Evaluation of clear corneal wound dynamics with contrast-enhanced spectral-domain optical coherence tomography.

Preeya K Gupta1, Justis P Ehlers, Terry Kim.   

Abstract

BACKGROUND AND
OBJECTIVE: To evaluate wound morphology and integrity using contrast-enhanced spectral-domain optical coherence tomography (SD-OCT). PATIENTS AND METHODS: Eighteen patients (18 eyes) undergoing unilateral routine cataract surgery were enrolled. SD-OCT imaging of the corneal wound was performed immediately following cataract surgery and on postoperative day 1. A topical contrast agent, prednisolone acetate 1%, was used to help evaluate wound integrity. OCT features of the corneal incision and changes in wound interface reflectivity after administration of topical prednisolone acetate 1% were analyzed. Main outcome measures were quantitative and qualitative description of wound morphology and contrast enhancement of the wound interface.
RESULTS: Internal wound gape was found in 89% of eyes. Gape area decreased by 43% from days 0 to 1 (462 vs 273 μm(2); P < .01). Contrast enhancement at the wound interface was seen in 56% of eyes on day 0 compared to 17% on day 1 following application of prednisolone acetate 1% (P < .05). Decreased intraocular pressure correlated with increased contrast enhancement (P < .05).
CONCLUSION: Wound morphology evolves during the early postoperative period. Wound gape and ingress of a topical contrast agent decrease during the first 24 hours following surgery. Contrast enhancement at the wound interface increases as the intraocular pressure decreases. Copyright 2012, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22421201     DOI: 10.3928/15428877-20120308-05

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  3 in total

1.  Visualisation of contrast-enhanced intraoperative optical coherence tomography with indocyanine green.

Authors:  Justis P Ehlers; Stephen McNutt; Suhail Dar; Yuankai K Tao; Sunil K Srivastava
Journal:  Br J Ophthalmol       Date:  2014-04-23       Impact factor: 4.638

2.  Comparison of new motorized injector vs manual injector for implantation of foldable intraocular lenses on wound integrity: an ASOCT study.

Authors:  S Khokhar; R Sharma; B Patil; N Aron; S Gupta
Journal:  Eye (Lond)       Date:  2014-07-18       Impact factor: 3.775

3.  Wound Integrity of 2.0 mm Transconjunctival Single-Plane Sclerocorneal Incision: A Comparison between a Motorized Injector with/without Pause Time and a Manual Injector.

Authors:  Hisanori Imai; Ayaka Fujii; Emiko Tani; Atsushi Azumi
Journal:  J Ophthalmol       Date:  2016-02-17       Impact factor: 1.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.