Pacing polarity and left ventricular mechanical activation sequence in cardiac resynchronization therapy.

Hyan Suk Yang1, Giuseppe Caracciolo, Partho P Sengupta, Ramil Goel, Krishnaswamy Chandrasekaran, Komandoor Srivathsan.   

Abstract

OBJECTIVE: The aim of this study is to evaluate the relationship between polarity of left ventricular (LV) pacing and the resultant regional, global, and transmural mechanical sequence of contraction.
BACKGROUND: Cardiac resynchronization therapy (CRT) is widely utilized in patients with drug refractory congestive heart failure with systolic dysfunction (EF <35 %) and intraventricular conduction delay (QRS duration >120 ms). However, little is known about polarity of pacing stimulation and the resultant differences in LV mechanics.
METHODS: The polarity of pacing was altered sequentially in 20 patients (73 ± 13, 16 males) with preexisting biventricular devices with potential choice of multiple vectors for pacing stimulation. Initial unipolar or extended bipolar configurations were switched to bipolar configuration or vice versa, and echocardiographic images were acquired for off-line analysis. Regional and global LV longitudinal and radial mechanics were assessed selectively from the subendocardial and subepicardial regions with 2D speckle-tracking echocardiography. Left ventricular capture by each vector configuration was confirmed by local lead capture and appropriate QRS alteration.
RESULTS: Unipolar pacing resulted in increased dispersion of LV regional endocardial strains with a higher base-to-apex gradients of longitudinal shortening strains (P < 0.05). LV longitudinal shortening strain magnitude was higher at LV base with bipolar stimulation in comparison with unipolar stimulation (-10.5 ± 10.5 vs. -4.2 ± 6.3, P = 0.02).
CONCLUSION: There is a difference in the mechanical activation sequence of the LV between unipolar vs. bipolar pacing stimulation. This may have important implications for CRT.

Entities:  

Mesh:

Year:  2012        PMID: 22580715     DOI: 10.1007/s10840-012-9686-1

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  16 in total

1.  Cardiac resynchronization in chronic heart failure.

Authors:  William T Abraham; Westby G Fisher; Andrew L Smith; David B Delurgio; Angel R Leon; Evan Loh; Dusan Z Kocovic; Milton Packer; Alfredo L Clavell; David L Hayes; Myrvin Ellestad; Robin J Trupp; Jackie Underwood; Faith Pickering; Cindy Truex; Peggy McAtee; John Messenger
Journal:  N Engl J Med       Date:  2002-06-13       Impact factor: 91.245

2.  Spread of activation in the left ventricular wall of the dog. I.

Authors:  D DURRER; L H VAN DER TWELL
Journal:  Am Heart J       Date:  1953-11       Impact factor: 4.749

3.  A comparison of QRS complexes resulting from unipolar and bipolar pacing: implications for pace-mapping.

Authors:  A H Kadish; S Schmaltz; F Morady
Journal:  Pacing Clin Electrophysiol       Date:  1991-05       Impact factor: 1.976

4.  Intraventricular conduction delay: a prognostic marker in chronic heart failure.

Authors:  W Shamim; D P Francis; M Yousufuddin; S Varney; M F Pieopli; S D Anker; A J Coats
Journal:  Int J Cardiol       Date:  1999-07-31       Impact factor: 4.164

5.  Development and prospective validation of a clinical index to predict survival in ambulatory patients referred for cardiac transplant evaluation.

Authors:  K D Aaronson; J S Schwartz; T M Chen; K L Wong; J E Goin; D M Mancini
Journal:  Circulation       Date:  1997-06-17       Impact factor: 29.690

6.  Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The Pacing Therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research Group.

Authors:  A Auricchio; C Stellbrink; M Block; S Sack; J Vogt; P Bakker; H Klein; A Kramer; J Ding; R Salo; B Tockman; T Pochet; J Spinelli
Journal:  Circulation       Date:  1999-06-15       Impact factor: 29.690

7.  Low settings of the ventricular pacing output in patients dependent on a pacemaker: are they really safe?

Authors:  Andreas Schuchert; Jens Frese; Ekkehard Stammwitz; Miroslav Novák; Arthur Schleich; Stefan M Wagner; Thomas Meinertz
Journal:  Am Heart J       Date:  2002-06       Impact factor: 4.749

8.  Natural history of abnormal conduction and its relation to prognosis in patients with dilated cardiomyopathy.

Authors:  H B Xiao; C Roy; S Fujimoto; D G Gibson
Journal:  Int J Cardiol       Date:  1996-02       Impact factor: 4.164

9.  Long-term prognosis of patients presenting to the emergency room with decompensated congestive heart failure.

Authors:  J M Brophy; G Deslauriers; J L Rouleau
Journal:  Can J Cardiol       Date:  1994-06       Impact factor: 5.223

10.  Electrically unexcitable scar mapping based on pacing threshold for identification of the reentry circuit isthmus: feasibility for guiding ventricular tachycardia ablation.

Authors:  Kyoko Soejima; William G Stevenson; William H Maisel; John L Sapp; Laurence M Epstein
Journal:  Circulation       Date:  2002-09-24       Impact factor: 29.690

View more
  1 in total

1.  The impact of the left ventricular pacing polarity and localization during cardiac resynchronization therapy on depolarization and repolarization parameters.

Authors:  Emin Evren Özcan; Ali Öztürk; Erdem Özel; Ömer Senarslan; Bela Merkely; Laszlo Geller
Journal:  Anatol J Cardiol       Date:  2018-03-21       Impact factor: 1.596

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.