Fretting corrosion studies of universal femoral head prostheses and cone taper spigots.

D C Fricker1, R Shivanath.   

Abstract

Short-duration, cyclically loaded, axial, fretting corrosion tests were designed and performed to compare the fretting behaviour of different metal Howmedica universal heads connected to coated and uncoated metal cone taper spigots. Concurrent axial push-on and pull-off taper friction tests were also performed. There was no measurable fretting debris found in any test and SEM studies of the contact surfaces showed no evidence of fretting. It is concluded that no significant, long-term fretting corrosion of the Vitalium Co-Cr-Mo heads on the Vitallium or uncoated Ti-6AI-4V spigots, nor of the Orthinox stainless steel heads on the Orthinox spigots, is likely to occur in vivo.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2242398     DOI: 10.1016/0142-9612(90)90064-w

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  3 in total

1.  Proximal component modularity in THA--at what cost? An implant retrieval study.

Authors:  A M Kop; C Keogh; E Swarts
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

2.  In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper Design.

Authors:  Massimiliano Baleani; Paolo Erani; Barbara Bordini; Federica Zuccheri; Mateusz Kordian Mąkosa; Dalila De Pasquale; Alina Beraudi; Susanna Stea
Journal:  Materials (Basel)       Date:  2017-07-01       Impact factor: 3.623

Review 3.  Mechanisms of Adverse Local Tissue Reactions to Hip Implants.

Authors:  Felipe Eltit; Qiong Wang; Rizhi Wang
Journal:  Front Bioeng Biotechnol       Date:  2019-07-30