Single-image axial localization precision analysis for individual fluorophores.

Michael C DeSantis1, Shannon Kian Zareh, Xianglu Li, Robert E Blankenship, Y M Wang.   

Abstract

Bio-mechanism investigations demand single particle tracking with high spatial and temporal resolutions which require single fluorophore 3D localization measurements with matching precision and speed. Although the precision for lateral-localization measurements is well described by an analytical expression, for the axial direction, it is often obtained by repeating location measurements or by estimating a lower bound. Here, we report a precision expression for an axial-localization method that analyzes the standard deviations of single fluorophores' intensity profiles. Like the lateral-localization precision, this expression includes all relevant experimental effects measurable from a gaussian intensity profile of the fluorophore. This expression completes the precision analysis for single-image 3D localization of individual fluorophores and lifts the temporal resolution to the typical exposure timescales of milliseconds.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22330542      PMCID: PMC3482922          DOI: 10.1364/OE.20.003057

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


1. Introduction

Single particle tracking (SPT), whose building blocks are consecutive 3D localization measurements of the investigated particle, is important for many biological investigations [1-3]. To achieve SPT studies with high spatial and temporal resolutions, it is essential to determine the precisions to the 3D localization measurements of the investigated particle in an accurate and timely manner [2]. In current single molecule localization measurements using standard imaging setups, such as in epifluorescence and total internal reflection fluorescence (TIRF) imaging measurements, the intensity profile of a stationary or a slowly moving molecule (relative to the imaging timescale) is called a point spread function (PSF) and can be approximated by a Gaussian function. The centroid of the Gaussian fit to the PSF yields the lateral location of the molecule, and the standard deviation (SD) can be used to determine the molecule’s axial position or defocusing distance (relative to the focal plane) [4] as well as the precision to the lateral location measurements [5, 6]. Although the lateral-localization precision can be expressed by fitting parameters of an individual fluorophore’s single image (PSF and background intensities) [5, 6], thus enabling lateral localization with lateral precision in the typical single-image exposure timescales of milliseconds, the axial-localization precision awaits an analogous expression to complete 3D localization with 3D precision in milliseconds. Current single fluorophore axial-localization methods – which include methods that analyze PSFs obtained using standard imaging setups, such as inferring the axial location from the measured PSF SD [4, 7–10] and using novel algorithms [11-14], and methods that require instrumentation modifications [4, 14–17] – use two approaches in quantifying precision: (i) experimentally repeating the axial-location measurement and using the SD of the axial-location distribution for precision [4, 11, 12, 15–17], and (ii) using an estimation, either the SD of the calculated axial locations according to novel algorithms [12, 14] or a Cramér-Rao lower bound that predicts the lowest possible error to the axial location measurements [9,12,13]. For the first approach, although the results include the instrumentation and fluorophore effects (camera’s readout noise and pixelation effect; PSF’s photon noise, respectively), it decreases the temporal resolution of SPT by at least 20-fold due to repeated imaging; for the second approach, (i) the Cramér-Rao calculation is an estimation of the lower bound, rather than a precise description of the axial-localization precision, and (ii) it does not consider the instrumentation and fluorophore effects, which can affect the precision determination considerably [5, 6]. Furthermore, due to the lack of an expression that includes all experimental noise parameters, neither method can be used to assess axial-localization precisions in alternative experimental settings. Here we report a precision expression for a fluorophore’s single-image axial location measurement that includes the instrumentation and fluorophore’s noise effects, based on the axial-localization method that measures PSF’s SDs. This axial-localization method involves minimal modifications to standard single-molecule imaging setups, and all the parameters in the axial precision expression can be obtained from a single Gaussian fit to the PSF of the fluorophore: SD, photon count, pixel size, and background noise values. This axial-precision expression is a new addition to our single-image molecular analysis (SIMA) studies [19]; it completes individual fluorophores’ single-image 3D localization measurements by providing the axial-direction precision, and lifts the temporal resolution of SPT to the typical single-image exposure timescales of milliseconds.

