The potential capabilities and analytical performance of lab-on-valve (LOV) manifold as a front end to amperometry have been explored for the on-line determination of morin. Meanwhile, the electrochemical behaviors of morin were investigated based on polyvinylpyrrolidone- (PVP-) doped carbon paste electrode (CPE), which found that PVP can significantly improve its oxidation peak current. The excellent amperometric current response was achieved when the potential difference (ΔE) of 0.6 V was implemented in pH 6.5 phosphate buffer solution (PBS) that served as the supporting electrolyte. A well-defined oxidation peak has been obtained in studies using PVP as a modifier of CPE based on the oxidation of morin. The present work introduces the LOV technique as a useful tool for amperometric measurement, documents advantages of using programmable flow, and outlines means for miniaturization of assays on the basis of PVP modified CPE. The proposed method was applied successfully to the determination of morin in real samples, and the spiked recoveries were satisfactory.Entities:
Year: 2012 PMID: 22567551 PMCID: PMC3335390 DOI: 10.1155/2012/257109
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Figure 1The structures of PVP (a) and morin (b).
Figure 2Schematic diagram of the SI-LOV manifold in conjunction with amperometry for the determination of morin.
Figure 3Cyclic voltammograms obtained at bare CPE (b) and PVP/CPE (a) surfaces using 1.0 × 10−5 mol L−1 morin solution in 0.05 mol L−1 PBS (pH 6.5).
Figure 4Cyclic voltammograms obtained at PVP/CPE in 0.05 mol L−1 PBS with various scan rates. The scan rates were 25, 50, 100, 150, 200, 250, 300 mV/s from a to g. Inset: plot of the peak current as a function of the scan rate.
Figure 5The amperometric current response with increasing concentrations of morin. ΔE of 0.6 V was applied. Sample volume was 200 μL with the flow rate of 30 μL s−1.
Sequential injection lab-on-valve format hyphenated with amperometry based on PVP/CPE with applications to the analysis of morin in real samplesa.
| Samples | Original | Spiked | Total found | Recovery (%) | RSD (%, |
|---|---|---|---|---|---|
| (10−5 mol L−1) | (10−5 mol L−1) | (10−5 mol L−1) | |||
| Tea | 1.76 ± 0.28 | 2.00 | 3.70 ± 0.35 | 97.0 ± 0.6 | 2.6 |
| 4.00 | 5.86 ± 0.44 | 102.5 ± 0.9 | 2.7 | ||
|
| |||||
| Mulberry leaves | 5.13 ± 0.27 | 6.00 | 11.27 ± 0.46 | 102.3 ± 1.1 | 3.1 |
| 8.00 | 13.42 ± 0.53 | 103.6 ± 1.2 | 3.6 | ||
aThe results were obtained at 95% confidence level.