Hydrostatic comparison of nonpenetrating titanium clips versus conventional suture for repair of spinal durotomies.

Nathan D Faulkner1, Michael A Finn, Paul A Anderson.   

Abstract

STUDY
DESIGN: Biomechanics.
OBJECTIVE: To compare the hydrostatic strength of suture and nonpenetrating titanium clip repairs of standard spinal durotomies. SUMMARY OF BACKGROUND DATA: Dural tears are a frequent complication of spine surgery and can be associated with significant morbidity. Primary repair of durotomies with suture typically is attempted, but a true watertight closure can be difficult to obtain because of leakage through suture tracts. Nonpenetrating titanium clips have been developed for vascular anastomoses and provide a close apposition of the tissues without the creation of a suture tract.
METHODS: Twenty-four calf spines were prepared with laminectomies and the spinal cord was evacuated leaving an intact dura. After Foley catheters were inserted from each end and inflated adjacent to a planned dural defect, the basal flow rate was measured and a 1-cm longitudinal durotomy was made with a scalpel. Eight repairs were performed for each material, which included monofilament suture, braided suture, and nonpenetrating titanium clips. The flow rate at 30, 60, and 90 cm of water and the time needed for each closure were measured.
RESULTS: There was no statistically significant difference in the baseline leak rate for all 3 groups. There was no difference in the leakage rate of durotomies repaired with clips and intact specimens at any pressure. Monofilament and braided suture repairs allowed significantly more leakage than both intact and clip-repaired specimens at all pressures. The difference in leak rate increased as the pressure increased. Closing the durotomy with clips took less than half the time of closure with suture.
CONCLUSION: Nonpenetrating titanium clips provide a durotomy closure with immediate hydrostatic strength similar to intact dura whereas suture repair with either suture was significantly less robust. The use of titanium clips was more rapid than that of suture repair.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22322376     DOI: 10.1097/BRS.0b013e31824cf756

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  8 in total

Review 1.  Spinal dura mater: biophysical characteristics relevant to medical device development.

Authors:  Sean J Nagel; Chandan G Reddy; Leonardo A Frizon; Matthieu K Chardon; Marshall Holland; Andre G Machado; George T Gillies; Matthew A Howard; Saul Wilson
Journal:  J Med Eng Technol       Date:  2018-03-23

2.  Primary Repair of Posteriorly Located Anterior Skull Base Dural Defects Using Nonpenetrating Titanium Clips in Cranial Trauma.

Authors:  Camille K Milton; Bethany J Andrews; Cordell M Baker; Kyle P O'Connor; Andrew K Conner; Michael E Sughrue; Kibwei A McKinney; Edward T El Rassi; Jose A Sanclement; Chad A Glenn
Journal:  J Neurol Surg B Skull Base       Date:  2020-11-26

3.  Mediastinal micro-vessels clipping during lymph node dissection may contribute to reduce postoperative pleural drainage.

Authors:  Shi Yan; Xing Wang; Chao Lv; Kevin Phan; Yuzhao Wang; Jia Wang; Yue Yang; Nan Wu
Journal:  J Thorac Dis       Date:  2016-03       Impact factor: 2.895

4.  Watertightness of wound closure in lumbar spine-a comparison of different techniques.

Authors:  Aline Suter; José Miguel Spirig; Paolo Fornaciari; Elias Bachmann; Tobias Götschi; Karina Klein; Mazda Farshad
Journal:  J Spine Surg       Date:  2019-09

5.  Effectiveness of Repair Techniques for Spinal Dural Tears: A Systematic Review.

Authors:  Elliot H Choi; Alvin Y Chan; Nolan J Brown; Brian V Lien; Ronald Sahyouni; Andrew K Chan; John Roufail; Michael Y Oh
Journal:  World Neurosurg       Date:  2021-02-25       Impact factor: 2.104

6.  Failure Pressures of Dural Repairs in a Porcine Ex Vivo Model: Novel Use of Titanium Clips Versus Tissue Glue.

Authors:  Kevin Chorath; Mason Krysinski; Leonid Bunegin; Jacob Majors; Erik Kent Weitzel; Kevin Christopher McMains; Philip G Chen
Journal:  Allergy Rhinol (Providence)       Date:  2019-09-30

7.  Tension pneumocephalus following endoscopic resection of a mediastinal thoracic spinal tumor: A case report.

Authors:  Chao-Yuan Chang; Cheng-Che Hung; Ju-Mien Liu; Cheng-Di Chiu
Journal:  World J Clin Cases       Date:  2022-01-14       Impact factor: 1.337

8.  Durotomy repair in minimally invasive transforaminal lumbar interbody fusion by nonpenetrating clips.

Authors:  Yen-Po Cheng; Ping-Yi Lin; Abel Po-Hao Huang; Chun-Yuan Cheng; Chien-Min Chen; Dueng-Yuan Hueng
Journal:  Surg Neurol Int       Date:  2014-03-20
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.