2. Materials and methods

2.1. Sample preparation and imaging

In this study, we used phycobilisome (PBS; Synechosystis PCC 6803) protein complexes for our z and Δz investigations. PBSs are large hemidiscoidal-shaped light harvesting antenna protein complexes in cyanobacteria; the molecular weight is ≈ 10 MDa and it measures 60 × 30 × 20 nm in width, height, and length [20]. A PBS molecule contains 144 fluorophores distributed throughout the complex; consequently, it serves as an ideal emitter to meet the demands of our study due to its brightness and long fluorescence lifetime. PBSs were purified following the method described in Ref. [21]; the purified PBSs were then crosslinked according to a protocol from the Noam Adir group [22]. The PBSs were diluted in 20 mM Tris-HCl buffer (pH 8.0) to approximately 0.1 nM. Manufacturer pre-cleaned fused-silica chips (6W675-575 20C, Hoya Corporation USA, San Jose, CA) were used, where isolated PBS molecules were adsorbed to surfaces at low concentration. A PBS solution of 5 μL was sandwiched between the fused-silica surface and an oxygen-plasma-cleaned coverslip (2.2 × 2.2 cm2), resulting in a 10.5 μm thick water layer. Because of the prism TIRF imaging setup, our sample on the fused-silica surface is 10.5 μm away from the coverslip surface; therefore, refractive index mismatch will affect the PBS SD versus z relation [17, 18]. Single-molecule imaging was performed using a Nikon Eclipse TE2000-S inverted microscope (Nikon, Melville, NY) in combination with a Nikon 100X objective (Nikon, 1.49 N.A., oil immersion). Samples were excited by prism-type TIRF microscopy with a linearly polarized 568 nm laser line (I70C-SPECTRUM Argon/Krypton laser, Coherent Inc., Santa Clara, CA) focused on a 40 × 20 μm2 region. The incident angle at the fused-silica water interface was 70° with respect to the normal. The 568 nm line was filtered from the multiline laser emission using a polychromatic acousto-optic filter (48062 PCAOM model, NEOS Technologies, Melbourne, FL). The laser excitation was pulsed with illumination intervals of 10 ms; the excitation intensity was 5.2 kW/cm2. Images were captured by an iXon back-illuminated electron multiplying charge coupled device (EMCCD) camera (DV897ECS-BV, Andor Technology, Belfast, Northern Ireland). An additional 2X expansion lens was placed before the EMCCD, producing a pixel size of 79 nm. The excitation filter was 568/20 nm, and the emission filter was a 580 nm long pass filter.

2.2. Data acquisition and selection

PBS movies were obtained by synchronizing the onset of camera exposure with laser illumination. The maximum gain level of the camera was used and the data acquisition rate was 1 MHz pixels/sec (≈ 3.3 frames/sec). Single-molecule images were checked such that there were no saturations in the intensity profiles. The axial position of the objective was controlled by a focus drive (H122, Prior Scientific Inc., Rockland, MA), moving one-way from z ≈ 700 nm to −350 nm in 50 nm increments and 1 s intervals while a PBS movie was recorded, taking individual snapshots at every axial location; then, 50-image movies were acquired at the next three consecutive axial locations (0.3 s intervals), before a last movie was recorded for z ≈ – 500 nm to −750 nm. Here, positive defocusing distance z, is defined as when the surface-adsorbed PBS molecules are on the side of the focal plane closer to the glass coverslip. For PBS analyses, 35 × 35 pixel boxes centered at the molecule were selected by hand using IMAGEJ (NIH, Bethesda, MD); the center 25 × 25 pixels containing the PSF were used for 2D Gaussian fitting, and the peripheral pixels were used for background analysis. Only PBS molecules with signal-to-noise ratios (SNR) ≥ 2.5 were used for analysis. Before analysis, the camera’s intensity count at each pixel in an image was converted into photon count by using the camera-to-photon count conversion factor calibrated the same day of the measurement as described in our previous article [5]. The number of detected photons in an image was obtained by subtracting the total photon count of the background from the total photon count of the image. The PBS intensity profiles were fit to a 2D Gaussian function using a least squares curve fitting algorithm (lsqcurvefit) provided by MATLAB (The Mathworks, Natick, MA) in order to obtain the SD values of the molecule: where f0 is the multiplication factor, s and s are SDs in the x- and y- directions, respectively, x0 and y0 are the centroid location of the molecule, and 〈b〉 is the mean background offset in photons.

3. Results

In the PSF SD-based axial-localization method, the measured fluorophore’s PSF SD in the x-or y-direction, s, is frequently described by a symmetric [7-9] or an asymmetric function of the defocusing distance z [4, 10]. Since z is a function of s, by using a single image of the fluorophore, we should be able to represent the precision in z, Δz, by the precision in s, Δs, via error propagation calculation. Since we have recently obtained an expression for a PSF’s root mean square (rms) Δs as a function of where s0 is the PSF SD without the pixelation effect, pixel size a, number of detected photons N, and the mean 〈b〉 and variance , of the background’s photon noise [5] we can obtain Δz by studying a single image of a fluorophore. (Note that for the remainder of this article, we use i to denote x or y and Δs to denote Δs.) In Fig. 1(A), we show images of a single PBS molecule at different defocusing distances |z|, separated by 350 nm; it is obvious that the size of the image (or s) increases with |z|. Figure 1(B) plots the mean s values of 6 simultaneously imaged PBS molecules for z moving from ≈ 700 to −750 nm (the z limits were determined to be before the onset of PSF diffraction rings). The data fits well to the following expression where is the PBS SD at focus including the pixelation effect, A and B are fitting parameters which can be related to the focus depth of the microscope, d, as A = 1/d2 and B = B′/d4 with B′ being the coefficient of a higher order term to correct for the refractive index mismatch effect and the non-ideality of an imaging system [4, 17].
Fig. 1

