Cardiac pacemakers in electric and magnetic fields of 400-kV power lines.

Leena Korpinen1, Harri Kuisti, Jarmo Elovaara, Vesa Virtanen.   

Abstract

BACKGROUND: The cardiac pacemaker (PM) implantation rate per million is high. Earlier studies have found interference to PMs by electromagnetic fields. The aim of the study is to investigate disturbances in cardiac PM using a human-shaped phantom in electric and magnetic fields of 400-kV power lines.
METHODS: The phantom was used in the following manner: isolated from the ground, grounded from left or right foot, or grounded from left or right hand.
RESULTS: Out of the tested PMs one had such a disturbance that it set the pace 60 times per minute, when the electric field was 6.7-7.5 kV/m and the magnetic field was 2.4-2.9 μT. The electrode configuration of the PM was unipolar. In bipolar configuration, the same PM had no disturbance. During the test period, other PMs only had minor disturbances or none at all. Some PMs do not record time information for minor disturbances. In such cases, it was impossible to link the disturbances to the exposure under the power line.
CONCLUSIONS: The electric field under a 400-kV power line may disturb a PM. However, only one type out of several tested PMs showed a major disturbance and that was only with a unipolar electrode configuration. The risk of disturbances is therefore not deemed to be high. ©2012, The Authors. Journal compilation ©2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22309463     DOI: 10.1111/j.1540-8159.2011.03327.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Occupational exposure to electric and magnetic fields during tasks at ground or floor level at 110 kV substations in Finland.

Authors:  Leena Korpinen; Rauno Pääkkönen
Journal:  Int J Occup Saf Ergon       Date:  2016-04-14

2.  Computation of Pacemakers Immunity to 50 Hz Electric Field: Induced Voltages 10 Times Greater in Unipolar Than in Bipolar Detection Mode.

Authors:  Cihan Gercek; Djilali Kourtiche; Mustapha Nadi; Isabelle Magne; Pierre Schmitt; Martine Souques
Journal:  Bioengineering (Basel)       Date:  2017-03-06

3.  In Vivo Study of Electromagnetic Interference With Cardiac Contractility Modulation Devices at Power Frequency.

Authors:  Matthias Daniel Zink; Dominik Stunder; Tobias Theiler; Thomas Kraus; Nikolaus Marx; Andreas Napp
Journal:  J Am Heart Assoc       Date:  2021-08-18       Impact factor: 5.501

4.  Testing of common electromagnetic environments for risk of interference with cardiac pacemaker function.

Authors:  Maria Tiikkaja; Aapo L Aro; Tommi Alanko; Harri Lindholm; Heli Sistonen; Juha E K Hartikainen; Lauri Toivonen; Jukka Juutilainen; Maila Hietanen
Journal:  Saf Health Work       Date:  2013-07-20

5.  The Possibility of Decreasing 50-Hz Electric Field Exposure near 400-kV Power Lines with Arc Flash Personal Protective Equipment.

Authors:  Leena Korpinen; Herkko Pirkkalainen; Timo Heiskanen; Rauno Pääkkönen
Journal:  Int J Environ Res Public Health       Date:  2016-09-23       Impact factor: 3.390

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.