Literature DB >> 22538138

Glucose measurement of intensive care unit patient plasma samples using a fixed-wavelength mid-infrared spectroscopy system.

James Krinsley1, Kelly Bochicchio, Christopher Calentine, Grant Bochicchio.   

Abstract

OBJECTIVE: Glycemic control is a rapidly developing field in intensive care medicine with the aim of reducing mortality, morbidity, and cost. Current intensive care unit (ICU) glucose measurement technologies are susceptible to interference from medications, volume expanders, and other substances present in critically ill patients. We hypothesized that a fixed-wavelength mid-infrared (mid-IR) spectroscopy system would be accurate for measuring glucose levels of ICU patients. RESEARCH DESIGN AND METHODS: This is a prospective investigation of plasma samples from two different institutions treating a heterogeneous population of ICU patients. The first 292 samples were collected from 86 patients admitted to Stamford Hospital, and the next 352 samples were collected from 75 patients from three ICUs at the University of Maryland. Plasma samples were measured on a Fourier-transform infrared or a proprietary spectrometer, with a glucose prediction algorithm to correct for spectral interference, which were compared with reference measurements taken using a YSI 2300 glucose analyzer.
RESULTS: Glucose values ranged from 24 to 343 mg/dl. Numerous medications and injury/disease states were observed in the patient populations, with metoprolol, fentanyl, and multiple organ failure the most prevalent. Despite these interferents, there was a high correlation (r ≥ 0.94) and low standard error (≤12.8 mg/dl) between the predicted glucose values and those of the YSI 2300 STAT Plus reference instrument in the three studies. A total of 95.1% of the 644 values in the three studies met International Organization for Standardization 15197 criteria.
CONCLUSION: These results suggest that a fixed-wavelength mid-IR spectrometer can measure glucose accurately in the plasma of ICU patients.
© 2012 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22538138      PMCID: PMC3380770          DOI: 10.1177/193229681200600212

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

Review 1.  [Glycaemia control in critically ill patients is justified and effective].

Authors:  J Rybka
Journal:  Vnitr Lek       Date:  2010-09

2.  Diverse influences on blood glucose measurements in the ICU setting.

Authors:  Tanya M Binette; George S Cembrowski
Journal:  Chest       Date:  2005-10       Impact factor: 9.410

3.  Prediction of glucose in whole blood by near-infrared spectroscopy: influence of wavelength region, preprocessing, and hemoglobin concentration.

Authors:  Yoen-Joo Kim; Gilwon Yoon
Journal:  J Biomed Opt       Date:  2006 Jul-Aug       Impact factor: 3.170

4.  Glucose and other constituents of blood determined by ATR-FTIR-spectroscopy.

Authors:  J D Kruse-Jarres; G Janatsch; U Gless; R Marbach; H M Heise
Journal:  Clin Chem       Date:  1990-02       Impact factor: 8.327

5.  Critical glucose control: the devil is in the details.

Authors:  Brenda G Fahy; Douglas B Coursin
Journal:  Mayo Clin Proc       Date:  2008-04       Impact factor: 7.616

Review 6.  Glycemic control in the ICU.

Authors:  Moritoki Egi; Simon Finfer; Rinaldo Bellomo
Journal:  Chest       Date:  2011-07       Impact factor: 9.410

7.  Plasma-generating glucose monitor accuracy demonstrated in an animal model.

Authors:  Peggy Magarian; Bernhard Sterling
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 8.  The impact of premorbid diabetic status on the relationship between the three domains of glycemic control and mortality in critically ill patients.

Authors:  James S Krinsley; Geert Meyfroidt; Greet van den Berghe; Mori Egi; Rinaldo Bellomo
Journal:  Curr Opin Clin Nutr Metab Care       Date:  2012-03       Impact factor: 4.294

9.  Automated near-continuous glucose monitoring measured in plasma using mid-infrared spectroscopy.

Authors:  Thomas Jax; Tim Heise; Leszek Nosek; Jennifer Gable; Gene Lim; Christopher Calentine
Journal:  J Diabetes Sci Technol       Date:  2011-03-01

10.  Computerized intensive insulin dosing can mitigate hypoglycemia and achieve tight glycemic control when glucose measurement is performed frequently and on time.

Authors:  Rattan Juneja; Corbin P Roudebush; Stanley A Nasraway; Adam A Golas; Judith Jacobi; Joni Carroll; Deborah Nelson; Victor J Abad; Samuel J Flanders
Journal:  Crit Care       Date:  2009-10-12       Impact factor: 9.097

View more
  4 in total

1.  Glucose Management Technologies for the Critically Ill.

Authors:  Pedro D Salinas; Carlos E Mendez
Journal:  J Diabetes Sci Technol       Date:  2019-01-13

Review 2.  Vascular Glucose Sensor Symposium: Continuous Glucose Monitoring Systems (CGMS) for Hospitalized and Ambulatory Patients at Risk for Hyperglycemia, Hypoglycemia, and Glycemic Variability.

Authors:  Jeffrey I Joseph; Marc C Torjman; Paul J Strasma
Journal:  J Diabetes Sci Technol       Date:  2015-06-15

3.  [Glucose control in the critically ill. Innovations and contemporary strategies].

Authors:  U Holzinger
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-05-19       Impact factor: 0.840

4.  Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically ill patients.

Authors:  Roosmarijn Tm van Hooijdonk; Tineke Winters; Johan C Fischer; Edmée C van Dongen-Lases; James S Krinsley; Jean-Charles Preiser; Marcus J Schultz
Journal:  Ann Intensive Care       Date:  2014-03-06       Impact factor: 6.925

  4 in total