PBS axial-localization precision studies. (A) Snapshots of a PBS molecule separated by 350 nm along z (the middle image is at z ≈ −50 nm). Scale bar, 500 nm. (B) Mean s versus z for 6 simultaneously imaged PBS molecules. The solid line is a fit to the data according to Eq. (3), yielding s′0 = 144.1 nm, A = 2.91 × 10−7 nm−2, and B = 1.87×10−11 nm−4. The y- and x-axis error bars are the average Δs and Δz values of the 6 PBS molecules. Note that the errors increase as z decreases because the PBS molecules gradually bleached with imaging time from 4800 to 1400 mean photons per PSF. (C) 50 consecutive s measurements for the PBS molecule in (A) at each of the three z locations in the blue circle in (B) (gray lines). The mean photon counts per image is ≈ 3000. (D) The corresponding z values to s values in (C) (gray lines). At each axial location in (C) and (D), the black horizontal lines outline the average s and z values, and the left (black) and right (red) error bars represent the respective experimental and theoretical Δs and Δz values. Insets to (C) and (D) show Gaussian fits to the distributions of the experimental s and z data for the middle axial location; the SDs of the fits (experimental error bars) are Δs = 6.3 nm and Δz = 21.5 nm, in good agreement with the theoretical values of Δs = 6.0 nm and Δz = 20.0 nm. Note that Δz is clearly less than the z increment size of 50 nm.

PBS axial-localization precision studies. (A) Snapshots of a PBS molecule separated by 350 nm along z (the middle image is at z ≈ −50 nm). Scale bar, 500 nm. (B) Mean s versus z for 6 simultaneously imaged PBS molecules. The solid line is a fit to the data according to Eq. (3), yielding s′0 = 144.1 nm, A = 2.91 × 10−7 nm−2, and B = 1.87×10−11 nm−4. The y- and x-axis error bars are the average Δs and Δz values of the 6 PBS molecules. Note that the errors increase as z decreases because the PBS molecules gradually bleached with imaging time from 4800 to 1400 mean photons per PSF. (C) 50 consecutive s measurements for the PBS molecule in (A) at each of the three z locations in the blue circle in (B) (gray lines). The mean photon counts per image is ≈ 3000. (D) The corresponding z values to s values in (C) (gray lines). At each axial location in (C) and (D), the black horizontal lines outline the average s and z values, and the left (black) and right (red) error bars represent the respective experimental and theoretical Δs and Δz values. Insets to (C) and (D) show Gaussian fits to the distributions of the experimental s and z data for the middle axial location; the SDs of the fits (experimental error bars) are Δs = 6.3 nm and Δz = 21.5 nm, in good agreement with the theoretical values of Δs = 6.0 nm and Δz = 20.0 nm. Note that Δz is clearly less than the z increment size of 50 nm. Having obtained s as a function of z, we proceed to express z as a function of s where , and then to calculate Δz by using the error propagation calculation of Δz = (∂z/∂s)Δs: Now we have a Δz expression as a function of experimental parameters of a single PSF. Using Eqs. (2), (3), and (5), we calculated and plotted the average Δs and Δz values for PBS molecules in Fig. 1 (including modification to the Δs values, and consequently the Δz values by the appropriate scaling factors; see Sec. 3.1) as the y- and x-axis error bars in Fig. 1(B), respectively. Since Δz diverges at low |z|, only Δz at |z| ≥ Δz(z) are shown. In order to validate our method in obtaining Δz by error propagation calculation of z(s), we have performed repeated measurements of a single PBS molecule at three axial locations near z = −400 nm separated by 50 nm. Figures 1(C) and 1(D) show the s and the corresponding z values [calculated from s using Eq. (4)], respectively, at the three axial locations. The two error bars at each axial location compare the calculated with the experimentally determined Δs and Δz [insets to Figs. 1(C) and 1(D)], showing agreement.

