BACKGROUND: The dominant frequency of the Fourier power spectrum is useful to analyze complex fractionated atrial electrograms (CFAE), but spectral resolution is limited and uniform from DC to the Nyquist frequency. Herein the spectral resolution of a recently described and relatively new spectral estimation technique is compared to the Fourier radix-2 implementation. METHODS: In 10 paroxysmal and 10 persistent atrial fibrillation patients, 216 CFAE were acquired from the pulmonary vein ostia and left atrial free wall (977 Hz sampling rate, 8192 sample points, 8.4 s duration). With these parameter values, in the physiologic range of 3-10 Hz, two frequency components can theoretically be resolved at 0.24 Hz using Fourier analysis and at 0.10 Hz on average using the new technique. For testing, two closely-spaced periodic components were synthesized from two different CFAE recordings, and combined with two other CFAE recordings magnified 2×, that served as interference signals. The ability to resolve synthesized frequency components in the range 3-4 Hz, 4-5 Hz, …, 9-10 Hz was determined for 15 trials each (105 total). RESULTS: With the added interference, frequency resolution averaged 0.29 ± 0.22 Hz for Fourier versus 0.16 ± 0.10 Hz for the new method (p < 0.001). The misalignment error of spectral peaks versus actual values was ±0.023 Hz for Fourier and ±0.009 Hz for the new method (p < 0.001). One or both synthesized peaks were lost in the noise floor 13/105 times using Fourier versus 4/105 times using the new method. CONCLUSIONS: Within the physiologically relevant frequency range for characterization of CFAE, the new method has approximately twice the spectral resolution of Fourier analysis, there is less error in estimating frequencies, and peaks appear more readily above the noise floor. Theoretically, when interference is not present, to resolve frequency components separated by 0.10 Hz using Fourier analysis would require an 18.2 s sequence duration, versus 8.4 s with the new method.
BACKGROUND: The dominant frequency of the Fourier power spectrum is useful to analyze complex fractionated atrial electrograms (CFAE), but spectral resolution is limited and uniform from DC to the Nyquist frequency. Herein the spectral resolution of a recently described and relatively new spectral estimation technique is compared to the Fourier radix-2 implementation. METHODS: In 10 paroxysmal and 10 persistent atrial fibrillationpatients, 216 CFAE were acquired from the pulmonary vein ostia and left atrial free wall (977 Hz sampling rate, 8192 sample points, 8.4 s duration). With these parameter values, in the physiologic range of 3-10 Hz, two frequency components can theoretically be resolved at 0.24 Hz using Fourier analysis and at 0.10 Hz on average using the new technique. For testing, two closely-spaced periodic components were synthesized from two different CFAE recordings, and combined with two other CFAE recordings magnified 2×, that served as interference signals. The ability to resolve synthesized frequency components in the range 3-4 Hz, 4-5 Hz, …, 9-10 Hz was determined for 15 trials each (105 total). RESULTS: With the added interference, frequency resolution averaged 0.29 ± 0.22 Hz for Fourier versus 0.16 ± 0.10 Hz for the new method (p < 0.001). The misalignment error of spectral peaks versus actual values was ±0.023 Hz for Fourier and ±0.009 Hz for the new method (p < 0.001). One or both synthesized peaks were lost in the noise floor 13/105 times using Fourier versus 4/105 times using the new method. CONCLUSIONS: Within the physiologically relevant frequency range for characterization of CFAE, the new method has approximately twice the spectral resolution of Fourier analysis, there is less error in estimating frequencies, and peaks appear more readily above the noise floor. Theoretically, when interference is not present, to resolve frequency components separated by 0.10 Hz using Fourier analysis would require an 18.2 s sequence duration, versus 8.4 s with the new method.
Accurate spectral resolution is important for analysis of atrial fibrillation (AF) in
the frequency domain [1]. The information can
be used, for example, to monitor periods of AF versus sinus rhythm in patients with
paroxysmal arrhythmia. For understanding the mechanism by which the arrhythmia is
maintained, it is also desirable to quantify any temporal variation in frequency
that can occur during AF [2]. Since frequency
shifts can be minute, sufficient resolving power is essential to detect these
differences over short time intervals. Yet this becomes more difficult when
signal-to-noise ratio is low, as is often the case. Another impediment to the
generation of an accurate spectral profile is the presence of transients in the
signal that are unrelated to physiologically relevant phenomena. One way to improve
continuity in the frequency domain representation of AF from one instance of time to
the next is to model and update the spectral profile based on new signal information
as it is acquired in atrial fibrillation signals [2,3]. By using Gaussian functions to
model the spectral peaks, after leveling the spectral baseline, spurious or
transient changes to the signal frequency content can be excluded from the updates.
