranon Fractures in Children: A Single-Institution Retrospective Study.

Authors:  Yang Li; Kelai Wang; Dong Sun; Yakun Liu; Zhe Wang; Aiwu Li
Journal:  Indian J Orthop       Date:  2022-02-18       Impact factor: 1.033

4.  Distal interphalangeal joint arthrodesis with nonaxial multiple small screws: a biomechanical analysis with axial headless compression screw and clinical result of 15 consecutive cases.

Authors:  Seung Hun Woo; Sang Ho Kwak; Hyo Seok Jang; Dong Hee Kim; Jang Hyeon Seo; Sang Hyun Lee
Journal:  BMC Musculoskelet Disord       Date:  2022-05-27       Impact factor: 2.562

5.  Effect of Screw Perpendicularity on Compression in Scaphoid Waist Fractures.

Authors:  Morgan M Swanstrom; Kyle W Morse; Joseph D Lipman; Krystle A Hearns; Michelle G Carlson
Journal:  J Wrist Surg       Date:  2016-12-01

6.  Outcome of isolated olecranon fractures in skeletally immature patients: comparison of open reduction and tension band wiring fixation versus closed reduction and percutaneous screw fixation.

Authors:  Marco Corradin; Lorenza Marengo; Antonio Andreacchio; Matteo Paonessa; Vanessa Giacometti; Antoine Samba; Marie Rousset; Federico Canavese
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-04-30

7.  Postoperative Outcomes of Volar Plate Fixation in Cases of Scaphoid Deformity or Nonunion: A Case Series.

Authors:  Saif A Ansari; James A Kennedy; Fizan Younis
Journal:  J Wrist Surg       Date:  2020-05-20

8.  Which Headless Compression Screw Produces the Highest Interfragmentary Compression Force in Scaphoid Fracture?

Authors:  Karthik Vishwanathan; Ravi Patel; Sumedh Talwalkar
Journal:  Indian J Orthop       Date:  2020-04-22       Impact factor: 1.251

9.  A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size.

Authors:  Asif M Ilyas; Jonathan M Mahoney; Brandon S Bucklen
Journal:  Hand (N Y)       Date:  2019-09-30

10.  Rotational Stability of Scaphoid Osteosyntheses: An In Vitro Comparison of Small Fragment Cannulated Screws to Novel Bone Screw Sets.

Authors:  Jochen Erhart; Ewald Unger; Philip Schefzig; Peter Varga; Inga Trulson; Anna Gormasz; Alexander Trulson; Martin Reschl; Michael Hagmann; Vilmos Vecsei; Winfried Mayr
Journal:  PLoS One       Date:  2016-06-03       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.