Manifestation and post hoc correction of gradient cross-term artifacts in DTI.

Govind Nair1, Xiaoping P Hu.   

Abstract

Cross-terms between imaging and diffusion gradients, unaccounted for during tensor calculations, can lead to erroneous estimation of diffusivity and fractional anisotropy (FA) in regions of isotropic and anisotropic diffusion. Cross-term of magnitude 136.8±1.6 s/mm(2), artificially introduced in the slice-encode direction, caused an increase in FA in isotropic phantom from 0.0546±0.0001 to 0.0996±0.0001, while the change in chimpanzee brain depended on the orientation of the white matter (WM). Mean diffusivity (MD) remained unchanged in isotropic phantom, but increased by ∼20% in the WM due to cross-terms. A bias was observed in the principal eigenvectors in both phantom and chimpanzee brain, resulting in significant increase in midline crossing fibers along the bias than perpendicular to it in tractography in chimpanzee brain. Post hoc correction of these artifacts was achieved by estimating the cross-term factors using calibration scans on an isotropic phantom and modifying the b-matrix before tensor calculation. Upon correction, the FA and MD values closely resembled the values obtained from sequence without cross-terms, and the bias in principal eigenvectors was eliminated. Customized sequences involving large b-values, high-resolution imaging, or long diffusion or echo times should therefore be evaluated and any residual cross-terms corrected before implementation.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22495240      PMCID: PMC3367036          DOI: 10.1016/j.mri.2012.02.021

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  16 in total

1.  A study of rotationally invariant and symmetric indices of diffusion anisotropy.

Authors:  N G Papadakis; D Xing; G C Houston; J M Smith; M I Smith; M F James; A A Parsons; C L Huang; L D Hall; T A Carpenter
Journal:  Magn Reson Imaging       Date:  1999-07       Impact factor: 2.546

2.  Optimal strategies for measuring diffusion in anisotropic systems by magnetic resonance imaging.

Authors:  D K Jones; M A Horsfield; A Simmons
Journal:  Magn Reson Med       Date:  1999-09       Impact factor: 4.668

3.  DTI at long diffusion time improves fiber tracking.

Authors:  Swati Rane; Govind Nair; Timothy Q Duong
Journal:  NMR Biomed       Date:  2010-06       Impact factor: 4.044

4.  A simple method for obtaining cross-term-free images for diffusion anisotropy studies in NMR microimaging.

Authors:  M Neeman; J P Freyer; L O Sillerud
Journal:  Magn Reson Med       Date:  1991-09       Impact factor: 4.668

5.  Double-spin-echo diffusion weighting with a modified eddy current adjustment.

Authors:  Jürgen Finsterbusch
Journal:  Magn Reson Imaging       Date:  2010-01-13       Impact factor: 2.546

Review 6.  Diffusion MRI: precision, accuracy and flow effects.

Authors:  T E Conturo; R C McKinstry; J A Aronovitz; J J Neil
Journal:  NMR Biomed       Date:  1995 Nov-Dec       Impact factor: 4.044

7.  A simplified method to measure the diffusion tensor from seven MR images.

Authors:  P J Basser; C Pierpaoli
Journal:  Magn Reson Med       Date:  1998-06       Impact factor: 4.668

8.  The b matrix in diffusion tensor echo-planar imaging.

Authors:  J Mattiello; P J Basser; D Le Bihan
Journal:  Magn Reson Med       Date:  1997-02       Impact factor: 4.668

9.  Analytical expressions for the NMR apparent diffusion coefficients in an anisotropic system and a simplified method for determining fiber orientation.

Authors:  E W Hsu; S Mori
Journal:  Magn Reson Med       Date:  1995-08       Impact factor: 4.668

10.  Estimation of the effective self-diffusion tensor from the NMR spin echo.

Authors:  P J Basser; J Mattiello; D LeBihan
Journal:  J Magn Reson B       Date:  1994-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.