When AF organization is reduced, the harmonic pattern is also expected to diminish
[2], while independent spectral peaks
caused for example by wavebreak can increase [4]. When resolving power is insufficient, a broad spectral peak
in the Fourier power spectrum may thus be indicative of either a temporal variation
in the frequency of a single component, or the merging of multi-component
independent sources [5]. Therefore, the need
for sufficient spectral resolution is essential for AF analysis. Improved resolution
could also be useful to study the evolution and spontaneous termination of
paroxysmal AF, as well as to estimate the organization of atrial activity
[6].In recent work we introduced a paradigm for spectral estimation and signal
transformation of complex fractionated atrial electrograms (CFAE) based upon
ensemble averaging [7-9]. This relatively new technique, like Fourier
analysis, depends upon the autocorrelation function for generation of a power
spectrum [9]. For spectral estimation,
Fourier analysis models the sinusoidal properties of the autocorrelation function,
while the new technique averages the autocorrelation function at lags w, 2w, 3w,
…, where w is the period [9]. The
spectral resolution of Fourier analysis is constant across bandwidth, while the
spectral resolution of the ensemble average method depends inversely on w. The 1/w
relationship results in greater spectral resolution at lower frequencies, which
includes the electrophysiologic range of interest
of ~ 3–10 Hz that is used for AF characterization. Herein,
the new technique is compared to Fourier analysis for discerning closely-spaced
frequency components in CFAE, which are of interest to characterize the arrhythmia
and to identify dominant frequencies.
Method
Clinical data acquisition
Atrial electrograms were recorded in a series of 20 patients referred to the
Columbia University Medical Center cardiac electrophysiology (EP) laboratory for
catheter ablation of AF. These recordings were obtained prospectively as
approved by the Internal Review Board at Columbia University Medical Center, but
were analyzed retrospectively after the catheter ablation procedures were
completed using standard clinical protocols. Ten patients had documented
clinical paroxysmal AF, and all 10 had normal sinus rhythm as their baseline
rhythm in the EP laboratory. AF was induced by burst atrial pacing from the
coronary sinus or right atrial lateral wall, and persisted for at least 10
minutes for those signals included in the retrospective analysis of this study.
Ten other patients had longstanding persistent AF, and had been in AF without
interruption for 1–3 years prior to the catheter mapping and
ablation procedure. The surface electro gram signals were acquired in analog
form using the GE CardioLab system (GE Healthcare, Waukesha, WI) and filtered
from 30-500 Hz with a single-pole band pass filter to remove baseline
drift and high frequency noise. The filtered signals were digitally sampled by
the system at 0.977 KHz and stored. Although the band pass high end was slightly
above the Nyquist frequency, negligible signal energy is expected to reside in
this frequency range [8].Only signals identified as CFAEs by two cardiac electro physiologists were
included in the retrospective analysis. Candidate CFAE recordings of at least 10
seconds in duration were obtained from two sites outside the ostia of each of
the four pulmonary veins (PV). Similar recordings were obtained at two sites on
the endocardial surface of the left atrial free wall, one in the mid-posterior
wall, and another on the anterior ridge at the base of the left atrial
appendage. From each of these recordings, 8.4-second sequences (8192 sample
points) were analyzed. A total of 240 such sequences were acquired during
electrophysiologic analysis – 120 from paroxysmal and 120 from
longstanding AFpatients. Subsequently, only 216 of the recordings were
determined to be CFAE, and only these were used for subsequent analysis. As in
previous studies, all CFAE signals were normalized to mean zero and unity
variance prior to further processing [9].
Frequency resolution
Using the radix-2 implementation for Fourier spectral estimation, the frequency
resolution is:This resolution is uniform throughout the frequency range. Based upon the Nyquist
theorem, the range extends from DC to a maximum of ½ the sample rate.
