Evaluation of implant-abutment microgap and bacterial leakage in five external-hex implant systems: an in vitro study.

Eduardo Cláudio Lopes de Chaves e Mello Dias1, Edecir Décio Cargnin Bisognin, Nassim David Harari, Sílvio Jorge Machado, Cristiano Pereira da Silva, Glória Dulce de Almeida Soares, Guaracilei Maciel Vidigal.   

Abstract

PURPOSE: The aim of the present study was to verify the presence of a microgap between implants and prosthetic abutments on their external surfaces in five different external-hex implant systems and to determine whether bacterial leakage occurs through the implant-abutment (I-A) interface.
MATERIALS AND METHODS: Ten samples of each implant system were used. Eight samples of each implant system were inoculated with 0.3 μL of a suspension of Escherichia coli for bacterial leakage evaluation. Two other samples were used as controls. Sample analysis was accomplished at 24 and 48 hours and on the 5th, 7th, and 14th day after inoculation. After microbiologic analysis, all samples were prepared for I-A interface evaluation by scanning electron microscopy. The mean of the microgap misfit was obtained from six points at x1,000 and x3,000 magnifications.
RESULTS: The Neodent implant system (0.51 ± 0.39 μm) showed the closest adaptation, followed by Dentoflex (1.44 ± 0.73 μm), Titanium Fix (1.88 ± 1.28 μm), SIN (2.46 ± 3.38 μm), and Conexão (2.68 ± 3.02 Μm). Twenty-five percent of the Dentoflex samples showed bacterial leakage through the I-A interface; no other implant system presented bacterial leakage.
CONCLUSION: The width of the microgap at the I-A interface was less than 3 μm in all systems. A direct correlation between bacterial leakage and I-A interface misfit was not observed.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22442774

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  6 in total

1.  A new system of implant abutment connection: how to improve a two piece implant system sealing.

Authors:  F Grecchi; M DI Girolamo; F Cura; V Candotto; F Carinci
Journal:  Oral Implantol (Rome)       Date:  2017-11-30

2.  Impact of Dynamic Loading on the Implant-abutment Interface Using a Gas-enhanced Permeation Test In Vitro.

Authors:  Anas Al-Jadaa; Thomas Attin; Timo Peltomäki; Christian Heumann; Patrick Roger Schmidlin
Journal:  Open Dent J       Date:  2015-03-31

Review 3.  Abutment misfit in implant-supported prostheses manufactured by casting technique: An integrative review.

Authors:  Lorena M S Pereira; Mariane B Sordi; Ricardo S Magini; Antônio R Calazans Duarte; Júlio C M Souza
Journal:  Eur J Dent       Date:  2017 Oct-Dec

4.  Microleakage Evaluation at Implant-Abutment Interface Using Radiotracer Technique.

Authors:  Hakimeh Siadat; Mahnaz Arshad; Hossein-Ali Mahgoli; Babak Fallahi
Journal:  J Dent (Tehran)       Date:  2016-06

5.  Microbiological Sealing Analysis of a Tapered Connection and External Hexagon System.

Authors:  Gardel Nepomuceno Costa; Elizabeth Ferreira Martinez; Aluísio Martins de Oliveira Ruellas; Daiane Cristina Peruzzo; Júlio Cesar Joly; Marcelo Henrique Napimoga
Journal:  Int J Dent       Date:  2017-02-28

6.  The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study.

Authors:  Sunjai Kim; Changhun Choi; Yunsu Cha; Jae-Seung Chang
Journal:  BMC Oral Health       Date:  2021-02-18       Impact factor: 2.757

  6 in total