D R Priehs1, G G Gum, R D Whitley, L E Moore.
Abstract
The Mackay-Marg, Tono-Pen, and Challenger applanation tonometers were evaluated in vivo in 12 clinically normal eyes of 6 dogs. Tonometric measures of intraocular pressure (IOP) were compared with closed manometric IOP measurements from the anterior chamber of anesthetized dogs. The tonometers were evaluated at IOP that ranged from 5 to 100 mm of Hg. The Mackay-Marg tonometer was the most reliable instrument when evaluated at IOP from 5 to 100 mm of Hg (r2 = 0.996) and from 10 to 30 mm of Hg (r2 = 0.962). The Tono-Pen tonometer was also reliable (r2 = 0.967) over the range of IOP, but consistently overestimated IOP at lower pressures and underestimated IOP at higher pressures. The Mackay-Marg and Tono-Pen measurements were essentially linear. When evaluated from 10 to 30 mm of Hg, r2 was 0.828 for the Tono-Pen tonometer. The Challenger tonometer, although reliable over the full range of IOP (r2 = 0.965), proved to be less accurate, as indicated by lack of a good linear equation.Entities:
Mesh:
Year: 1990 PMID: 2240775
Source DB: PubMed Journal: Am J Vet Res ISSN: 0002-9645 Impact factor: 1.156