Linear attenuation coefficient and buildup factor of MCP-96 alloy for dose accuracy, beam collimation, and radiation protection.

Deidre N Hopkins1, Muhammad Maqbool, Mohammed S Islam.   

Abstract

The linear attenuation coefficients and buildup factor of MCP-96 alloy were determined for (60)Co, (54)Mn, and (137)Cs gamma emitters and a NaI detector. The thickness of the MCP-96 attenuator was varied from 1 to 4 cm. A collimated beam of gamma rays was allowed to pass through various thicknesses of the MCP-96 alloy. The attenuated beam was detected by a NaI detector, and data were recorded by a multichannel analyzer. The run was repeated without the collimator for broad-beam geometry. For each run, the attenuated beam intensity was normalized by the intensity of the unattenuated incident beam obtained by removing the attenuators. Linear attenuation coefficients were determined by plotting of the intensity of the collimated beam against the attenuator thickness. For every thickness of the alloy, the ratio of the attenuated to the unattenuated beam was found to be higher in broad-beam geometry as compared to the same ratio in narrow-beam geometry. We used the difference in these ratios in broad and narrow-beam geometries to calculate the buildup factor. The buildup factor was found to increase with beam energy and attenuator thickness. Variation in the source-to-detector distance gave a lower value of the buildup factor for a small and a large distance and a higher value for an intermediate distance. The buildup factor was found to be greater than 1 in all cases. We conclude that the buildup factor must be calculated and incorporated for dose correction and precision when the MCP-96 alloy is used for tissue compensation or radiation shielding and protection purposes.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22585280     DOI: 10.1007/s12194-012-0158-1

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  12 in total

1.  Clinical implementation of a computer controlled milling machine for compensating filter production.

Authors:  M Partridge; E Donovan; N Fenton; S Reise; S Blane
Journal:  Br J Radiol       Date:  1999-11       Impact factor: 3.039

2.  Late normal tissue injury from permanent interstitial implants.

Authors:  R J Yaes
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-03-15       Impact factor: 7.038

3.  Photon scatter in intensity modulating filters evaluated by first Compton scatter and Monte Carlo calculations and experiments in broad beams.

Authors:  Y Mejaddem; S Hyödynmaa; A Brahme
Journal:  Phys Med Biol       Date:  2000-10       Impact factor: 3.609

4.  Compensators for three-dimensional treatment planning.

Authors:  G S Mageras; R Mohan; C Burman; G D Barest; G J Kutcher
Journal:  Med Phys       Date:  1991 Mar-Apr       Impact factor: 4.071

5.  Intensity-modulated radiation therapy for mesothelioma: impact of multileaf collimator leaf width and pencil beam size on planning quality and delivery efficiency.

Authors:  X Ronald Zhu; Karl Prado; H Helen Liu; Thomas M Guerrero; Melenda Jeter; Zhongxing Liao; David Rice; Kenneth Forster; Craig W Stevens
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-08-01       Impact factor: 7.038

6.  Compensation in three-dimensional non-coplanar treatment planning.

Authors:  G J Kutcher; C Burman; R Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  1991-01       Impact factor: 7.038

7.  Collimator (head) scatter at extended distances in linear accelerator-generated photon beams.

Authors:  F M Khan; J P Gibbons; D M Roback
Journal:  Int J Radiat Oncol Biol Phys       Date:  1996-06-01       Impact factor: 7.038

8.  Procedure for accurate fabrication of tissue compensators with high-density material.

Authors:  Y Mejaddem; I Lax; K Shamsuddin Adakkai
Journal:  Phys Med Biol       Date:  1997-02       Impact factor: 3.609

9.  Calculation of dose distributions in the vicinity of high-Z interfaces for photon beams.

Authors:  O A Sauer
Journal:  Med Phys       Date:  1995-10       Impact factor: 4.071

10.  Precision in reporting the dose given in a course of radiotherapy.

Authors:  S Dische; M I Saunders; C Williams; A Hopkins; E Aird
Journal:  Radiother Oncol       Date:  1993-12       Impact factor: 6.280

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.