Atrial fibrillation signals are typically acquired at 977 Hz digitization,
as was done in this study. For evaluation of the frequency content of CFAE, time
durations of approximately 8 seconds are desirable for analysis [10]. Using these parameter values, the
resolution would be:This frequency resolution is uniform throughout the range of DC to
977/2 = 489 Hz. The total number of data points in the
resulting power spectrum will be 8192/2 = 4096. Thus the frequency
resolution above the electrophysiologic range of interest, 10 Hz, to the
high limit of 489 Hz, is the same as the frequency resolution within the
physiologic range of 3–10 Hz.A recently described technique that utilizes signal averaging was also used for
spectral estimation [7-9]. Briefly, an ensemble average vector
ew with dimension w × 1 is obtained by
averaging the n successive mean zero segments of a signal x having
dimension N × 1, with each segment being of length w:Where an underline denotes a vector, bold font indicates a matrix, Uw
is the summing matrix with dimension w × N, and Iw
are w × w identity submatrices used to extract the signal
segments from x. The number of signal segments in the summation is:where int is the integer function. If N/w is not an integer, then in forming
Uw, 0’s are added to the N – (n · w) columns at
the matrices’ right edge [9]. The
segment length w can be converted to a frequency:The power in the ensemble average is given by:To construct the power spectrum, the root mean square (RMS) power can be used
[7-9]:which has units of millivolts. The power spectrum can be displayed by plotting
√n·PwRMS versus frequency f. The √n term levels
the spectral baseline, which would otherwise decrease by 1/√n, the noise
falloff per number of summations n used for ensemble averaging. Computer code to
calculate this spectrum has been presented and described elsewhere
[9].From Eq. 5, it is apparent that the frequency resolution of the new spectral
estimation technique is proportional to 1/w. This resolution can be calculated
as:where RN is the resolution of the spectral estimator, SR is the sample
rate, and fw and fw+1 are any two adjacent points on its
frequency spectrum. The frequency resolution will be higher (i.e., the
difference fw − fw+1 will be smaller) at
points along the spectrum where period w is longer (i.e. at low
frequencies).Akin to zero padding in Fourier analysis, it would be possible to interpolate
between discrete sample points and use real values of w to boost the number of
spectral points generated by the new method, but like padding for Fourier
analysis, this would not add information to the system. The integer values of w
range from 2 to N/2, where the upper bound of N/2 is given by the requirement
that at least two signal segments are needed to compute an ensemble average. The
value at w = 1 is the DC level. Thus analogous to the Fourier power
spectrum, the power spectrum using the new technique also has 4096 points of
resolution. Yet, unlike Fourier analysis, this resolution is not uniform. Based
upon the 1/x relationship between frequency f and period w, the resolution will
be poor at high frequencies but excellent at low frequencies. This property can
be exploited for analysis of CFAE, since much of the frequency content of
interest is in the range 3–10 Hz [7,8]. In this range the average
frequency resolution of the new technique can be estimated from the resolution
at representative frequency values. For example the resolution at 3.5 Hz
is given by:Similarly, the resolution at 4.5 Hz is:The average resolution as determined from the resolution at 3.5, 4.5,
5.5,…, 9.5 Hz is 0.05 Hz. However to resolve two peaks, there
must be at least one point of power spectral resolution separating them. Thus
the minimum frequency at which two spectral peaks can be resolved would be
2 × 0.12 Hz = 0.24 Hz for Fourier, and
on the average 2 × 0.05 Hz = 0.10 Hz
for the new method in the range 3–10 Hz (see second column, Table
1). As a note of comparison, to theoretically resolve
frequency components separated by 0.10 Hz using Fourier analysis would
therefore require an 18.2 s sequence duration, versus 8.4 s with the
new method.
Table 1
Resolution to discern closely-spaced frequency components
Freq. (Hz)
Res-theor (Hz)
Res-NT (Hz)
Res-FT (Hz)
Significance
9.37 ± 0.47
0.19
0.21 ± 0.05
0.25 ± 0.08
P = 0.307
8.49 ± 0.24
0.15
0.23 ± 0.09
0.28 ± 0.08
P = 0.105
7.51 ± 0.32
0.11
0.18 ± 0.07
0.29 ± 0.18
P = 0.081
6.47 ± 0.27
0.09
0.14 ± 0.04
0.32 ± 0.20
P = 0.003
5.45 ± 0.32
0.06
0.13 ± 0.09
0.27 ± 0.18
P < 0.001
4.37 ± 0.27
0.04
0.08 ± 0.03
0.22 ± 0.07
P < 0.001
3.51 ± 0.29
0.03
0.07 ± 0.04
0.19 ± 0.06
P < 0.001
6.41 ± 2.01 (MN)
0.10
0.16 ± 0.10 (MN)
0.29 ± 0.22 (MN)
P < 0.001
Freq.: the mean frequency used for the test, Res-theor – the
theoretical resolving power of the new method, Res – NT: the
resolving power of the new method, Res – FT: the resolving
power of Fourier analysis, Significance – the significance of
the difference between the resolving power of the techniques based
on the unpaired t-test. MN – mean value.
Resolution to discern closely-spaced frequency componentsFreq.: the mean frequency used for the test, Res-theor – the
theoretical resolving power of the new method, Res – NT: the
resolving power of the new method, Res – FT: the resolving
power of Fourier analysis, Significance – the significance of
the difference between the resolving power of the techniques based
on the unpaired t-test. MN – mean value.
