Literature DB >> 22513105

Defining a magnetic resonance scan sequence for permanent seed prostate brachytherapy postimplant assessment.

David Bowes1, Juanita M Crook, Rasika Rajapakshe, Cynthia Araujo, Brent Parker.   

Abstract

PURPOSE: We describe a magnetic resonance (MR) scan sequence for prostate brachytherapy postimplant assessment. METHODS AND MATERIALS: One brachytherapy team at the British Columbia Cancer Agency has incorporated MR-CT fusion into their permanent seed prostate brachytherapy quality assurance procedure. Several attempts were required to ensure that the diagnostic MR scanner at the adjoining general hospital performed the desired sequence, providing many examples of suboptimal scans and underlining the pitfalls for a center trying to incorporate the use of MR scanning into their brachytherapy program.
RESULTS: The recommended sequence (Fast Spin Echo T2-weighted, repetition time [TR]/echo time [TE] 4500/90, echo train length [ETL] 10, 20×20 field of view [FOV], 80 bandwidth [BW]) is associated with superior edge detection when compared with those images in which a typical diagnostic sequence was used. The use of a low bandwidth sequence does not compromise edge detection or seed identification when compared with a higher bandwidth.
CONCLUSIONS: We have defined a magnetic resonance imaging sequence, which appears to optimize both prostate delineation and identification of seeds, lending itself to straightforward fusion with CT images and allowing for less uncertainty in permanent seed prostate brachytherapy quality assurance.
Copyright © 2013 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22513105     DOI: 10.1016/j.brachy.2012.03.001

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  4 in total

1.  Effect of pulse sequence parameter selection on signal strength in positive-contrast MRI markers for MRI-based prostate postimplant assessment.

Authors:  Tze Yee Lim; Rajat J Kudchadker; Jihong Wang; R Jason Stafford; Christopher MacLellan; Arvind Rao; Geoffrey S Ibbott; Steven J Frank
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

2.  Radiotherapy planning using MRI.

Authors:  Maria A Schmidt; Geoffrey S Payne
Journal:  Phys Med Biol       Date:  2015-10-28       Impact factor: 3.609

3.  Hemiablative Focal Low Dose Rate Brachytherapy: A Phase II Trial Protocol.

Authors:  Ana Fernandez Ots; Joseph Bucci; Yaw Sinn Chin; David Malouf; Andrew Howie; Komiti Ese Enari
Journal:  JMIR Res Protoc       Date:  2016-06-13

4.  Development of a magnetic resonance imaging protocol to visualize encapsulated contrast agent markers in prostate brachytherapy recipients: initial patient experience.

Authors:  Tze Yee Lim; Rajat J Kudchadker; Jihong Wang; Tharakeswara Bathala; Janio Szklaruk; Thomas J Pugh; Usama Mahmood; Geoffrey S Ibbott; Steven J Frank
Journal:  J Contemp Brachytherapy       Date:  2016-06-13
  4 in total