Cone beam computed tomography image guidance system for a dedicated intracranial radiosurgery treatment unit.

Mark Ruschin1, Philip T Komljenovic, Steve Ansell, Cynthia Ménard, Gregory Bootsma, Young-Bin Cho, Caroline Chung, David Jaffray.   

Abstract

PURPOSE: Image guidance has improved the precision of fractionated radiation treatment delivery on linear accelerators. Precise radiation delivery is particularly critical when high doses are delivered to complex shapes with steep dose gradients near critical structures, as is the case for intracranial radiosurgery. To reduce potential geometric uncertainties, a cone beam computed tomography (CT) image guidance system was developed in-house to generate high-resolution images of the head at the time of treatment, using a dedicated radiosurgery unit. The performance and initial clinical use of this imaging system are described. METHODS AND MATERIALS: A kilovoltage cone beam CT system was integrated with a Leksell Gamma Knife Perfexion radiosurgery unit. The X-ray tube and flat-panel detector are mounted on a translational arm, which is parked above the treatment unit when not in use. Upon descent, a rotational axis provides 210° of rotation for cone beam CT scans. Mechanical integrity of the system was evaluated over a 6-month period. Subsequent clinical commissioning included end-to-end testing of targeting performance and subjective image quality performance in phantoms. The system has been used to image 2 patients, 1 of whom received single-fraction radiosurgery and 1 who received 3 fractions, using a relocatable head frame.
RESULTS: Images of phantoms demonstrated soft tissue contrast visibility and submillimeter spatial resolution. A contrast difference of 35 HU was easily detected at a calibration dose of 1.2 cGy (center of head phantom). The shape of the mechanical flex vs scan angle was highly reproducible and exhibited <0.2 mm peak-to-peak variation. With a 0.5-mm voxel pitch, the maximum targeting error was 0.4 mm. Images of 2 patients were analyzed offline and submillimeter agreement was confirmed with conventional frame.
CONCLUSIONS: A cone beam CT image guidance system was successfully adapted to a radiosurgery unit. The system is capable of producing high-resolution images of bone and soft tissue. The system is in clinical use and provides excellent image guidance without invasive frames.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22560556     DOI: 10.1016/j.ijrobp.2012.03.022

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Stereotactic radiosurgery of glomus jugulare tumors: current concepts, recent advances and future perspectives.

Authors:  Omer Sager; Ferrat Dincoglan; Murat Beyzadeoglu
Journal:  CNS Oncol       Date:  2015

2.  Simulational study of a dosimetric comparison between a Gamma Knife treatment plan and an intensity-modulated radiotherapy plan for skull base tumors.

Authors:  Hisato Nakazawa; Yoshimasa Mori; Masataka Komori; Takahiko Tsugawa; Yuta Shibamoto; Tatsuya Kobayashi; Chisa Hashizume; Yukio Uchiyama; Masahiro Hagiwara
Journal:  J Radiat Res       Date:  2013-12-17       Impact factor: 2.724

3.  Preliminary Evaluation of a Novel Thermoplastic Mask System with Intra-fraction Motion Monitoring for Future Use with Image-Guided Gamma Knife.

Authors:  Winnie Li; Gregory Bootsma; Oscar Von Schultz; Per Carlsson; Normand Laperriere; Barbara-Ann Millar; David Jaffray; Caroline Chung
Journal:  Cureus       Date:  2016-03-13

4.  Evaluation of the stability of the stereotactic Leksell Frame G in Gamma Knife radiosurgery.

Authors:  Alvaro Rojas-Villabona; Katherine Miszkiel; Neil Kitchen; Rolf Jäger; Ian Paddick
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

5.  Significant tumor shift in patients treated with stereotactic radiosurgery for brain metastasis.

Authors:  Eline D Hessen; Laurens D van Buuren; Jasper A Nijkamp; Kim C de Vries; Wai Kong Mok; Luc Dewit; Anke M van Mourik; Alejandro Berlin; Uulke A van der Heide; Gerben R Borst
Journal:  Clin Transl Radiat Oncol       Date:  2017-01-26

6.  The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases.

Authors:  Ans C C Swinnen; Michel C Öllers; Chin Loon Ong; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2020-04-01       Impact factor: 2.102

7.  Cone-Beam CT image contrast and attenuation-map linearity improvement (CALI) for brain stereotactic radiosurgery procedures.

Authors:  SayedMasoud Hashemi; Christopher Huynh; Arjun Sahgal; William Y Song; Håkan Nordström; Markus Eriksson; James G Mainprize; Young Lee; Mark Ruschin
Journal:  J Appl Clin Med Phys       Date:  2018-10-19       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.