| Literature DB >> 22433063 |
Qinan Bao1, Brian A Hrycushko, Joseph P Dugas, Frederick H Hager, Timothy D Solberg.
Abstract
BACKGROUND: Total skin electron irradiation (TSEI) is a special radiotherapy technique which has generally been used for treating adult patients with mycosis fungoides. Recently, two infants presented with leukemia cutis isolated to the skin requiring TSEI. This work discusses the commissioning and quality assurance (QA) methods for implementing a modified Stanford technique using a rotating harness system to position sedated pediatric patients treated with electrons to the total skin. METHODS ANDEntities:
Mesh:
Year: 2012 PMID: 22433063 PMCID: PMC3349519 DOI: 10.1186/1748-717X-7-40
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Combined beam profile at 330 cm SSD. The gantry was directed at ± 20° from the horizontal.
Figure 2The patient in treatment positions for (a) PA and (b) RAO locations.
OSL detector measured dose on 20 cm diameter cylindrical phantom following delivery of a prescribed 50 cGy
| OSL Detector Location | NanoDot 1 Dose (cGy) | NanoDot 2 Dose (cGy) | Average Dose (cGy) | % Diff from Prescription |
|---|---|---|---|---|
| AP | 54.3 | 53.3 | 53.8 | 7.6 |
| Lt Lat | 53.5 | 54.4 | 54.0 | 8.0 |
| PA | 53.9 | 53.1 | 53.5 | 7.0 |
| Rt Lat | 54.5 | 52.4 | 53.4 | 6.8 |
Figure 3Film measurement for the composite TSEI beam. (a) An XV film was uniformly irradiated with 50 cGy dose and (b) two orthogonal profiles indicates the dose penetration of the TSEI beam.
OSL detector dose measurements for cycle one of first patient
| OSL Detector Location | AP + LPO + RPO Day 1 Dose (cGy) | PA + LAO + RAO Day 2 Dose (cGy) | Total Dose (cGy) | % Diff from Prescription |
|---|---|---|---|---|
| Forehead | 106.2 | 119.1 | 225.3 | 12.7 |
| Posterior of head | 152.1 | 96.2 | 248.3 | 24.2 |
| Sternum | 78.9 | 136.9 | 215.8 | 7.9 |
| Posterior of sternum | 137.8 | 85.9 | 223.7 | 11.9 |
| Umbilicus | 80.5 | 132.7 | 213.2 | 6.6 |
| Posterior of Umbilicus | 146.9 | 78.7 | 225.6 | 12.8 |
| Left thigh | 75.8 | 139.1 | 214.9 | 7.5 |
| Posterior Left Thigh | 145.9 | 81.5 | 227.4 | 13.7 |
| Right thigh | 77.1 | 133.7 | 210.8 | 5.4 |
| Posterior Right Thigh | 130.0 | 86.5 | 216.5 | 8.3 |
| Right Foot | 84.6 | 141.6 | 226.2 | 13.1 |
| Bottom of Right Foot | 135.9 | 82.6 | 218.5 | 9.3 |
OSL detector dose measurements for cycle two of first patient for reduced MUs
| OSL Detector Location | AP + LPO + RPO Day 3 Dose (cGy) | PA + LAO + RAO Day 4 Dose (cGy) | Total Dose (cGy) | % Diff from Prescription |
|---|---|---|---|---|
| Forehead | 96.1 | 117.2 | 213.3 | 6.6 |
| Posterior of Head | 121.9 | 84 | 205.9 | 3.0 |
| Umbilicus | 79.1 | 117.3 | 196.3 | 18 |
| Posterior of Umbilicus | 125.6 | 72.6 | 198.2 | 09 |
| Back of Right Hand | 70.5 | 125.3 | 195.7 | 21 |