Completeness map evaluation demonstrated with candidate next-generation cardiac CT architectures.

Baodong Liu1, James Bennett, Ge Wang, Bruno De Man, Kai Zeng, Zhye Yin, Paul Fitzgerald, Hengyong Yu.   

Abstract

PURPOSE: In this report, the authors introduce the general concept of the completeness map, as a means to evaluate the completeness of data acquired by a given CT system design (architecture and scan mode). They illustrate the utility of completeness map by applying the completeness map concept to a number of candidate CT system designs, as part of a study to advance the state-of-the-art in cardiac CT.
METHODS: In order to optimally reconstruct a point within a volume of interest (VOI), the Radon transform on all possible planes through that point should be measured. The authors quantified the extent to which this ideal condition is satisfied for the entire image volume. They first determined a Radon completeness number for each point in the VOI, as the percentage of possible planes that is actually measured. A completeness map is then defined as a 3D matrix of the completeness numbers for the entire VOI. The authors proposed algorithms to analyze the projection datasets in Radon space and compute the completeness number for a fixed point and apply these algorithms to various architectures and scan modes that they are evaluating. In this report, the authors consider four selected candidate architectures, operating with different scan modes, for a total of five system design alternatives. Each of these alternatives is evaluated using completeness map.
RESULTS: If the detector size and cone angle are large enough to cover the entire cardiac VOI, a single-source circular scan can have ≥99% completeness over the entire VOI. However, only the central z-slice can be exactly reconstructed, which corresponds to 100% completeness. For a typical single-source architecture, if the detector is limited to an axial dimension of 40 mm, a helical scan needs about five rotations to form an exact reconstruction region covering the cardiac VOI, while a triple-source helical scan only requires two rotations, leading to a 2.5x improvement in temporal resolution. If the source and detector of an inverse-geometry (IGCT) system have the same axial extent, and the spacing of source points in the axial and transaxial directions is sufficiently small, the IGCT can also form an exact reconstruction region for the cardiac VOI. If the VOI can be covered by the x-ray beam in any view, a composite-circling scan can generate an exact reconstruction region covering the VOI.
CONCLUSIONS: The completeness map evaluation provides useful information for selecting the next-generation cardiac CT system design. The proposed completeness map method provides a practical tool for analyzing complex scanning trajectories, where the theoretical image quality for some complex system designs is impossible to predict, without yet-undeveloped reconstruction algorithms.

Entities:  

Mesh:

Year:  2012        PMID: 22559610      PMCID: PMC3338591          DOI: 10.1118/1.3700172

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Helical cardiac cone beam reconstruction using retrospective ECG gating.

Authors:  M Grass; R Manzke; T Nielsen; P Koken; R Proksa; M Natanzon; G Shechter
Journal:  Phys Med Biol       Date:  2003-09-21       Impact factor: 3.609

2.  Interior Reconstruction Using the Truncated Hilbert Transform via Singular Value Decomposition.

Authors:  Hengyong Yu; Yangbo Ye; Ge Wang
Journal:  J Xray Sci Technol       Date:  2008-01-01       Impact factor: 1.535

3.  A prototype table-top inverse-geometry volumetric CT system.

Authors:  Taly Gilat Schmidt; Josh Star-Lack; N Robert Bennett; Samuel R Mazin; Edward G Solomon; Rebecca Fahrig; Norbert J Pelc
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

Review 4.  Approximate and exact cone-beam reconstruction with standard and non-standard spiral scanning.

Authors:  Ge Wang; Yangbo Ye; Hengyong Yu
Journal:  Phys Med Biol       Date:  2007-02-19       Impact factor: 3.609

5.  A three-dimensional-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT-helical scanning.

Authors:  Xiangyang Tang; Jiang Hsieh; Roy A Nilsen; Sandeep Dutta; Dmitry Samsonov; Akira Hagiwara
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

6.  Handling data redundancy in helical cone beam reconstruction with a cone-angle-based window function and its asymptotic approximation.

Authors:  Xiangyang Tang; Jiang Hsieh
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

7.  A scheme for multisource interior tomography.

Authors:  Ge Wang; Hengyong Yu; Yangbo Ye
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

8.  Exact interior reconstruction from truncated limited-angle projection data.

Authors:  Yangbo Ye; Hengyong Yu; Ge Wang
Journal:  Int J Biomed Imaging       Date:  2008

9.  Extending Three-Dimensional Weighted Cone Beam Filtered Backprojection (CB-FBP) Algorithm for Image Reconstruction in Volumetric CT at Low Helical Pitches.

Authors:  Xiangyang Tang; Jiang Hsieh; Roy A Nilsen; Scott M McOlash
Journal:  Int J Biomed Imaging       Date:  2006-09-05

10.  A general local reconstruction approach based on a truncated hilbert transform.

Authors:  Yangbo Ye; Hengyong Yu; Yuchuan Wei; Ge Wang
Journal:  Int J Biomed Imaging       Date:  2007
View more
  6 in total

1.  Dynamic detector offsets for field of view extension in C-arm computed tomography with application to weight-bearing imaging.

Authors:  Magdalena Herbst; Frank Schebesch; Martin Berger; Jang-Hwan Choi; Rebecca Fahrig; Joachim Hornegger; Andreas Maier
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

2.  Quantification of Tomographic Incompleteness in Cone-Beam Reconstruction.

Authors:  Rolf Clackdoyle; Frédéric Noo
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2019-05-22

3.  The feasibility of an inverse geometry CT system with stationary source arrays.

Authors:  Scott S Hsieh; Joseph A Heanue; Tobias Funk; Waldo S Hinshaw; Brian P Wilfley; Edward G Solomon; Norbert J Pelc
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

4.  Quest for the ultimate cardiac CT scanner.

Authors:  Paul FitzGerald; Peter Edic; Hewei Gao; Yannan Jin; Jiao Wang; Ge Wang; Bruno De Man
Journal:  Med Phys       Date:  2017-07-17       Impact factor: 4.071

5.  Cardiac CT: A system architecture study.

Authors:  Paul FitzGerald; James Bennett; Jeffrey Carr; Peter M Edic; Daniel Entrikin; Hewei Gao; Maria Iatrou; Yannan Jin; Baodong Liu; Ge Wang; Jiao Wang; Zhye Yin; Hengyong Yu; Kai Zeng; Bruno De Man
Journal:  J Xray Sci Technol       Date:  2016       Impact factor: 1.535

6.  Non-circular CT orbit design for elimination of metal artifacts.

Authors:  Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2020-03-16
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.