Guilherme Machado de Carvalho1, João Paulo Peral Valente, Alexandre Scalli Mathias Duarte, Eder Barbosa Muranaka, Alexandre Caixeta Guimarães, Marcelo Naoki Soki, Walter Adriano Bianchini, Arthur Menino Castilho, Jorge Rizzato Paschoal.
Abstract
UNLABELLED: A new era has arrived in auditory rehabilitation with the introduction of new technologies such as electroacoustic stimulation (EAS). EAS is indicated for patients with residual hearing at low frequencies and severe or profound hearing loss at high frequencies. These patients have no indication for conventional cochlear implant and have difficulties in adapting to individual sound amplification devices. Preservation of hearing is vital in this process; the surgical technique must be based on this concept.Entities:
Mesh:
Year: 2012 PMID: 22392237 PMCID: PMC9443873 DOI: 10.1590/s1808-86942012000100007
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Figure 1Chart showing an example of an expected audiogram in candidates for hybrid implants.11
Figure 2Schematic representation of the Electrode Med-El MAnd FLEXEAS ™(21mm).
Figure 3Amplified schematic showing the wave shape of electrodes (sinus) in the silicone sheath with a wire structure.
Technical specifications of the Med-El MAnd FLEXEAS cochlear implant
| Model: SONATAti100 ELECTRODE FLEXEAS | |
|---|---|
| Material: TITANIUM AND SILICONE | |
| Removable magnet | NO |
| Size | Length: 45.7 mm |
| Width: 24.8 mm | |
| Thickness: 5.9 mm | |
| Weight: 8.6g | |
| Receiver well for the internal unit | Diameter: 24.8 × 17.4 (mm × mm) |
| Depth: 2 mm | |
| Electrode array | Perimodiolar: no |
| Straight: yes | |
| Dimensions of electrode array | Total length of electrode array: 24 mm |
| Length of electrodes that remain within the cochlea: 20.9 mm | |
| Length of active electrodes: 20.9 mm | |
| Number of electrodes: 12 channels (19 electrodes – 5 apex channels are not double for residual preservation) | |
| Apex: electrode number 1 (number of the most apical electrode). | |
| Base: electrode number 12 | |
| Diameter of electrode array on the apex: 0.3 mm | |
| Diameter of electrode array on the base: 0.8 mm | |
| Diameter of the cochleostomy: 1.3 mm | |
| Allow MR | yes |
| Technical specifications: 0.2T, 1.0T e 1.5T | |
Please, fill out the questionnaire and prior instructions from Med El.
Figure 4Representative drawing of surgical landmarks to the left (tip of the mastoid, temporal line, retroauricular incision area, area of the internal component, and microphone area with a mock-up).
Figure 5Illustration of preparing the receiver well for the internal component of the EAS cochlear implant in the squamous portion of the temporal bone, with a specific mock-up;
Figure 6Ilustration of irrigation of the mastoid cavity and middle ear with povidone-iodine solution (10% povidone-iodine / 1% active iodine).