Characteristics of Gafchromic XRQA2 films for kV image dose measurement.

Tawfik Giaddui1, Yunfeng Cui, James Galvin, Wenzhou Chen, Yan Yu, Ying Xiao.   

Abstract

PURPOSE: In this study, the relevant characteristics of the new Gafchromic XRQA2 film for its application in measuring kV cone beam computed tomography (CBCT) image doses were thoroughly investigated.
METHODS: The film was calibrated free in air to air kerma levels between 0 and 9 cGy using 120 kVp photon beams produced by the x-ray volume imager. Films were scanned using transmission and reflection scanning modes with the Epson Expression 10000 XL flat-bed document scanner. The impact of film size, region of interest for the analysis, scan uniformity, scan resolution, scan orientation and alternate scanning sides on the analysis process were investigated. Energy dependence, postirradiation growth of reflectance with time and irradiation angular dependence of the film were tested at different air kerma levels.
RESULTS: The net reflectance changed by ∼3% when the size of the film piece changed from 1 cm × 2 cm to 10 cm × 11 cm and changed by ∼1% when ROI changed from 0. 7 cm × 0. 7 cm to 8 cm × 8 cm, suggesting a good uniformity of the film. The film was successfully analyzed using the transmission scanning mode, calibration curves from both transmission and reflection scanning modes showed similar behavior. The calibration uncertainty was somewhat lower when the film was scanned using reflection mode (6% and 8% for reflection and transmission modes, respectively.) Higher scanning resolution came with increasing calibration uncertainty. The calibration uncertainty for reflection and transmission modes increased from ∼3.5% to 7% and from ∼3.5% to 9%, respectively when scanning resolution was changed from 50 to 400 dpi. Scanning the film on alternate sides using transmission mode led to variation of 16%-19% in the net optical density at doses commonly used for CBCT procedures. The film response changed by almost 10% when it was exposed to beams of two different energies (100 and 120 kVp.) Other features of the film such as film orientation, postexposure growth, and irradiation angular dependence were also investigated.
CONCLUSIONS: The size of film piece and analysis ROI used for calibration slightly affected the film response. Both transmission and reflection scanning modes can be used to analyze the Gafchromic XRQA2, with the reflection mode having a somewhat lower calibration uncertainty. Scanning films on alternate sides using transmission mode significantly affects the optical density. The film response was shown to be energy dependent. The films reached stability in about 6 h after exposure. The film response was proven to be independent of irradiation angle except when the beam is parallel to the film surface.

Entities:  

Mesh:

Year:  2012        PMID: 22320794     DOI: 10.1118/1.3675398

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Investigation into image quality and dose for different patient geometries with multiple cone-beam CT systems.

Authors:  Stephen J Gardner; Matthew T Studenski; Tawfik Giaddui; Yunfeng Cui; James Galvin; Yan Yu; Ying Xiao
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

2.  Dosimetric changes with computed tomography automatic tube-current modulation techniques.

Authors:  Sofia Spampinato; Anna Maria Gueli; Pietro Milone; Luigi Angelo Raffaele
Journal:  Radiol Phys Technol       Date:  2018-04-06

3.  Dosimetry of a novel converging X-ray source for kilovoltage radiotherapy.

Authors:  Tyler Stalbaum; Larry Partain; Michael D Weil; Joon Kim; Jae Y Han; Michael J Plies; Hong Chen; Vitaliy Ziskin; Magdalena Bazalova-Carter; Samuel Song; Roy Rand; Douglas Boyd
Journal:  Med Phys       Date:  2021-08-25       Impact factor: 4.506

4.  Evaluation of different techniques for CT radiation profile width measurement.

Authors:  Steven R Jackson; Salahuddin Ahmad; Yida Hu; Chun Ruan
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

5.  Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies.

Authors:  Jarryd G Buckley; Dean Wilkinson; Alessandra Malaroda; Peter Metcalfe
Journal:  J Appl Clin Med Phys       Date:  2017-12-19       Impact factor: 2.102

6.  Gafchromic XR-QA2 film as a complementary dosimeter for hand-monitoring in CTF-guided biopsies.

Authors:  Sandra Sarmento; Joana Pereira; Maria José Sousa; Luís Cunha; Anabela G Dias; Miguel F Pereira; Augusto D Oliveira; João V Cardoso; Luis M Santos; Margarida Gouvêa; Joana Lencart; João G Alves; João A M Santos
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system.

Authors:  Christian Fedon; Marco Caballo; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2018-06-29       Impact factor: 4.071

8.  Real-time dosimetry with radiochromic films.

Authors:  Pierluigi Casolaro; Luigi Campajola; Giovanni Breglio; Salvatore Buontempo; Marco Consales; Andrea Cusano; Antonello Cutolo; Francesco Di Capua; Francesco Fienga; Patrizio Vaiano
Journal:  Sci Rep       Date:  2019-03-29       Impact factor: 4.379

9.  Radiochromic film based dosimetry of image-guidance procedures on different radiotherapy modalities.

Authors:  Ahmad Nobah; Saad Aldelaijan; Slobodan Devic; Nada Tomic; Jan Seuntjens; Mohammed Al-Shabanah; Belal Moftah
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

10.  Asymmetric breast dose in coronary angiography.

Authors:  Kam Shan Mong; Anthony B Wallace; Rick D Franich
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.