Synthesis of simulated CFAE and power spectra
CFAE with closely-spaced frequency components were artificially synthesized to
test the resolution of the methods as follows. First a sequence of random length
ω was extracted from one of the 216 CFAE recordings, which was also
selected at random. The extracted sequence of length ω was repeated to
N = 8192 sample points to form one signal component. Then another
random sequence, with length ω + 2, was extracted from another
randomly selected CFAE recording. The extracted sequence of length
ω + 2 was also repeated to N = 8192 sample points
to form a second signal component. Two other of the 216 CFAE recordings selected
at random were each amplified with a gain of ×2 to be used as interference
as in a previous study [8]. Thus the
following four randomly obtained sequences of length 8192 sample points were
summed and used for subsequent analysis:1. signal component 1 with repeating period w2. signal component 2 with repeating period w + 23. interference 1 with 2× gain4. interference 2 with 2× gainAn example of the synthesis process is shown in Figure 1
for frequencies toward the low end of the spectral range. Repeating segments
with ω = 180 sample points and
ω + 2 = 182 sample points are shown in panels A and
B, respectively (f = 5.43 Hz and 5.37 Hz). The sum of
two additive interferences with a gain of 2× is shown in panel C. The
result from combining the components of panels A, B, and C is shown in panel D.
The original periodicities are mostly unrecognizable in panel D. Similarly, an
example for components with frequencies toward the high end of the
electrophysiologic range of interest is shown in Figure 2.
Repeating segments with ω = 113 sample points and
ω + 2 = 115 sample points are shown in panels A and
B, respectively (f = 8.65 Hz and 8.50 Hz). Additive
interferences are shown in panel C. The result from combining the components of
panels A, B, and C is shown in panel D. The original periodicities are again
mostly unrecognizable.
Figure 1
Example of synthesized signals at the low range in
frequency.A-B. Synthesized frequency components.
C. Interferences. D. Combination of synthesized frequency
components and interferences.
Figure 2
Example of synthesized signals at the high range in
frequency.A-B. Synthesized frequency components.
C. Interferences. D. Combination of synthesized frequency
components and interferences.
Example of synthesized signals at the low range in
frequency.A-B. Synthesized frequency components.
C. Interferences. D. Combination of synthesized frequency
components and interferences.Example of synthesized signals at the high range in
frequency.A-B. Synthesized frequency components.
C. Interferences. D. Combination of synthesized frequency
components and interferences.This paradigm was repeated so that a total of 15 simulated CFAE with
closely-spaced frequency components in the range 3–4 Hz, 15 in the
range 4–5 Hz, …, and 15 in the range 9–10 Hz were
constructed. Thus the total number of simulation signals to be analyzed, of the
type shown in Figure 1D and 2D, was
7 × 15 = 105.Prior to Fourier analysis, each simulated signal was zero-padded to a length
N = 65536, for a final interval of 0.015 Hz between spectral
points. By zero padding, the frequency resolution of the FFT was not increased,
only the point-to-point interval in the resulting spectrum, because the length
of the observed signal was unchanged. Zero padding ensured that closely-spaced
spectral components would be evident within the limit given by the frequency
resolution constraint. The Fourier power spectrum was generated using a radix-2
implementation [11]. Initially the
simulated CFAE were preprocessed with a Han window prior to Fourier analysis.
However, while Han windowing can improve periodicity by forcing the signal ends
to zero, it also adds distortion in the form of amplitude modulation. The
amplitude modulation can impart spectral sidebands which reduce frequency
resolution. It was also observed visually that discernment of closely-spaced
frequency components was improved without this filter. Therefore a rectangular
window (i.e. no filter) was used for Fourier analysis. Likewise, no window
preprocessing was done prior to spectral estimation with the new method.Consider the spectral profiles of both Fourier and new technique, each of which
consists of 4096 data points for parameter values of 977 Hz sample rate
and a signal length of 8192 sample points. As shown in Figures 1 and 2, the initially synthesized signals had
frequency components of ω and ω + γ, where
γ = 2, the minimum resolving distance. To determine the ability
of each method to resolve closely-spaced peaks, γ was increased by 1 in
each of the simulated signals until two distinct peaks appeared in the spectrum.
