| Literature DB >> 22472530 |
Xi Chen1, Jin Chen1, Fubin Wang2, Xia Xiang1, Ming Luo1, Xinghu Ji1, Zhike He3.
Abstract
In this work, we first employ a drying method combining with the bienzyme colorimetric detection of glucose and uric acid on microfluidic paper-based analysis devices (μPADs). The channels of 3D μPADs are also designed by us to get better results. The color results are recorded by both Gel Documentation systems and a common camera. By using Gel Documentation systems, the limits of detection (LOD) of glucose and uric acid are 3.81 × 10(-5)M and 4.31 × 10(-5)M, respectively one order of magnitude lower than that of the reported methods on μPADs. By using a common camera, the limits of detection (LOD) of glucose and uric acid are 2.13 × 10(-4)M and 2.87 × 10(-4)M, respectively. Furthermore, the effects of detection conditions have been investigated and discussed comprehensively. Human serum samples are detected with satisfactory results, which are comparable with the clinical testing results. A low-cost, simple and rapid colorimetric method for the simultaneous detection of glucose and uric acid on the μPADs has been developed with enhanced sensitivity.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22472530 DOI: 10.1016/j.bios.2012.03.018
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618