| Literature DB >> 22608683 |
William L Griffin1, Thomas K Fehring, James C Kudrna, Robert H Schmidt, Michael J Christie, Susan M Odum, Anne C Dennos.
Abstract
The purpose of this study was to determine if cobalt and chromium ion levels can predict soft tissue damage at total hip revision. This study included 90 metal-on-metal total hip patients with preoperative cobalt and chromium ion levels. Tissue damage noted at revision surgery was graded on a 4-point scale. Sensitivity, specificity, and predictive values were calculated for various threshold values. Receiver operating characteristic analysis was conducted. Using 7 ppb as a threshold, cobalt and chromium ion levels had poor sensitivity and specificity (Co, 65% and 56%; Cr, 29% and 75%). Positive predictive values for cobalt and chromium were only 48% and 26% respectively. The area under the curve was 0.37 for cobalt and 0.44 for chromium. The length of time to revision significantly correlated with tissue damage (P = .001). Ion levels are unreliable predictors of periarticular soft tissue damage and should not be used in isolation as surgical intervention triggers.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22608683 DOI: 10.1016/j.arth.2012.03.020
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757