Literature DB >> 22542606

Clinical evaluation of the Oculus Keratograph.

N Best1, L Drury, J S Wolffsohn.   

Abstract

AIM: To determine the validity and reliability of the measurement of corneal curvature and non-invasive tear break-up time (NITBUT) measures using the Oculus Keratograph.
METHOD: One hundred eyes of 100 patients had their corneal curvature assessed with the Keratograph and the Nidek ARKT TonorefII. NITBUT was then measured objectively with the Keratograph with Tear Film Scan software and subjectively with the Keeler Tearscope. The Keratograph measurements of corneal curvature and NITBUT were repeated to test reliability. The ocular sensitivity disease index questionnaire was completed to quantify ocular comfort.
RESULTS: The Keratograph consistently measured significantly flatter corneal curvatures than the ARKT (MSE difference: +1.83 ± 0.44D), but was repeatable (p > 0.05). Keratograph NITBUT measurements were significantly lower than observation using the Tearscope (by 12.35 ± 7.45 s; p < 0.001) and decreased on subsequent measurement (by -1.64 ± 6.03 s; p < 0.01). The Keratograph measures the first time the tears break up anywhere on the cornea with 63% of subjects having NITBUTs <5 s and a further 22% having readings between 5 and 10 s. The Tearscope results were found to correlate better with the patients symptoms (r = -0.32) compared to the Keratograph (r = -0.19).
CONCLUSIONS: The Keratograph requires a calibration off-set to be comparable to other keratometry devices. Its current software detects very early tear film changes, recording significantly lower NITBUT values than conventional subjective assessment. Adjustments to instrumentation software have the potential to enhance the value of Keratograph objective measures in clinical practice.
Copyright © 2012 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22542606     DOI: 10.1016/j.clae.2012.04.002

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  33 in total

1.  A Non-invasive Way to Isolate and Phenotype Cells from the Conjunctiva.

Authors:  Tanima Bose; Aihua Hou; Ryan Lee; Louis Tong; K George Chandy
Journal:  J Vis Exp       Date:  2017-07-05       Impact factor: 1.355

2.  Agreement between Automated and Traditional Measures of Tear Film Breakup.

Authors:  Stephanie M Cox; Kelly K Nichols; Jason J Nichols
Journal:  Optom Vis Sci       Date:  2015-09       Impact factor: 1.973

3.  Repeatability of Non-invasive Keratograph Break-Up Time measurements obtained using Oculus Keratograph 5M.

Authors:  José Vicente García-Marqués; Noelia Martínez-Albert; Cristian Talens-Estarelles; Santiago García-Lázaro; Alejandro Cerviño
Journal:  Int Ophthalmol       Date:  2021-03-16       Impact factor: 2.031

4.  Agreement and repeatability of objective systems for assessment of the tear film.

Authors:  Joaquín Fernández; Manuel Rodríguez-Vallejo; Javier Martínez; Ana Tauste; Javier García-Montesinos; David P Piñero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-04-18       Impact factor: 3.117

5.  Subbasal nerve morphology, corneal sensation, and tear film evaluation after refractive femtosecond laser lenticule extraction.

Authors:  Anders H Vestergaard; Keea T Grønbech; Jakob Grauslund; Anders R Ivarsen; Jesper Ø Hjortdal
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-06-22       Impact factor: 3.117

6.  A novel noninvasive ocular surface analyzer for the assessment of dry eye with Meibomian gland dysfunction.

Authors:  Yuanyuan Qi; Chen Zhang; Shaozhen Zhao; Yue Huang; Ruibo Yang
Journal:  Exp Ther Med       Date:  2017-04-19       Impact factor: 2.447

7.  Relationships among Tear Film Stability, Osmolarity, and Dryness Symptoms.

Authors:  Thao N Yeh; Andrew D Graham; Meng C Lin
Journal:  Optom Vis Sci       Date:  2015-09       Impact factor: 1.973

8.  Reliability of Chinese web-based ocular surface disease index questionnaire in dry eye patients: a randomized, crossover study.

Authors:  Xin-Mei Zhang; Lan-Ting Yang; Qing Zhang; Qing-Xia Fan; Can Zhang; Yue You; Chen-Guang Zhang; Tie-Zhu Lin; Ling Xu; Salissou Moutari; Jonathan E Moore; Emmanuel E Pazo; Wei He
Journal:  Int J Ophthalmol       Date:  2021-06-18       Impact factor: 1.779

9.  Non-invasive assessment of tear film stability with a novel corneal topographer in Indian subjects.

Authors:  Vipul Bhandari; Jagdeesh K Reddy; Kirti Relekar; Ameya Ingawale; Neeraj Shah
Journal:  Int Ophthalmol       Date:  2016-02-18       Impact factor: 2.031

10.  Spatial Distribution of Noninvasive Break Up Times and Clinical Relevance in Healthy Participants and Mild Dry Eye.

Authors:  Louis Tong; Calesta Hui Yi Teo; Ryan Khee Jin Lee
Journal:  Transl Vis Sci Technol       Date:  2019-10-17       Impact factor: 3.283

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.