Physical evaluation of a needle photostimulable phosphor based CR mammography system.

Nicholas W Marshall1, Kim Lemmens, Hilde Bosmans.   

Abstract

PURPOSE: Needle phosphor based computed radiography (CR) systems promise improved image quality compared to powder phosphor based CR units for x-ray screening mammography. This paper compares the imaging performance of needle CR cassettes, powder based CR cassettes and a well established amorphous selenium (a-Se) based flat panel based mammography system, using consistent beam qualities.
METHODS: Detector performance was assessed using modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE). Mammography system performance was assessed against levels from the European Guidelines, including threshold gold thickness (c-d), relative signal difference to noise (SdNR) and mean glandular dose, for automatic exposure control settings suggested by the manufacturers. The needle based Agfa HM5.0 CR detector was compared against the single sided readout Agfa MM3.0R and dual sided readout Fuji Profect CS powder CR plates using a 28 kV Mo/Rh spectrum, while a 28 kV W/Rh spectrum was used to compare the Agfa HM5.0 against the Siemens MAMMOMAT Inspiration a-Se based system.
RESULTS: MTF at 5 mm(-1) was 0.16 and 0.24 for the needle CR detector in the fast and slow scan directions, respectively, indicating a slight improvement (∼20%) over the two powder CR systems but remained 50% lower than the result at 5 mm(-1) for the a-Se detector (∼0.55). Structured screen noise was lower for the needle phosphor compared to the powder plates. CR system gain, estimated from the measured absorption fraction and NNPS results, was 6.3 for the (single sided) needle phosphor and 5.1 and 7.2 for the single sided and dual sided powder phosphor systems. Peak DQE at ∼100 μGy was 0.47 for the needle system compared to peak DQE figures of 0.33 and 0.46 for the single sided readout powder plates and dual sided readout plates. The high frequency DQE (at 5 mm(-1)) was 0.19 for the needle CR plates, a factor of approximately 3 greater than for the powder CR plates. At 28 kV W/Rh, 2 mm Al, peak DQE for the needle CR system was 0.45 against a value of 0.50 for the a-Se detector. The needle CR detector reached the Acceptable limit for 0.1 mm details in the European Guidelines at a mean glandular dose (MGD) of approximately 1.31 mGy imaged at 28 kV Mo/Rh, compared to figures of 2.19 and 1.43 mGy for the single sided and dual sided readout powder CR systems. The a-Se detector could reach the limit at 0.65 mGy using a 28 kV W/Rh spectrum, while the needle CR system required 1.09 mGy for the same spectrum.
CONCLUSIONS: Imaging performance for the needle CR phosphor technology, characterized using MTF and DQE and threshold gold thickness demonstrated a clear improvement compared to both single and dual sided reading powder phosphor based CR systems.

Entities:  

Mesh:

Year:  2012        PMID: 22320791     DOI: 10.1118/1.3675403

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  The role of activator concentration and precipitate formation on optical and dosimetric properties of KCl:Eu(2+) storage phosphor detectors.

Authors:  Rachael A Hansel; Zhiyan Xiao; Yanle Hu; Olga Green; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

2.  Effect of image quality on calcification detection in digital mammography.

Authors:  Lucy M Warren; Alistair Mackenzie; Julie Cooke; Rosalind M Given-Wilson; Matthew G Wallis; Dev P Chakraborty; David R Dance; Hilde Bosmans; Kenneth C Young
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

3.  Technical and clinical breast cancer screening performance indicators for computed radiography versus direct digital radiography.

Authors:  Hilde Bosmans; An De Hauwere; Kim Lemmens; Federica Zanca; Hubert Thierens; Chantal Van Ongeval; Koen Van Herck; Andre Van Steen; Patrick Martens; Luc Bleyen; Gretel Vande Putte; Eliane Kellen; Griet Mortier; Erik Van Limbergen
Journal:  Eur Radiol       Date:  2013-05-21       Impact factor: 5.315

4.  On the relevance of modulation transfer function measurements in digital mammography quality control.

Authors:  Kristina T Wigati; Nicholas W Marshall; Kim Lemmens; Joke Binst; Annelies Jacobs; Lesley Cockmartin; Guozhi Zhang; Liesbeth Vancoillie; Dimitar Petrov; Dirk A N Vandenbroucke; Djarwani S Soejoko; Hilde Bosmans
Journal:  J Med Imaging (Bellingham)       Date:  2021-04-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.