Closely-spaced synthesized frequency components were considered to be resolved
for the minimum value of γ for which any resulting spectral peaks with
periods w and w + α met the following four criteria. The first
requirement for discernability was that:where s is the amplitude of the tallest spectral peak in proximity to the actual
location of a synthesized frequency component, referenced to that foot which
gives the largest peak amplitude, s1 is the amplitude of the higher
peak, and s2 is the amplitude of the lower peak. An example of peak
amplitude measurement is shown in Figure 3B (solid
vertical arrows). The second requirement for discernability is that the minimum
value between the two peaks:
Figure 3
Examples of frequency spectra used for analysis for CFAE with two
closely-spaced frequency components at the low end of range. A.
The power spectrum using the new technique in the range
1–12 Hz. B. A close-up of the power spectrum with a
range of 2 Hz. The parameters used to determine whether the
synthetic frequency components are resolved are shown. C. The
Fourier power spectrum in the range 1–12 Hz. D. A
close-up of the Fourier power spectrum with a range of 2 Hz. The
synthetic frequency components are not resolved in the Fourier
spectrum.
is within or below the maximum range b of the neighboring background level
(Figure 3B). The neighboring background level is defined
as the range in magnitude −0.25 Hz in frequency away from the left
foot of peak w + α, and +0.25 Hz in frequency away from
the right foot of peak w (horizontal lines, Figure 3B).
The third criterion is that:that is, the magnitudes sw and sw+α are greater than
the maximum range b of the neighboring background level (Figure 3B). The final criterion is that the peaks at w and
w + α are each within ±0.15 Hz of the two
synthesized frequency components having actual periods of ω and
ω + γ (noted with bars at the top of the graph in Figure
3B).Examples of frequency spectra used for analysis for CFAE with two
closely-spaced frequency components at the low end of range. A.
The power spectrum using the new technique in the range
1–12 Hz. B. A close-up of the power spectrum with a
range of 2 Hz. The parameters used to determine whether the
synthetic frequency components are resolved are shown. C. The
Fourier power spectrum in the range 1–12 Hz. D. A
close-up of the Fourier power spectrum with a range of 2 Hz. The
synthetic frequency components are not resolved in the Fourier
spectrum.The spectral resolution of Fourier analysis versus the new technique in the
presence of interference were determined as follows. The criteria as given by
Equations 11, 12 and 13 were tested for two synthesized spectral components with
periods ω and ω + γ, γ = 2 through
20 spectral points. Again, the lower limit of 2 was necessitated by the fact
that two frequency components occurring at adjacent spectral points
(γ = 1) cannot be resolved. The minimum value of γ for
which these criteria were met was defined as the spectral resolution to discern
the closely-spaced simulated spectral peaks. If these criteria were not met to
γ = 20, the trial was tabulated as being one in which the
resolution was indeterminable due to the loss of one or more spectral peaks in
the noise floor. For both methods, the estimate error was defined as the average
of the absolute difference in fw - fω and
fw+α – fω+γ, where fw
and fω are the measured and actual frequencies of one of the
peaks, and fw+α and fω+γ are the measured
and actual frequencies of the other peak, respectively. For Fourier analysis, a
correction was made by subtracting 0.06 Hz from each error after
calculating the absolute difference of measured to actual frequency, with the
minimum error after subtraction being 0.00 Hz. This allowed for the fact
that for simplicity, the synthetic frequencies that were used were always 1/w,
with w being an integer, so that on average for Fourier analysis, having
constant increments of 0.12 Hz between spectral points, the actual
frequency components could be different by as much as ±0.06 Hz. For
the new technique, the spectral points occurred at 1/w for all w in range, so no
correction was done.The overall paradigm was repeated for 15 trials in each frequency range
3–4 Hz, 4–5 Hz, …, 9–10 Hz for both
Fourier analysis and the new technique. The results were averaged and tabulated
as mean ± standard deviation. The unpaired t-test was
used to determine the significance of the difference between methods (SigmaPlot
2004 for Windows Ver. 9.01, Systat Software, Chicago, and MedCalc ver. 9.5,
2008, MedCalc Software bvba, Mariakerke, Belgium).
Results
Spectral properties
A power spectrum using the new spectral estimation technique is shown in Figure
3A. Note that the highest frequency resolution occurs
at lower frequencies due to the 1/w relationship of resolution to frequency for
this method. By comparison, the Fourier power spectrum is uniform in resolution
across the range (Figure 3C). Figures 3B and D show close-ups of the respective spectra in the range of
the synthesized components. The actual synthesized components have frequencies
of 5.34 Hz (ω + γ = 183 sample points at
977 Hz sampling rate) and 5.43 Hz (ω = 180 sample
points), noted by vertical bars at the tops of panels 3B and 3D. The two
components are correctly resolved by the new technique (panel 3B), that is,
w = ω and
w + ∝ = ω + γ. However,
Fourier analysis does not resolve at this component spacing (panel 3D). In
Figure 4, using the same CFAE and with the high frequency
remaining at 5.43 Hz (ω = 180 sample points), the result
is shown for γ = 19 when the low frequency is 4.91 Hz
(ω + γ = 199 sample points). The spectrum and
close-up using the new method are shown in Figure 4A-B,
and the frequency components are readily resolved as in Figure 3A-B. The Fourier spectrum is shown in Figure 4 C-D and now distinct peaks appear (Figure 4D), meeting the criteria set forth in the Methods. This was the
minimum distance γ at which two corresponding Fourier spectral peaks met
the criteria, and therefore the resolution for the Fourier spectrum. The
measurements for sw+α, sw, smin, and b are
shown.