3.1. Theoretical Δsi scaling factor calculations

As shown in our previous article on Δs of single fluorophore images [5], the theoretical Δs values [Eq. (2)] underreports the true experimental Δs value. A scaling factor that depends on a/s0 should be multiplied to the theoretical Δs in order to obtain the expected experimental Δs. Although we have studied the a/s0 dependence of the scaling factor by varying a [5], in this article, it is s0 that changes at different z and the a/s0 dependence may vary. We have performed Δs simulations for a/s0 = 0.20 to 0.76, which covers our measured PBS a/s0 range of 0.20 to 0.55 for a = 79 nm, s0 = 404 nm, and s0 = 142 nm. The simulation results are shown in Fig. 2, where each Δs datum is the SD of the s distribution of 2000 simulated PSFs using the average PBS parameters of a = 79 nm, N = 3000, 〈 b〉 = 4, and σ = 1 photons. The simulated Δs results were compared to the calculated Δs results, yielding a scaling factor expression of 1.51 + 0.17a/s0. The higher limit of a/s0 was chosen to be 0.76, which is up to where the scaling factor expression remains valid.
Fig. 2

Simulation (circles) and theoretical (solid black line) Δs versus a/s0. The dashed (red) line is the theoretical Δs results multiplied by 1.51 + 0.17a/s0 as a best fit to the simulated Δs, showing agreement.

Simulation (circles) and theoretical (solid black line) Δs versus a/s0. The dashed (red) line is the theoretical Δs results multiplied by 1.51 + 0.17a/s0 as a best fit to the simulated Δs, showing agreement.

4. Discussion

Unlike the lateral-localization precision, which is independent of the lateral position (Δx, Refs. [5,6]), the axial-localization precision varies with z. In Fig. 3, we plot Δz versus |z| for PBS PSFs for a range of photon counts, providing a direct guideline for the axial-localization precisions that can be achieved by using single fluorophore images at specific axial locations and photon counts.
Fig. 3

PBS Δz versus |z| calculations according to Eqs. (2), (3), and (5) at photon counts N, of 100, 500, 1000, 5000, and 2×104 (top to bottom). No background noise is included. Only Δz at |z| ≥ Δz(z) are shown.

PBS Δz versus |z| calculations according to Eqs. (2), (3), and (5) at photon counts N, of 100, 500, 1000, 5000, and 2×104 (top to bottom). No background noise is included. Only Δz at |z| ≥ Δz(z) are shown. We discuss two features of Fig. 3 that will aid in generalizing our method to all SD-based axial-location precision studies using different fluorophores and experimental settings. (i) At high N and large |z|, the single-image determined Δz is in the nanometer range, allowing 3D localization measurements with simultaneous high spatial and temporal resolutions. (ii) As seen in Fig. 2, Δz diverges as |z| approaches 0 (this has been previously reported in Refs. [9, 12]). In order to obtain Δz at low |z|, one solution is to use astigmatism by introducing a cylindrical lens [23] that shifts the foci of s(z) and s(z) to be on opposite sides of z = 0 [4]. In Fig. 4(A) we shift the PBS s(z) and s(z) foci by 300 nm, assuming that s(z) and s(z) remain the same shape and measurable range after the cylindrical lens modification. Figure 4(B) plots the corresponding Δz(z) curves to the shifted s(z) and s(z) for a range of photon counts. At all measurable z for PBS, appropriate s(z) values can be used to obtain the minimum Δz, marked as bold lines, and nanometer precision for single-image axial localization measurements can be achieved at high photon counts.
Fig. 4

(A) PBS s(z) and s(z) curves with shifted foci at z = ±300 nm. (B) Δz versus z for the shifted s(z) and s(z) for N = 100, 500, 1000, 5000, and 2×104 photons (top to bottom). Only Δz at |z ± 300 nm| = Δz(z) at N = 500 and 2×104 photons are shown, and since both s and s are required for the Δs calculation, only Δz where both PBS s and s are valid in (A) are shown. The bold lines mark the minimal Δz values at all measurable z.

