Rami Doukky1, Mouyyad Rahaby, Dave Chawla, Raj Vashistha, Tareq Alyousef, Amit P Amin.
Abstract
BACKGROUND: Little is known about soft tissue attenuation artifacts when an upright patient-position SPECTmyocardial perfusion imaging (MPI) system is used. In this investigation we sought to describe the patterns and frequency of attenuation artifacts associated with this type of instruments and we explored the impact of gender and body habitus on these artifacts.Entities:
Keywords: Myocardial Perfusion Imaging; SPECT; artifacts; interpretation.; soft tissue attenuation; upright
Year: 2012 PMID: 22435079 PMCID: PMC3308262 DOI: 10.2174/1874192401206010022
Source DB: PubMed Journal: Open Cardiovasc Med J ISSN: 1874-1924
Baseline Characteristics
| 212 | |
| 95 (44.8%) | |
| 55.8 (±12.5) | |
| 64.2% | |
| 25.0% | |
| 46.1% | |
| 11.8% | |
| 21.7% | |
| 110.0 (±11.2) | |
| 110.9 (±9.7) | |
| 109.0 (±12.7) | |
| 31.6 (±6.0) | |
| 30.1 (±5.6) | |
| 32.8 (±6.7) | |
| 134 (63.2%) | |
| 16.21 (±8.1) | |
| 12.42 (±9.2) | |
| 0.52 | |
| 65.3% (±8.8) |
CAD: Coronary artery disease; BMI: Body Mass Index; CHD: coronary Heart Disease.
Relevant to symptomatic patients only and was based on Diamond and Forrester prediction tables 12.
Rates of Attenuation Defects Observed
| Total, |
Men, |
Women, | Chi-square p-value (Men vs. Women) | |
|---|---|---|---|---|
| 13 (6.1%) | 4 (3.4%) | 9 (9.5%) | 0.07 | |
| 135 (63.7%) | 88 (75.2%) | 47 (49.5%) | < 0.001 | |
| 51 (24.1%) | 18 (15.4%) | 33 (34.7%) | 0.001 |
Characteristics of the Attenuation Patterns Observed
| Attenuation Pattern | Prevalence | Gender | Effect of Chest Size on Prevalence | Effect of BMI on Prevalence | Factors that Increase Severity |
|---|---|---|---|---|---|
| Anterior | Uncommon | ?F > M | + | - | Chest size |
| Inferior | Very common | M > F | - | - | BMI Chest size |
| Lateral | Common | F > M | - | + | None |