Evaluation of three applanation tonometers in dogs.

D R Priehs1, G G Gum, R D Whitley, L E Moore.   

Abstract

The Mackay-Marg, Tono-Pen, and Challenger applanation tonometers were evaluated in vivo in 12 clinically normal eyes of 6 dogs. Tonometric measures of intraocular pressure (IOP) were compared with closed manometric IOP measurements from the anterior chamber of anesthetized dogs. The tonometers were evaluated at IOP that ranged from 5 to 100 mm of Hg. The Mackay-Marg tonometer was the most reliable instrument when evaluated at IOP from 5 to 100 mm of Hg (r2 = 0.996) and from 10 to 30 mm of Hg (r2 = 0.962). The Tono-Pen tonometer was also reliable (r2 = 0.967) over the range of IOP, but consistently overestimated IOP at lower pressures and underestimated IOP at higher pressures. The Mackay-Marg and Tono-Pen measurements were essentially linear. When evaluated from 10 to 30 mm of Hg, r2 was 0.828 for the Tono-Pen tonometer. The Challenger tonometer, although reliable over the full range of IOP (r2 = 0.965), proved to be less accurate, as indicated by lack of a good linear equation.

Entities:  

Mesh:

Year:  1990        PMID: 2240775

Source DB:  PubMed          Journal:  Am J Vet Res        ISSN: 0002-9645            Impact factor:   1.156


  5 in total

1.  Intraocular pressure measurements in cattle, sheep, and goats with 2 different types of tonometers.

Authors:  Nina Peche; Johanna Corinna Eule
Journal:  Can J Vet Res       Date:  2018-07       Impact factor: 1.310

2.  Validation and comparison of four handheld tonometers in normal ex vivo canine eyes.

Authors:  Andrea L Minella; Julie A Kiland; Shawna Gloe; Gillian J McLellan
Journal:  Vet Ophthalmol       Date:  2020-06-01       Impact factor: 1.644

3.  Comparison of two rebound tonometers in healthy horses.

Authors:  Minna P Mustikka; Elina M Pietilä; Anna K Mykkänen; Thomas S C Grönthal
Journal:  Vet Ophthalmol       Date:  2020-09-04       Impact factor: 1.644

4.  Comparative intraocular pressure measurements using three different rebound tonometers through in an ex vivo analysis and clinical trials in canine eyes.

Authors:  Jaeho Shim; Seonmi Kang; Yoonji Park; Sunhyo Kim; Seokmin Go; Eunji Lee; Kangmoon Seo
Journal:  Vet Ophthalmol       Date:  2020-06-05       Impact factor: 1.644

5.  Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.

Authors:  João Victor Goulart Consoni Passareli; Felipe Franco Nascimento; Giovana José Garcia Estanho; Claudia Lizandra Ricci; Glaucia Prada Kanashiro; Rogério Giuffrida; Silvia Franco Andrade
Journal:  Vet World       Date:  2021-09-21
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.