Literature DB >> 22549030

Mechanical failure after locking plate fixation of unstable intertrochanteric femur fractures.

Philipp N Streubel1, Michael J Moustoukas, William T Obremskey.   

Abstract

OBJECTIVES: The aim of this study was analyze modes of mechanical failure in a consecutive cohort of patients and establish possible risk factors.
DESIGN: This was a retrospective cohort study.
SETTING: The study was conducted at an academic level-1 trauma center. PATIENTS: Twenty-nine patients (mean age 56 years, range 21-92; 45% males, 41% smokers, 17% diabetic, mean body mass index 26.9, range 20-56) with 30 OTA 31A3 fractures treated between 2003 and 2007 were included. TREATMENT: Operative fixation using 4.5-mm locking compression plate (LCP) proximal femur plate (Synthes, Paoli, PA). MAIN OUTCOME MEASUREMENTS: Mechanical failure was defined as loss of alignment of at least 10° or shortening of at least 2 cm. Secondary outcomes included patient and fixation construct variables as possible predictors for mechanical failure.
RESULTS: At 20 months of follow-up, 11 failures (37%) occurred. Mean time to failure was 18 weeks (range 2-84). Cumulative failure rates were 10%, 20%, 27%, and 33% at 1, 2, 6, and 12 months, respectively. The most frequent failure mode was varus collapse with screw cut out (5 cases). There was no statistically significant difference between groups with regards to age, body mass index, diabetes, or smoking habit. The presence of a "kickstand screw" and medial cortical reduction were not significantly different in cases with and without failure. Proximal screw number and type were similar in both groups.
CONCLUSIONS: A high rate of mechanical failure can be expected with proximal locking plate fixation of unstable proximal femur fractures. The use of a "kickstand" screw could not be established to reduce the risk for mechanical failure. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Year:  2013        PMID: 22549030     DOI: 10.1097/BOT.0b013e318251930d

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  18 in total

1.  Proximal femoral nails compared with reverse distal femoral locking plates in intertrochanteric fractures with a compromised lateral wall; a randomised controlled trial.

Authors:  Rehan Ul Haq; Vikrant Manhas; Amite Pankaj; Amit Srivastava; Ish Kumar Dhammi; Anil Kumar Jain
Journal:  Int Orthop       Date:  2014-03-22       Impact factor: 3.075

2.  Management of unstable pertrochanteric fractures with proximal femoral locking compression plates and affect of neck-shaft angle on functional outcomes.

Authors:  Emrah Kovalak; Cenk Ermutlu; Tolga Atay; Özgür Başal
Journal:  J Clin Orthop Trauma       Date:  2017-07-15

3.  Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study.

Authors:  Pratik Nag; Souptick Chanda
Journal:  Med Biol Eng Comput       Date:  2021-01-08       Impact factor: 2.602

4.  Cephalomedullary nail versus sliding hip screw for fixation of AO 31 A1/2 intertrochanteric femoral fracture: a 12-year comparison of failure, complications, and mortality.

Authors:  Casey S Whale; D Andrew Hulet; Michael J Beebe; David L Rothberg; Chong Zhang; Angela P Presson; Ami R Stuart; Erik N Kubiak
Journal:  Curr Orthop Pract       Date:  2016 Nov-Dec

5.  Comment on "Evaluation of proximal femoral locking plate in unstable extracapsular proximal femoral fractures: Surgical technique & mid term follow up results" [J. Clin. Orthop. Trauma 5 (3) (2014) 137-145].

Authors:  Ankit Chawla; Abhipray Gahlowt
Journal:  J Clin Orthop Trauma       Date:  2015-07-03

6.  [Treatment strategy for posttraumatic complex deformity. After bilateral femoral shaft fractures].

Authors:  M Ahrend; A Ateschrang; U Stöckle; S Schröter
Journal:  Unfallchirurg       Date:  2016-02       Impact factor: 1.000

7.  Subtrochanteric Fractures: Comparison of Proximal Femur Locking Plate and Intramedullary Locking Nail Fixation Outcome.

Authors:  Ahmadreza Mirbolook; Babak Siavashi; Adel Ebrahimpour Jafarinezhad; Sina Khajeh Jahromi; Maral Farahmand; Melina Roohi Rad; Mohammad Amin Kahe
Journal:  Indian J Surg       Date:  2013-11-16       Impact factor: 0.656

8.  Complications following proximal femoral locking compression plating in unstable proximal femur fractures: medium-term follow-up.

Authors:  Sandro Hodel; Frank J P Beeres; Reto Babst; Björn-Christian Link
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-05-25

9.  Comparison of Intra- and Extramedullary Implants in Treatment of Unstable Intertrochanteric Fractures.

Authors:  Tahir Mutlu Duymus; Suavi Aydogmus; İbrahim Ulusoy; Tolga Kececi; Levent Adiyeke; Bahar Dernek; Serhat Mutlu
Journal:  J Clin Orthop Trauma       Date:  2018-04-10

10.  A biomechanical comparison of proximal femoral nails and locking proximal anatomic femoral plates in femoral fracture fixation: A study on synthetic bones.

Authors:  Korhan Ozkan; İsmail Türkmen; Adem Sahin; Yavuz Yildiz; Selim Erturk; Mehmet Salih Soylemez
Journal:  Indian J Orthop       Date:  2015 May-Jun       Impact factor: 1.251

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.