Literature DB >> 22610718

Causes of IOL explantation in Spain.

Roberto Fernández-Buenaga1, Jorge L Alio, Francisco J Muñoz-Negrete, Rafael I Barraquer Compte, Jorge L Alio-Del Barrio.   

Abstract

PURPOSE: To study the reasons and the demography of pseudophakic intraocular lens (IOL) explantation in Spain.
METHODS: In this observational multicenter retrospective study, the cases studied correspond to the 15 centers that constitute the Nodo Calidad Visual y Cirugia Refractiva of the Red Tematica de Investigacion Cooperativa (RETICS) sponsored by the Spanish Ministry of Health. Clinical data from all the patients who underwent explantation were assessed. The different reasons that caused the decision of explantation were analyzed.
RESULTS: A total of 257 explanted pseudophakic IOLs have been studied. Patients' mean age when explantation occurred was 67.5 years (SD 13.5 [22-99]) and 135 were female (52.5%). The main causes for explantation were dislocation/decentration in 145 cases (56.3%), incorrect lens power in 33 cases (12.8%), IOL opacification in 29 eyes (11.3%), neuroadaptation failure in 16 cases (6.2%), pseudophakic bullous keratopathy in 6 eyes (2.3%), endophthalmitis in 5 cases (1.9%), and "other causes" in 23 eyes (8.9%). Treatment after explantation was posterior chamber IOL implantation in 149 eyes (58%), anterior chamber IOL implantation in 49 eyes (19.1%), aphakia in 39 eyes (15.2%), and missing information in 20 cases (7.8%). Mean time from implantation to explantation was 3.97 (SD 4.68 [0.005-21.1]) years.
CONCLUSIONS: Dislocation/decentration (most of the time with lens in the bag) is the main cause for explantation in Spain, followed by incorrect lens power (which decreased greatly over past years) and IOL opacification. Posterior chamber IOL implantation is the most elected treatment after explantation.

Entities:  

Mesh:

Year:  2012        PMID: 22610718     DOI: 10.5301/ejo.5000168

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  10 in total

1.  Population-based Incidence of Intraocular Lens Exchange in Olmsted County, Minnesota.

Authors:  Erick D Bothun; Lilian C B Cavalcante; David O Hodge; Sanjay V Patel
Journal:  Am J Ophthalmol       Date:  2018-01-02       Impact factor: 5.258

2.  A pilot study of intraocular lens explantation in 69 eyes in Chinese patients.

Authors:  Fang Chai; Bo Ma; Xin-Guang Yang; Juan Li; Mei-Fang Chu
Journal:  Int J Ophthalmol       Date:  2017-04-18       Impact factor: 1.779

3.  Intraocular lens explantation in Chinese patients: different patterns and different responses.

Authors:  Tommy C Y Chan; Jerry K H Lok; Vishal Jhanji; Victoria W Y Wong
Journal:  Int Ophthalmol       Date:  2014-09-03       Impact factor: 2.031

4.  Intraocular lens explantation in Spain: indications and outcomes at a tertiary referral center from 2010 to 2018.

Authors:  Mª Victoria de Rojas; Sara Viña; Antía Gestoso; Patricia Simón; Marcelino Álvarez
Journal:  Int Ophthalmol       Date:  2019-09-29       Impact factor: 2.031

5.  Perioperative complications and clinical outcomes of intraocular lens exchange in patients with opacified lenses.

Authors:  Roberto Fernández-Buenaga; Jorge L Alió; Laura Pinilla-Cortés; Rafael I Barraquer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-07-03       Impact factor: 3.117

6.  Intraocular Lens Fragmentation Using Femtosecond Laser: An In Vitro Study.

Authors:  Chandra Bala; Jeffrey Shi; Kerrie Meades
Journal:  Transl Vis Sci Technol       Date:  2015-06-11       Impact factor: 3.283

Review 7.  Plasticity in the human visual cortex: an ophthalmology-based perspective.

Authors:  Andreia Martins Rosa; Maria Fátima Silva; Sónia Ferreira; Joaquim Murta; Miguel Castelo-Branco
Journal:  Biomed Res Int       Date:  2013-09-25       Impact factor: 3.411

Review 8.  Multifocal intraocular lenses: Types, outcomes, complications and how to solve them.

Authors:  Liberdade C Salerno; Mauro C Tiveron; Jorge L Alió
Journal:  Taiwan J Ophthalmol       Date:  2017 Oct-Dec

9.  Patients' dissatisfaction with multifocal intraocular lenses managed by exchange with other multifocal lenses of different optical profiles.

Authors:  Olena Al-Shymali; Colm McAlinden; Jorge L Alio Del Barrio; Mario Canto-Cerdan; Jorge L Alio
Journal:  Eye Vis (Lond)       Date:  2022-03-01

10.  Fundamentals of Presbyopia: visual processing and binocularity in its transformation.

Authors:  Olga I Rozanova; Andrey G Shchuko; Tatyana S Mischenko
Journal:  Eye Vis (Lond)       Date:  2018-01-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.