Vincent Auboiroux1, Lorena Petrusca, Magalie Viallon, Thomas Goget, Christoph D Becker, Rares Salomir.
Abstract
Magnetic resonance imaging (MRI) and ultrasonography have been used simultaneously in this ex vivo study for the image-guidance of high intensity focused ultrasound (HIFU) treatment in moving tissue. A ventilator-driven balloon produced periodic and non-rigid (i.e. breathing-like) motion patterns in phantoms. MR-compatible ultrasound (US) imaging enabled near real-time 2D motion tracking based on optical flow detection, while near-harmonic reference-free proton resonance frequency shift (PRFS) MR thermometry (MRT) was used to monitor the thermal buildup on line. Reference-free MRT was applied to gradient-echo echo-planar imaging phase maps acquired at the frame rate of 250 to 300 ms/slice with voxel size 1.25×1.25×5 mm(3). The MR-US simultaneous imaging was completely free of mutual interferences while minor RF interferences from the HIFU device were detected in the far field of the US images. The effective duty-cycle of the HIFU sonication was close to 100 % and no off-interval was required to temporally decouple it from the ultrasonography. The motion compensation of the HIFU sonication was achieved with an 8 Hz frame rate and sub-millimeter spatial accuracy, both for single-focus mode and for an iterated multi-foci line scan. Near harmonic reference-less PRFS MRT delivered motion-robust thermal maps perpendicular or parallel to the HIFU beam (0.7 °C precision, 0.5 °C absolute accuracy). Out-of-plane motion compensation was not addressed in this study.Mesh:
Year: 2012 PMID: 22517112 DOI: 10.1088/0031-9155/57/10/N159
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609