Response to albuterol MDI delivered through an anti-static chamber during nocturnal bronchospasm.

Sreekala Prabhakaran1, Jonathan Shuster, Sarah Chesrown, Leslie Hendeles.   

Abstract

BACKGROUND: Decreasing electrostatic charge on valved holding chambers increases the amount of drug delivered. However, there are no data demonstrating that this increases bronchodilatation.
OBJECTIVE: To investigate the influence of reducing electrostatic charge on the bronchodilator response to albuterol inhaler during nocturnal bronchospasm.
METHODS: This randomized double-blind, double-dummy crossover study included subjects, 18-40 years old, with nocturnal bronchospasm (20% overnight decrease in peak flow on 3 of 7 nights during run-in), FEV(1) 60-80% predicted during the day, and ≥ 12% increase after albuterol. Subjects slept in the clinical research center up to 3 nights for each treatment. FEV(1) and heart rate were measured upon awakening spontaneously or at 4:00 am, and 15 min after each dose of 1, 2, and 4 cumulative puffs of albuterol via metered-dose inhaler. The drug was administered through an anti-static valved holding chamber (AeroChamber Plus Z-Stat) or a conventional valved holding chamber containing a static charge (AeroChamber Plus).
RESULTS: Of 88 consented subjects, 11 were randomized and 7 completed the study. Most exclusions were due to lack of objective evidence of nocturnal bronchospasm. Upon awakening, FEV(1) was 44 ± 9% of predicted before the anti-static chamber and 48 ± 7% of predicted before the static chamber. The mean ± SD percent increase in FEV(1) after 1, 2, and 4 cumulative puffs using the anti-static versus the static chamber, respectively, were 52 ± 26% versus 30 ± 19%, 73 ± 28% versus 48 ± 26%, and 90 ± 34% versus 64 ± 35%. The point estimates for the differences (and 95% CIs) between the devices (anti-static vs static) were 21% (4-38%) (P = .03), 23% (6-41%) (P = .02), and 25% (7-42%) (P = .01) for 1, 2, and 4 cumulative puffs, respectively. There was no significant difference in heart rate between treatments.
CONCLUSIONS: Delivery of albuterol through an anti-static chamber provides a clinically relevant improvement in bronchodilator response during acute, reversible bronchospasm such as nocturnal bronchospasm.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22348270      PMCID: PMC3806197          DOI: 10.4187/respcare.01572

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  16 in total

1.  Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves delivery.

Authors:  F Piérart; J H Wildhaber; I Vrancken; S G Devadason; P N Le Souëf
Journal:  Eur Respir J       Date:  1999-03       Impact factor: 16.671

Review 2.  Consensus statement: aerosols and delivery devices. American Association for Respiratory Care.

Authors:  M A Dolovich; N R MacIntyre; P J Anderson; C A Camargo; N Chew; C H Cole; R Dhand; J B Fink; N J Gross; D R Hess; A J Hickey; C S Kim; T B Martonen; D J Pierson; B K Rubin; G C Smaldone
Journal:  Respir Care       Date:  2000-06       Impact factor: 2.258

3.  Response to inhaled albuterol during nocturnal asthma.

Authors:  Leslie Hendeles; Russell Beaty; Richard Ahrens; Gary Stevens; Eloise M Harman
Journal:  J Allergy Clin Immunol       Date:  2004-06       Impact factor: 10.793

Review 4.  Effect of add-on devices for aerosol drug delivery: deposition studies and clinical aspects.

Authors:  S P Newman; M T Newhouse
Journal:  J Aerosol Med       Date:  1996

5.  On comparing inhaled beta adrenergic agonists.

Authors:  R C Ahrens
Journal:  Ann Allergy       Date:  1991-09

6.  The importance of nonelectrostatic materials in holding chambers for delivery of hydrofluoroalkane albuterol.

Authors:  Joseph L Rau; Dominic P Coppolo; Mark W Nagel; Valentina I Avvakoumova; Cathy C Doyle; Kimberly J Wiersema; Jolyon P Mitchell
Journal:  Respir Care       Date:  2006-05       Impact factor: 2.258

7.  Spirometric reference values from a sample of the general U.S. population.

Authors:  J L Hankinson; J R Odencrantz; K B Fedan
Journal:  Am J Respir Crit Care Med       Date:  1999-01       Impact factor: 21.405

8.  Electrostatic charge on a plastic spacer device influences the delivery of salbutamol.

Authors:  J H Wildhaber; S G Devadason; M J Hayden; R James; A P Dufty; R A Fox; Q A Summers; P N LeSouëf
Journal:  Eur Respir J       Date:  1996-09       Impact factor: 16.671

9.  Effect of electrostatic charge in plastic spacers on the lung delivery of HFA-salbutamol in children.

Authors:  J Anhøj; H Bisgaard; B J Lipworth
Journal:  Br J Clin Pharmacol       Date:  1999-03       Impact factor: 4.335

10.  Problems patients have using pressurized aerosol inhalers.

Authors:  G K Crompton
Journal:  Eur J Respir Dis Suppl       Date:  1982
View more
  3 in total

Review 1.  Optimizing the Delivery of Inhaled Medication for Respiratory Patients: The Role of Valved Holding Chambers.

Authors:  R Andrew McIvor; Hollie M Devlin; Alan Kaplan
Journal:  Can Respir J       Date:  2018-04-04       Impact factor: 2.409

2.  A Retrospective Study of the Effectiveness of the AeroChamber Plus® Flow-Vu® Antistatic Valved Holding Chamber for Asthma Control.

Authors:  Chakkarin Burudpakdee; Vladimir Kushnarev; Dominic Coppolo; Jason A Suggett
Journal:  Pulm Ther       Date:  2017-07-14

Review 3.  A review of the in vitro and in vivo valved holding chamber (VHC) literature with a focus on the AeroChamber Plus Flow-Vu Anti-static VHC.

Authors:  Sanjeeva Dissanayake; Jason Suggett
Journal:  Ther Adv Respir Dis       Date:  2018 Jan-Dec       Impact factor: 4.031

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.