Long-term efficacy of single procedure remote magnetic catheter navigation for ablation of ischemic ventricular tachycardia: a retrospective study.

Borislav Dinov1, Robert Schönbauer, Agnieska Wojdyla-Hordynska, Frieder Braunschweig, Sergio Richter, David Altmann, Philipp Sommer, Thomas Gaspar, Andreas Bollmann, Ulrike Wetzel, Sascha Rolf, Christopher Piorkowski, Gerhard Hindricks, Arash Arya.   

Abstract

BACKGROUND: Remote magnetic navigation (RMN) aims to reduce some inherent limitations of manual radiofrequency (RF) ablation. However, data comparing the effectiveness of both methods are scarce. This study evaluated the acute and long-term success of RMN guided versus manual RF ablation in patients with ischemic sustained ventricular tachycardia (sVT).
METHODS: One hundred two consecutive patients (age 68 ± 10 years, LVEF 32 ± 12%, 88 men) with ischemic sVT were ablated with RMN (Stereotaxis; 49%) or manually (51%) using substrate and/or activation mapping (Carto) and open-irrigated-tip catheters. All received implantable defibrillators or loop recorders. Acute success was defined as noninducibility of any sVT at the end of the ablation procedure and long-term success as freedom from VT upon follow-up.
RESULTS: There was no difference in the baseline characteristics between the groups. Three patients died in hospital. Acute success rate was similar for RMN and manual ablation (82% vs 71%, P = 0.246). RMN was associated with significantly shorter fluoroscopy time (13 ± 12 minutes vs 32 ± 17 minutes, P = 0.0001) and RF time (2337.59 ± 1248.22 seconds vs 1589.95 ± 1047.42 seconds, P = 0.049), although total procedure time was similar (157 ± 40 minutes vs 148 ± 50 minutes, P = 0.42). There was a nonsignificant trend toward better long-term success in RMN group: after a median of 13 (range 1-34) months, 63% in the RMN and 53% in the manual ablation group were free from VT recurrence (P = 0.206).
CONCLUSION: RMN guided RF ablation of ischemic sustained VT is equally efficient compared with manual ablation in terms of acute and long-term success rate. These results are achieved with a significantly reduced fluoroscopy time and shorter RF time.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22313170     DOI: 10.1111/j.1540-8167.2011.02243.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  15 in total

1.  Magnetic resonance imaging compatible remote catheter navigation system with 3 degrees of freedom.

Authors:  M A Tavallaei; M K Lavdas; D Gelman; M Drangova
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-12-24       Impact factor: 2.924

2.  Managing difficult anatomy: remote-controlled ablation of atrioventricular nodal reentry tachycardia in a patient with agenesis of the inferior vena cava.

Authors:  Burkhard Hügl; Dmitrij Velikan; Björn Buchter; Zdravena Findeisen
Journal:  Clin Res Cardiol       Date:  2013-06-15       Impact factor: 5.460

3.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Saenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  J Interv Card Electrophysiol       Date:  2020-10       Impact factor: 1.900

4.  [Non-fluoroscopic catheter tracking: the MediGuide™ system].

Authors:  Philipp Sommer; S Rolf; S Richter; G Hindricks; C Piorkowski
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2012-11-21

Review 5.  A meta-analysis of manual versus remote magnetic navigation for ventricular tachycardia ablation.

Authors:  Mohit K Turagam; Donita Atkins; Roderick Tung; Moussa Mansour; Jeremy Ruskin; Jie Cheng; Luigi Di Biase; Andrea Natale; Dhanunjaya Lakkireddy
Journal:  J Interv Card Electrophysiol       Date:  2017-06-17       Impact factor: 1.900

6.  Prediction and prognosis of ventricular tachycardia recurrence after catheter ablation with remote magnetic navigation for electrical storm in patients with ischemic cardiomyopathy.

Authors:  Qi Jin; Peter Karl Jacobsen; Steen Pehrson; Xu Chen
Journal:  Clin Cardiol       Date:  2017-08-17       Impact factor: 2.882

7.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Sáenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  Europace       Date:  2019-08-01       Impact factor: 5.214

8.  Radiofrequency atrial fibrillation ablation with irrigated tip catheter using remote magnetic navigation compared with conventional manual method.

Authors:  Rugheed Ghadban; Kristina Gifft; Zachary Luebbering; Sandeep Sodhi; Daniel Cooper; Tariq Enezate
Journal:  J Interv Card Electrophysiol       Date:  2020-09-22       Impact factor: 1.900

9.  Non-fluoroscopic catheter visualization using MediGuide™ technology: experience from the first 600 procedures.

Authors:  P Sommer; S Richter; G Hindricks; S Rolf
Journal:  J Interv Card Electrophysiol       Date:  2014-01-16       Impact factor: 1.900

10.  Remote Navigation for Complex Arrhythmia.

Authors:  Irina Suman-Horduna; Sonya V Babu-Narayan; Sabine Ernst
Journal:  Arrhythm Electrophysiol Rev       Date:  2013-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.