Organ-based computed tomographic (CT) radiation dose reduction to the lenses: impact on image quality for CT of the head.

Anja Judith Reimann1, Chris Davison, Thor Bjarnason, Yogesh Thakur, Thakur Yogesh, Karl Kryzmyk, John Mayo, Savvas Nicolaou.   

Abstract

PURPOSE: Recently, a new specific organ dose adaption and reduction protocol, or SODAR tool (X-CARE, Siemens Healthcare), which reduces dose to the anterior aspect of the body of patients, was installed on our computed tomographic scanner. The purpose of this pilot project was to evaluate image quality and dose distribution in the acquired data with the new protocol.
MATERIALS AND METHODS: Sixteen consecutive patients were scanned with the new SODAR head protocol. The findings were compared with 16 matched patients who were imaged with the standard computed tomographic head trauma protocol. Image quality was assessed qualitatively using a scale of 1 to 4 (1, excellent; 2, good; 3, fair; 4, nondiagnostic). Additionally, 1-cm regions of interest were placed in the white matter of the cerebral hemispheres, the cerebellar hemispheres, and the brain stem at the level of the pons for a quantitative analysis. The standard deviation of each measurement was recorded as an indicator for image noise. Dose measurement trials were performed using optically stimulated luminescence dosimeters on head phantoms and then on patients.
RESULTS: Subjective image quality ranged between 1 and 3; no scan areas were considered nondiagnostic. Overall image quality of the posterior fossa averaged at 1.656 was slightly reduced compared to the cerebral hemispheres (mean, 1.141). The mean standard protocol brain stem image quality was 1.604, with only minimal deterioration to 1.708 in the SODAR group.No significant difference in image noise could be found between the SODAR group with a mean noise of 4.515 and standard images with a mean of 4.721 (P > 0.05).The dose to the anterior aspect of the patient was lowered to 3.2 mGy compared to 4.5 mGy on the lateral aspect of the scan (P > 0.05). To compensate for the photon loss in the posterior aspect, the dose has to be slightly increased to a mean of 6 mGy, but overall, a significant dose reduction with stable image quality could be achieved by reducing the dose length product from 1489 to 1347 mGy·cm using SODAR (P < 0.0001).
CONCLUSION: Using the SODAR protocol resulted not only in an impressive 46% to 59% frontal dose reduction but also in the overall dose reduction. This dose reduction was obtained without sacrificing image quality, providing diagnostic images of the brain while protecting radiosensitive structures like the eye lenses in trauma brain imaging. Future applications will be reducing dose to other radiosensitive structures such as the thyroid gland and breast tissue from potentially harmful low-energy radiation without compromising image quality.

Entities:  

Mesh:

Year:  2012        PMID: 22592620     DOI: 10.1097/RCT.0b013e318251ec61

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  8 in total

1.  Attenuation correction using deep learning for brain perfusion SPECT images.

Authors:  Kenta Sakaguchi; Hayato Kaida; Shuhei Yoshida; Kazunari Ishii
Journal:  Ann Nucl Med       Date:  2021-03-09       Impact factor: 2.668

2.  Introduction of a Dedicated Emergency Department MR Imaging Scanner at the Barrow Neurological Institute.

Authors:  M Buller; J P Karis
Journal:  AJNR Am J Neuroradiol       Date:  2017-05-11       Impact factor: 3.825

3.  Impact of organ-specific dose reduction on the image quality of head and neck CT angiography.

Authors:  L Schimmöller; R S Lanzman; P Heusch; S Dietrich; F Miese; J Aissa; T A Heusner; G Antoch; P Kröpil
Journal:  Eur Radiol       Date:  2013-01-09       Impact factor: 5.315

4.  Evaluation of the impact of organ-specific dose reduction on image quality in pediatric chest computed tomography.

Authors:  Johannes Boos; Patric Kröpil; Dirk Klee; Philipp Heusch; Lars Schimmöller; Jörg Schaper; Gerald Antoch; Rotem S Lanzman
Journal:  Pediatr Radiol       Date:  2014-03-29

5.  Evaluation of Virtual Noncontrast Images Obtained from Dual-Energy CTA for Diagnosing Subarachnoid Hemorrhage.

Authors:  X Y Jiang; S H Zhang; Q Z Xie; Z J Yin; Q Y Liu; M D Zhao; X L Li; X J Mao
Journal:  AJNR Am J Neuroradiol       Date:  2015-01-22       Impact factor: 3.825

6.  The effect of vertical centering and scout direction on automatic tube voltage selection in chest CT: a preliminary phantom study on two different CT equipments.

Authors:  Touko Kaasalainen; Teemu Mäkelä; Mika Kortesniemi
Journal:  Eur J Radiol Open       Date:  2018-12-17

7.  Lens Identification to Prevent Radiation-Induced Cataracts Using Convolutional Neural Networks.

Authors:  Ross Filice
Journal:  J Digit Imaging       Date:  2019-08       Impact factor: 4.056

8.  Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study.

Authors:  Ying-Lan Liao; Nan-Ku Lai; Yeu-Sheng Tyan; Hui-Yu Tsai
Journal:  Biomed J       Date:  2019-11-04       Impact factor: 4.910

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.