Figure 4
Same data as in Figure3except that the
synthetic frequency components are spaced further apart (γ is
larger). The panels correspond to those in Figure 3. B. The synthetic frequency components remain
resolved using the new technique. D. The synthetic frequency
components resolve using Fourier analysis.
Same data as in Figure3except that the
synthetic frequency components are spaced further apart (γ is
larger). The panels correspond to those in Figure 3. B. The synthetic frequency components remain
resolved using the new technique. D. The synthetic frequency
components resolve using Fourier analysis.The power spectrum from 1–12 Hz using the new technique with a
different set of signals, and synthesized components near the higher end of the
frequency range is shown in Figure 5A. The components
reside at fω+γ = 8.50 Hz and
fω = 8.65 Hz. Several subharmonics present
in the range 1–4 Hz are evident in Figure 5A.
The Fourier power spectrum from the same synthesized signal is shown in the
range 1–12 Hz in Figure 5C. There are
substantial superharmonics in the high frequency range from
9.5–12 Hz. Close-ups are shown in the right-hand panels for both
methods. Even though the spectral resolution for the new technique is diminished
in this frequency range, it resolves the peaks when γ = 2
(Figure 5B). The parameters used for measurement are shown
in this panel. The Fourier power spectrum does not resolve at this level (panel
5D). There is only a single blunted peak from about 8.5 Hz to
8.75 Hz. The same data is shown for γ = 4 in Figure 6. The frequency components remain involved for the new
method (Figure 6B) and first resolve for Fourier analysis
(Figure 6D). The actual frequency components have values
of fω+γ = 8.35 Hz and
fω = 8.65 Hz. The parameters indicating that
the Fourier spectral peaks resolve are noted (Figure 6D).
Figure 5
Examples of frequency spectra used for analysis for CFAE with two
closely-spaced frequency components at the high end of range.
Panels correspond to those in Figure 3. The
synthetic frequency components reside at approximately 8.5 Hz, and
are resolved by the new technique (panel B). Measurement parameters are
shown. The synthetic frequency components are not resolved by Fourier
analysis (panel D).
Figure 6
Same data as in Figure5except that the
synthetic frequency components are spaced further apart. The
panels correspond to those in Figure 3. The
synthetic frequency components remain resolved using the new technique
in panel B. The synthetic frequency components resolve using Fourier
analysis in panel D.
Examples of frequency spectra used for analysis for CFAE with two
closely-spaced frequency components at the high end of range.
Panels correspond to those in Figure 3. The
synthetic frequency components reside at approximately 8.5 Hz, and
are resolved by the new technique (panel B). Measurement parameters are
shown. The synthetic frequency components are not resolved by Fourier
analysis (panel D).Same data as in Figure5except that the
synthetic frequency components are spaced further apart. The
panels correspond to those in Figure 3. The
synthetic frequency components remain resolved using the new technique
in panel B. The synthetic frequency components resolve using Fourier
analysis in panel D.
Summary statistics
For a total of 105 trials, excluding those in which the synthesized components
could not be resolved, the mean frequency resolution with the presence of
additive interference was 0.29 ± 0.22 Hz for Fourier and
0.16 ± 0.10 Hz for the new technique
(p < 0.001). Thus the resolving power of each method, in
the presence of interference, was similar to their respective theoretical
resolutions of 0.10 Hz and 0.24 Hz. A complete summary of the
resolving power of both techniques is provided in Table 1.
