Literature DB >> 22559652

Cone-beam breast computed tomography with a displaced flat panel detector array.

Giovanni Mettivier1, Paolo Russo, Nico Lanconelli, Sergio Lo Meo.   

Abstract

PURPOSE: In cone-beam computed tomography (CBCT), and in particular in cone-beam breast computed tomography (CBBCT), an important issue is the reduction of the image artifacts produced by photon scatter and the reduction of patient dose. In this work, the authors propose to apply the detector displacement technique (also known as asymmetric detector or "extended view" geometry) to approach this goal. Potentially, this type of geometry, and the accompanying use of a beam collimator to mask the unirradiated half-object in each projection, permits some reduction of radiation dose with respect to conventional CBBCT and a sizeable reduction of the overall amount of scatter in the object, for a fixed contrast-to-noise ratio (CNR).
METHODS: The authors consider a scan configuration in which the projection data are acquired from an asymmetrically positioned detector that covers only one half of the scan field of view. Monte Carlo simulations and measurements, with their CBBCT laboratory scanner, were performed using PMMA phantoms of cylindrical (70-mm diameter) and hemiellipsoidal (140-mm diameter) shape simulating the average pendant breast, at 80 kVp. Image quality was evaluated in terms of contrast, noise, CNR, contrast-to-noise ratio per unit of dose (CNRD), and spatial resolution as width of line spread function for high contrast details.
RESULTS: Reconstructed images with the asymmetric detector technique deviate less than 1% from reconstruction with a conventional symmetric detector (detector view) and indicate a reduction of the cupping artifact in CT slices. The maximum scatter-to-primary ratio at the center of the phantom decreases by about 50% for both small and large diameter phantoms (e.g., from 0.75 in detector view to 0.40 in extended view geometry at the central axis of the 140-mm diameter PMMA phantom). Less cupping produces an increase of the CT number accuracy and an improved image detail contrast, but the associated increase of noise observed may produce a decrease of detail CNR. By simulating the energy deposited inside the phantoms, the authors evaluated a maximum 50% reduction of the absorbed dose at the expense of a decrease of CNR, for the half beam irradiation of the object performed with the displaced detector technique with respect to full beam irradiation. The decrease in CNR, and in absorbed dose as well, translates into a detail CNRD showing values comparable to or higher than the ones obtained for a conventional symmetric detector technique, attributed to the effect of decreased scatter in particular at the axis of the irradiated object. An estimate is provided (about 12%) for the average dose reduction possible in CBBCT at constant CNR for the average uncompressed breast (14 cm diameter, 50% glandularity), in case of minimum image overlapping in extended view.
CONCLUSIONS: Simulations and experiments show that CBCT reconstructions with the displaced detector technique and with a half beam collimator are less affected by scatter artifacts, which could lead to some decrease of the radiation dose to the irradiated object with respect to a conventional reconstruction. This dose reduction is associated with increase of noise, decrease of CNR, but equal or improved CNRD values. The use of a small area detector would allow also to reduce the apparatus cost and to improve the data transfer speed with a corresponding increment of frame rate.

Entities:  

Mesh:

Year:  2012        PMID: 22559652     DOI: 10.1118/1.4704641

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Cone-beam breast computed tomography using ultra-fast image reconstruction with constrained, total-variation minimization for suppression of artifacts.

Authors:  Hsin Wu Tseng; Srinivasan Vedantham; Andrew Karellas
Journal:  Phys Med       Date:  2020-04-28       Impact factor: 2.685

2.  Radiochromic film dosimetry in synchrotron radiation breast computed tomography: a phantom study.

Authors:  Giovanni Mettivier; Marica Masi; Fulvia Arfelli; Luca Brombal; Pasquale Delogu; Francesca Di Lillo; Sandro Donato; Christian Fedon; Bruno Golosio; Piernicola Oliva; Luigi Rigon; Antonio Sarno; Angelo Taibi; Paolo Russo
Journal:  J Synchrotron Radiat       Date:  2020-04-22       Impact factor: 2.616

3.  Cone-beam breast CT using an offset detector: effect of detector offset and image reconstruction algorithm.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Phys Med Biol       Date:  2022-04-07       Impact factor: 4.174

4.  Dedicated breast CT: geometric design considerations to maximize posterior breast coverage.

Authors:  Srinivasan Vedantham; Andrew Karellas; Margaret M Emmons; Lawrence J Moss; Sarwat Hussain; Stephen P Baker
Journal:  Phys Med Biol       Date:  2013-05-17       Impact factor: 3.609

5.  Personalized estimates of radiation dose from dedicated breast CT in a diagnostic population and comparison with diagnostic mammography.

Authors:  Srinivasan Vedantham; Linxi Shi; Andrew Karellas; Avice M O'Connell; David L Conover
Journal:  Phys Med Biol       Date:  2013-10-29       Impact factor: 3.609

6.  Dedicated cone-beam breast CT using laterally-shifted detector geometry: Quantitative analysis of feasibility for clinical translation.

Authors:  Srinivasan Vedantham; Hsin-Wu Tseng; Souleymane Konate; Linxi Shi; Andrew Karellas
Journal:  J Xray Sci Technol       Date:  2020       Impact factor: 1.535

7.  Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2021-01-26       Impact factor: 4.506

8.  X-ray interferometry without analyzer for breast CT application: a simulation study.

Authors:  Jingzhu Xu; Kyungmin Ham; Joyoni Dey
Journal:  J Med Imaging (Bellingham)       Date:  2020-03-26

9.  A residual dense network assisted sparse view reconstruction for breast computed tomography.

Authors:  Zhiyang Fu; Hsin Wu Tseng; Srinivasan Vedantham; Andrew Karellas; Ali Bilgin
Journal:  Sci Rep       Date:  2020-12-03       Impact factor: 4.379

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.