Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia.

O Mimoz1, T Benard, A Gaucher, D Frasca, B Debaene.   

Abstract

BACKGROUND: Respiratory rate should be monitored continuously in the post-anaesthesia care unit (PACU) to avoid any delay in the detection of respiratory depression. Capnometry is the standard of care but in extubated patients requires a nasal cannula or a face mask that may be poorly tolerated or can be dislodged, leading to errors in data acquisition and false alarms. The value of a new non-invasive acoustic monitor in this setting has not been fully investigated.
METHODS: Adult patients admitted to the PACU after general anaesthesia were included. After tracheal extubation, an adhesive sensor with an integrated acoustic transducer (RRa™) was placed on the patient's throat and connected to its monitor while the patient breathed through a face mask with a carbon dioxide sampling port (Capnomask™) connected to a capnometer. Both the acoustic monitor and the capnometer were connected to a computer to record one pair of data per second for up to 60 min.
RESULTS: Fifty-two patients, mean (range) age 54 (22-84) yr and BMI 26 (19-39) kg m(-2), were studied. Compared with capnometry, the bias and limits of agreement of the acoustic method were 0 (-1.4-1.4) bpm. The acoustic sensor was well tolerated while the face mask was removed by eight patients, leading to study discontinuation in two patients.
CONCLUSIONS: In extubated patients, continuous assessment of respiration rate with an acoustic monitor correlated well with capnometry.

Entities:  

Mesh:

Year:  2012        PMID: 22323525     DOI: 10.1093/bja/aer510

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  35 in total

1.  Remote monitoring of breathing dynamics using infrared thermography.

Authors:  Carina Barbosa Pereira; Xinchi Yu; Michael Czaplik; Rolf Rossaint; Vladimir Blazek; Steffen Leonhardt
Journal:  Biomed Opt Express       Date:  2015-10-16       Impact factor: 3.732

2.  Acoustic sensor versus electrocardiographically derived respiratory rate in unstable trauma patients.

Authors:  Shiming Yang; Ashley Menne; Peter Hu; Lynn Stansbury; Cheng Gao; Nicolas Dorsey; William Chiu; Stacy Shackelford; Colin Mackenzie
Journal:  J Clin Monit Comput       Date:  2016-06-07       Impact factor: 2.502

3.  Assessment of noninvasive acoustic respiration rate monitoring in patients admitted to an Emergency Department for drug or alcoholic poisoning.

Authors:  Youcef Guechi; Amélie Pichot; Denis Frasca; Fatima Rayeh-Pelardy; Jean-Yves Lardeur; Olivier Mimoz
Journal:  J Clin Monit Comput       Date:  2015-01-23       Impact factor: 2.502

4.  Comparison of Nellcor™ PM1000N and Masimo Radical-7® for detecting apnea in volunteers.

Authors:  Chanatthee Kitsiripant; Tomoko Fukada; Hiroko Iwakiri; Yuri Tsuchiya; Makoto Ozaki; Minoru Nomura
Journal:  J Anesth       Date:  2017-07-20       Impact factor: 2.078

5.  Flash mob research: a single-day, multicenter, resident-directed study of respiratory rate.

Authors:  Matthew W Semler; Daniel G Stover; Andrew P Copland; Gina Hong; Michael J Johnson; Michael S Kriss; Hannah Otepka; Li Wang; Brian W Christman; Todd W Rice
Journal:  Chest       Date:  2013-06       Impact factor: 9.410

6.  Acoustic method respiratory rate monitoring is useful in patients under intravenous anesthesia.

Authors:  Kentaro Ouchi; Shigeki Fujiwara; Kazuna Sugiyama
Journal:  J Clin Monit Comput       Date:  2016-01-12       Impact factor: 2.502

7.  Listening to breathing again!

Authors:  Michael Ramsay
Journal:  J Clin Monit Comput       Date:  2014-12       Impact factor: 2.502

8.  Non-invasive human vital signs monitoring based on twin-core optical fiber sensors.

Authors:  Fengze Tan; Shuyang Chen; Weimin Lyu; Zhengyong Liu; Changyuan Yu; Chao Lu; Hwa-Yaw Tam
Journal:  Biomed Opt Express       Date:  2019-10-29       Impact factor: 3.732

9.  Non-invasive continuous respiratory monitoring using temperature-based sensors.

Authors:  Daniel E Hurtado; Angel Abusleme; Javier A P Chávez
Journal:  J Clin Monit Comput       Date:  2019-06-03       Impact factor: 2.502

10.  Detection of respiratory compromise by acoustic monitoring, capnography, and brain function monitoring during monitored anesthesia care.

Authors:  Pedro P Tanaka; Maria Tanaka; David R Drover
Journal:  J Clin Monit Comput       Date:  2014-01-14       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.