The mean frequency for the random trials in each frequency bin that resolved are
shown in the rows of column 1 from top to bottom (9–10 Hz,
8–9 Hz, …). The expected resolution at 9.5 Hz,
8.5 Hz, …, using the new method, are noted in the second column. The
actual resolving power for the new technique and for the Fourier transform are
shown in the next two columns, respectively. At all frequencies, the new
technique resolving power is improved over the Fourier method. The significance
of the difference in mean resolving power between the methods is noted in the
right-hand column. At lower frequencies up to 7 Hz there is a highly
significant difference (p ≤ 0.003). At higher
frequencies the difference trends toward significance.The average of the absolute misalignment of spectral peaks with actual values of
synthesized components (the estimate error) was
0.023 ± 0.039 Hz for Fourier and
0.009 ± 0.030 Hz for the new technique
(p < 0.001). A complete summary of the estimate error
for both methods is provided in Table 2. The mean
frequency for the trials in each frequency bin that resolved is again shown in
the rows of column 1 from top to bottom, as in Table 1.
The estimate error values for each frequency bin are shown for the new technique
and for Fourier analysis in the middle columns. In each case, the error, or
misalignment of spectral peak with respect to actual component frequencies, is
substantially less using the new technique. The significance of the difference
in error between the new technique and Fourier analysis is noted in the
right-hand column. There is a high degree of significance at all frequencies
except the 9–10 Hz bin.
Table 2
Estimate error in the frequency of peak values
Freq. (Hz)
Error NT (Hz)
Error FT (Hz)
Significance
9.37 ± 0.47
0.032 ± 0.074
0.039 ± 0.070
P = 0.708
8.49 ± 0.24
0.004 ± 0.003
0.016 ± 0.015
P < 0.001
7.51 ± 0.32
0.004 ± 0.002
0.027 ± 0.043
P = 0.006
6.47 ± 0.27
0.005 ± 0.007
0.017 ± 0.023
P = 0.011
5.45 ± 0.32
0.003 ± 0.002
0.018 ± 0.023
P = 0.001
4.37 ± 0.27
0.003 ± 0.002
0.019 ± 0.019
P < 0.001
3.51 ± 0.29
0.013 ± 0.035
0.043 ± 0.053
P = 0.016
6.41 ± 2.01 (MN)
0.009 ± 0.030 (MN)
0.023 ± 0.039 (MN)
P < 0.001
Freq.: the mean frequency used for the test, Error – NT: the
difference between the actual and estimated frequency using the new
method, Error – FT: the absolute difference between actual and
estimated frequency using Fourier analysis, Significance – the
significance of the difference between the techniques based on the
unpaired t-test. MN – mean value.
Estimate error in the frequency of peak valuesFreq.: the mean frequency used for the test, Error – NT: the
difference between the actual and estimated frequency using the new
method, Error – FT: the absolute difference between actual and
estimated frequency using Fourier analysis, Significance – the
significance of the difference between the techniques based on the
unpaired t-test. MN – mean value.Overall, of 15 trials for each frequency bin (105 trials in all), one or both
frequency components did not appear above the noise floor in 13/105 trials using
the Fourier method (12.4%) versus 4/105 trials using the new technique
(3.8%).
Discussion
Summary
In this study a comparison was made between the ability to resolve two
closely-spaced frequency components in the physiologic range of interest using
Fourier power spectral analysis, versus a new technique that utilizes signal
averaging. The synthesized closely-spaced frequency components and two additive
interferences were selected at random from a set of 216 CFAE. The values for
digital sampling rate (977 Hz) and sequence length (N = 8192,
8.4 s sequences) are typical of those used for frequency analysis of CFAE
obtained during clinical EP study. Tests were made in the range
3–10 Hz, the electrophysiologic range for evaluation of atrial
electrical activity. From 105 tests, the mean resolving power of Fourier versus
the new technique (0.29 Hz versus 0.16 Hz;
p < 0.001), were higher than the theoretical values but
in accord with the presence of large interferences that could act to mask the
frequency components. In 13/105 trials, interference masked frequency components
in the Fourier power spectrum. By comparison, this occurred in only 4/105 trials
using the new technique. The error in estimating the synthesized components was
±0.023 Hz using Fourier versus ±0.009 Hz using the new
technique (p < 0.001).
Signal segment not used for spectral estimation
Based upon Eq. 4, whenever n is not an integer, a portion of the 8.4 second CFAE
signal at its end is not used for spectral estimation. This can be described
as:where L is the unused signal segment length in seconds. In Figure 7 is shown a plot of segment length versus frequency in the
physiologic range of interest (3–10 Hz). The maximum unused segment
length, 0.3 seconds, occurs at 3.1 Hz, near the lowest frequency in the
range. The mean unused segment length is approximately 0.1 seconds. Thus on the
average, 8.3 seconds of the 8.4 second signal is used for power spectral
estimation, meaning that only about 1% of the signal is not included in the
measurement.
Figure 7
Plot of the segment length at the end of the each CFAE signal that is
unused for each power spectrum calculation using the new technique,
versus frequency in the physiologic range.