(A) PBS s(z) and s(z) curves with shifted foci at z = ±300 nm. (B) Δz versus z for the shifted s(z) and s(z) for N = 100, 500, 1000, 5000, and 2×104 photons (top to bottom). Only Δz at |z ± 300 nm| = Δz(z) at N = 500 and 2×104 photons are shown, and since both s and s are required for the Δs calculation, only Δz where both PBS s and s are valid in (A) are shown. The bold lines mark the minimal Δz values at all measurable z. The studies in this article are based on PBS molecules with a specific z(s) relation [Eq. (4)]. Because PBS differs from a single fluorophore in that it is an aggregate of 144 fluorophores, and our TIRF imaging setup further introduces the additional refractive index mismatch effect, our PBS s(z) relation, and thus our z(s) relation may differ slightly from those of single fluorophores at different depths in water (relative to the refractive index mismatch interface). Our group and two additional groups have observed that for single fluorophores in water, the s(z) curve is asymmetric and the degree of asymmetry varies with their depths [4, 10, 18, 24]. As a consequence, single fluorophore studies with different refractive index mismatch effects may have different z(s) and thus Δz relations. What our study has shown is that regardless of the sample or imaging setup, when a z(s) expression can be obtained for the molecule of interest, error propagation can be used to determine the axial-location precision from a single image of the fluorophore. In some complicated situations, such as when s(z) is asymmetric [4, 10, 18], multiple s(z) functions can be used to describe different regions of the s(z) curve in order to obtain simple z(s) expressions for Δz calculation. In summary, we have developed a precision expression for a fluorophore’s single-image axial localization measurement that includes all relevant experimental parameters. The improvement in temporal resolution will enable investigations of previously inaccessible SPT studies in the regime of milliseconds.
  18 in total

1.  Precise nanometer localization analysis for individual fluorescent probes.

Authors:  Russell E Thompson; Daniel R Larson; Watt W Webb
Journal:  Biophys J       Date:  2002-05       Impact factor: 4.033

2.  Three-dimensional tracking of fluorescent nanoparticles with subnanometer precision by use of off-focus imaging.

Authors:  Michael Speidel; Alexandr Jonás; Ernst-Ludwig Florin
Journal:  Opt Lett       Date:  2003-01-15       Impact factor: 3.776

3.  Single-image diffusion coefficient measurements of proteins in free solution.

Authors:  Shannon Kian Zareh; Michael C DeSantis; Jonathan M Kessler; Je-Luen Li; Y M Wang
Journal:  Biophys J       Date:  2012-04-03       Impact factor: 4.033

4.  Single molecule measurements of repressor protein 1D diffusion on DNA.

Authors:  Y M Wang; Robert H Austin; Edward C Cox
Journal:  Phys Rev Lett       Date:  2006-07-27       Impact factor: 9.161

Review 5.  Three-dimensional particle tracking via bifocal imaging.

Authors:  Erdal Toprak; Hamza Balci; Benjamin H Blehm; Paul R Selvin
Journal:  Nano Lett       Date:  2007-06-21       Impact factor: 11.189

6.  Effect of aberration on height calibration in three-dimensional localization-based microscopy and particle tracking.

Authors:  Yi Deng; Joshua W Shaevitz
Journal:  Appl Opt       Date:  2009-04-01       Impact factor: 1.980

7.  Structural organisation of phycobilisomes from Synechocystis sp. strain PCC6803 and their interaction with the membrane.

Authors:  Ana A Arteni; Ghada Ajlani; Egbert J Boekema
Journal:  Biochim Biophys Acta       Date:  2009-01-22

8.  High accuracy 3D quantum dot tracking with multifocal plane microscopy for the study of fast intracellular dynamics in live cells.

Authors:  Sripad Ram; Prashant Prabhat; Jerry Chao; E Sally Ward; Raimund J Ober
Journal:  Biophys J       Date:  2008-10-03       Impact factor: 4.033

Review 9.  Advances in single-molecule fluorescence methods for molecular biology.

Authors:  Chirlmin Joo; Hamza Balci; Yuji Ishitsuka; Chittanon Buranachai; Taekjip Ha
Journal:  Annu Rev Biochem       Date:  2008       Impact factor: 23.643

10.  Whole-cell 3D STORM reveals interactions between cellular structures with nanometer-scale resolution.

Authors:  Bo Huang; Sara A Jones; Boerries Brandenburg; Xiaowei Zhuang
Journal:  Nat Methods       Date:  2008-11-23       Impact factor: 28.547

View more
  3 in total

Review 1.  Precisely and accurately localizing single emitters in fluorescence microscopy.

Authors:  Hendrik Deschout; Francesca Cella Zanacchi; Michael Mlodzianoski; Alberto Diaspro; Joerg Bewersdorf; Samuel T Hess; Kevin Braeckmans
Journal:  Nat Methods       Date:  2014-03       Impact factor: 28.547

2.  Optimized protocol for combined PALM-dSTORM imaging.

Authors:  O Glushonkov; E Réal; E Boutant; Y Mély; P Didier
Journal:  Sci Rep       Date:  2018-06-08       Impact factor: 4.379

3.  Depth-resolved cellular microrheology using HiLo microscopy.

Authors:  Jarett Michaelson; Heejin Choi; Peter So; Hayden Huang
Journal:  Biomed Opt Express       Date:  2012-05-03       Impact factor: 3.732

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.