Abstract
A simple, accurate, and precise HPLC method was developed and validated for determination of rimonabant in human plasma. Following liquid-liquid extraction, chromatographic separation was accomplished using C18 column with mobile phase consisting of acetonitrile : water (90 : 10, v/v), drug was detected at 260 nm using UVdetector. The LOD and LOQ were 3.0 and 10.0 μg/L, respectively. The method is linear in the interval 50.0-1000.0 μg/L. The average extraction recovery of drug from plasma was found to be 92.2%. The percent CV of the method was found to be less than 10.8%, and accuracy was found between 94.5 and 106.7%. The assay may be applied to a pharmacokinetic and bioequivalence study of rimonabant.Entities:
Year: 2012 PMID: 22619744 PMCID: PMC3347727 DOI: 10.1155/2012/625979
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Figure 1Structure of rimonabant.
Figure 2Chromatogram of blank plasma.
Figure 3Chromatogram of low quality control sample.
Imprecision and inaccuracy results of validation.
| Spiked conc. ( | Conc. found mean ± SD ( | CV (%) | Accuracy (%) |
|---|---|---|---|
| Intraday imprecision and inaccuracy ( | |||
| 150 | 151.8 ± 6.00 | 4.0 | 101.2 |
| 450 | 442.3 ± 19.51 | 4.4 | 98.3 |
| 800 | 780.6 ± 25.30 | 3.2 | 97.6 |
| Interbatch imprecision and inaccuracy ( | |||
| 150 | 153.9 ± 5.20 | 3.4 | 102.6 |
| 450 | 430.5 ± 16.70 | 3.9 | 95.7 |
| 800 | 810.6 ± 30.3 | 3.7 | 101.3 |
| Interday imprecision and inaccuracy ( | |||
| 150.0 | 156.4 ± 8.04 | 5.1 | 104.3 |
| 450.0 | 435.2 ± 17.90 | 4.1 | 96.7 |
| 800.0 | 756.2 ± 20.45 | 2.7 | 94.5 |
| Analyst to analyst imprecision and inaccuracy ( | |||
| 150.0 | 160.0 ± 10.30 | 6.4 | 106.7 |
| 450.0 | 458.3 ± 9.80 | 2.1 | 101.8 |
| 800.0 | 780.2 ± 24.90 | 3.2 | 97.5 |
| Global imprecision and inaccuracy ( | |||
| 150.0 | 152.9 ± 6.60 | 4.3 | 101.9 |
| 450.0 | 436.4 ± 22.05 | 5.1 | 97.0 |
| 800.0 | 795.6 ± 36.59 | 4.6 | 99.5 |
Extraction recovery of rimonabant from plasma (n = 6).
| Analyte | Conc. added ( | Mean recovery ± SD (%) | CV (%) |
|---|---|---|---|
| Rimonabant | 150.0 | 93.9 ± 12.00 | 8.5 |
| 450.0 | 91.3 ± 22.20 | 5.4 | |
| 800.0 | 91.3 ± 14.60 | 2.0 |
The average global recovery = 92.2%, and percent CV of global recoveries = 1.6%.
Bench top and freeze-thaw stability of rimonabant (n = 6).
| Nominal conc. ( | Mean found conc. ( | CV (%) | CV (%) | ||
|---|---|---|---|---|---|
| Bench top stability | 0 hrs | 12 hrs | 24 hrs | 12 hrs | 24 hrs |
| 150.0 | 155.2 | 144.5 | 135.8 | 7.9 | 10.8 |
| 800.0 | 785.3 | 767.6 | 760.9 | 1.0 | 1.6 |
| Freeze-thaw stability | Day 0 | Day 1 | Day 2 | CV day (%) 1 | CV day (%) 2 |
| 150.0 | 149.5 | 147.4 | 143.2 | 4.6 | 5.4 |
| 800.0 | 789.9 | 795.6 | 789.1 | 0.5 | 1.4 |