Plot of the segment length at the end of the each CFAE signal that is
unused for each power spectrum calculation using the new technique,
versus frequency in the physiologic range.
Clinical correlates
The dynamics of internal fibrillatory activity are of great interest to establish
AF mechanisms as well as for catheter ablation, and can be measured using
markers of organization [12]. It would
be desirable to correlate these invasive measurements with noninvasive
electrocardiogram measurements, to determine the extent to which AF can be
characterized noninvasively with for example, a Holter monitor [12]. It is also desirable to distinguish
slightly different levels of organization during AF by improving temporal
resolution and sensitivity [13]. Such
improvements could lead to new ways of analyzing and understanding AF, and
improved AF treatment methods [13]. The
parameters used in our study for analysis of CFAE are typical for clinical
investigations. The capability of the Fourier transform to resolve two
independent frequency components in close proximity, 0.24 Hz, is
unacceptable for describing the temporal evolution of paroxysmal AF or to
differentiate this type from persistent AF. Using Fourier analysis, the dominant
frequency, defined as the tallest spectral peak in the physiologic range of
interest, often varies with a standard deviation of 0.2–0.3 Hz when
CFAE are recorded from each of the four pulmonary vein ostia [14]. Thus the resolving power using the
Fourier spectrum is likely to be insufficient for detailed, accurate
measurement. To determine whether differences in the frequencies of spectral
peaks are real or artifact requires an increase in resolving power, as is
possible with our new technique. Furthermore, the ability to resolve multiple
characteristic frequencies in AF signals, which may be caused by multiple
wavelets propagation as are commonly found in AF data, is more difficult when
the electrical activity is highly unorganized [15]. Applying the new technique to these signals can
potentially increase the ability to spectrally resolve the multiple activation
wavelets that can be present in the arrhythmogenic substrate.
Conclusions
In this study we demonstrated that the resolving power to discern two closely-spaced
synthesized frequency components was significantly greater using a new spectral
estimation technique as compared with Fourier analysis. Furthermore, the new
technique estimates the actual frequencies of the components to a significantly
greater degree of accuracy as compared with Fourier. The simulations suggest that
two closely-spaced frequency components, as might be encountered by drivers of
electrical activation in close spatial proximity to one another within the
arrhythmogenic substrate that is causing atrial fibrillation, can be discerned by
the new technique, even in combination with substantial interference levels. Such
interferences may represent, for example, independent drivers at the periphery of
the substrate that are inconsequential to the maintenance of AF. Thus this method
may be helpful to detect and identify independent drivers in the atrial substrate
that can be masked by interference from bystander regions. Such information is
potentially important to develop a mechanistic understanding of the pattern of
electrical activation during atrial fibrillation. Although only CFAE simulations
were done in this study, these observations are probably applicable to other types
of signals including ventricular electrograms [16] and electrocardiogram data [17]. Furthermore, better spectral resolution can be
important for improved recognition of repeating patterns in CFAE when analyzed in
the frequency domain [18].
Limitations
For simplicity, the number of trials for discerning closely-spaced frequency
components was limited to 15 for each of seven frequency bins in the
electrophysiologic range of interest. The types of interference signals and the
types of periodic components that were used were limited to those present in a
pool of 216 CFAE recordings obtained from the pulmonary vein ostia and left
atrial free wall in both paroxysmal and longstanding AFpatients. The use of
synthesized components was necessitated by the fact that frequency values needed
to be known with certainty. Since synthesized signals were tested, they will not
necessarily represent the performance of Fourier spectral analysis versus the
new technique on real data. Still, since the synthesized frequency components
and interferences were extracted from real CFAE, we believe that they are
representative of the types of signals that would be encountered in real
measured data.
Competing interests
The authors declare that they have no competing interests.
Author contributions
EJC developed the mathematical methods, conducted the data analysis, and wrote the
manuscript. ABB, WW, and HG made helpful suggestions, provided the clinical data,
and determined which recordings were complex fractionated atrial electrograms. All
authors have read and approved the final manuscript.
Authors: Edward J Ciaccio; Angelo B Biviano; William Whang; Andrew L Wit; Hasan Garan; James Coromilas Journal: Heart Rhythm Date: 2008-10-15 Impact factor: 6.343
Authors: Edward J Ciaccio; Angelo B Biviano; Alok Gambhir; Jason T Jacobson; Hasan Garan Journal: Pacing Clin Electrophysiol Date: 2013-08-26 Impact factor: 1.976
Authors: João L Salinet; Nicholas Masca; Peter J Stafford; G André Ng; Fernando S Schlindwein Journal: Biomed Eng Online Date: 2016-03-08 Impact factor: 2.819